I C Ruset1, L L Tsai, R W Mair, S Patz, M I Hrovat, M S Rosen, I Muradian, J Ng, G P Topulos, J P Butler, R L Walsworth, F W Hersman.
Abstract
We describe a prototype system built to allow open-access very-low-field MRI of human lungs using laser-polarized (3)He gas. The system employs an open four-coil electromagnet with an operational B(0) field of 4 mT, and planar gradient coils that generate gradient fields up to 0.18 G/cm in the x and y direction and 0.41 G/cm in the z direction. This system was used to obtain (1)H and (3)He phantom images and supine and upright (3)He images of human lungs. We include discussion on challenges unique to imaging at 50 -200 kHz, including noise filtering and compensation for narrow-bandwidth coils.Entities:
Year: 2006 PMID: 20354575 PMCID: PMC2846659 DOI: 10.1002/cmr.b.20075
Source DB: PubMed Journal: Concepts Magn Reson Part B Magn Reson Eng ISSN: 1552-5031 Impact factor: 1.176