Literature DB >> 20351388

Vascular complications after percutaneous coronary intervention following hemostasis with the Mynx vascular closure device versus the AngioSeal vascular closure device.

Shah Azmoon1, Anthony L Pucillo, Wilbert S Aronow, Ramin Ebrahimi, Joseph Vozzolo, Archana Rajdev, Kumar Kalapatapu, Jae H Ro, Craig Hjemdahl-Monsen.   

Abstract

UNLABELLED: We investigated the prevalence of vascular complications after PCI following hemostasis in 190 patients (67% men and 33% women, mean age 64 years) treated with the AngioSeal vascular closure device (St. Jude Medical, Austin, Texas) versus 238 patients (67% men and 33% women, mean age 64 years) treated with the Mynx vascular closure device (AccessClosure, Mountain View, California).
RESULTS: Death, myocardial infarction or stroke occurred in none of the 190 patients (0%) treated with the AngioSeal versus none of 238 patients (0%) treated with the Mynx. Major vascular complications occurred in 4 of 190 patients (2.1%) treated with the AngioSeal versus 5 of 238 patients (2.1%) treated with the Mynx (p not significant). Major vascular complications in patients treated with the AngioSeal included removal of a malfunctioning device (1.1%), hemorrhage requiring intervention (0.5%) and hemorrhage with a loss of > 3g Hgb (0.5%). The major vascular complications in patients treated with the Mynx included retroperitoneal bleeding requiring surgical intervention (0.8%), pseudoaneurysm with surgical repair (0.8%) and hemorrhage with a loss of > 3g Hgb (0.4%). These complications were not significantly different between the two vascular closure devices (p = 0.77). Minor complications included hematoma > 5 cm (0.5%, n = 1) within the AngioSeal group, as well as procedure failure requiring > 30 minutes of manual compression after device deployment, which occurred in 7 out of 190 patients (3.7%) treated with the AngioSeal versus 22 of 238 patients with the Mynx (9.2%) (p = 0.033).
CONCLUSIONS: Major vascular complications after PCI following hemostasis with vascular closure devices occurred in 2.1% of 190 patients treated with the AngioSeal vascular closure device versus 2.1% of 238 patients treated with the Mynx vascular closure device (p not significant). The Mynx vascular closure device appears to have a higher rate of device failure.

Entities:  

Mesh:

Year:  2010        PMID: 20351388

Source DB:  PubMed          Journal:  J Invasive Cardiol        ISSN: 1042-3931            Impact factor:   2.022


  9 in total

1.  Delayed complication from a percutaneous vascular closure device following a neuro-interventional procedure.

Authors:  Ahmed Khaldi; B Waldau; C Skowlund; G J Velat; J Mocco; B L Hoh
Journal:  Interv Neuroradiol       Date:  2011-12-16       Impact factor: 1.610

2.  Large-scale experience with an anchorless vascular closure device in a real-life clinical setting.

Authors:  Verena Schelp; Sandra Freitag-Wolf; Dieter Hinzmann; Peter Bramlage; Norbert Frey; Derk Frank
Journal:  Clin Res Cardiol       Date:  2014-10-08       Impact factor: 5.460

3.  Initial experience with the AXERA 2 Femoral Access System in neurovascular procedures.

Authors:  Ramesh Grandhi; Nathan T Zwagerman; Xiaoran Zhang; Stephanie H Chen; Ashutosh P Jadhav; Tudor Jovin; Brian T Jankowitz
Journal:  Interv Neuroradiol       Date:  2015-05-26       Impact factor: 1.610

4.  Incidence of delayed angiographic femoral artery complications using the EXOSEAL vascular closure device.

Authors:  Ramesh Grandhi; Xiaoran Zhang; David Panczykowski; Phillip Choi; Christopher T Hunnicutt; Ashutosh P Jadhav; Andrew F Ducruet; Tudor Jovin; Brian Jankowitz
Journal:  Interv Neuroradiol       Date:  2015-05-26       Impact factor: 1.610

5.  Arterial closure device to achieve hemostasis in children following percutaneous femoral arterial puncture.

Authors:  Somnath J Prabhu; Siddharth A Padia; Karim Valji; Michael F McNeeley; Sandeep Vaidya; Nghia J Vo
Journal:  Pediatr Radiol       Date:  2013-01-16

6.  [Arterial closure devices. What device for which clinical situation?].

Authors:  M Treitl; K M Eberhardt; D Maxien; B Behrends; M F Reiser
Journal:  Radiologe       Date:  2013-03       Impact factor: 0.635

7.  Vascular closure device failure in contemporary practice.

Authors:  Venkatesan D Vidi; Michael E Matheny; Usha S Govindarajulu; Sharon-Lise T Normand; Susan L Robbins; Vikram V Agarwal; Sripal Bangalore; Frederic S Resnic
Journal:  JACC Cardiovasc Interv       Date:  2012-08       Impact factor: 11.195

8.  Successful Use of the MYNXGRIP Closure Device during Repeated Transbrachial Percutaneous Peripheral Intervention.

Authors:  Klaus Hertting; Werner Raut
Journal:  Case Rep Vasc Med       Date:  2015-08-26

9.  Endovascular repair of a common femoral artery puncture after losing access using a vascular closure device.

Authors:  Vamsi Reddy; Nathan Todnem; Camila Franco-Mesa; Young Erben; Rabih G Tawk; Charles Ritchie
Journal:  J Vasc Surg Cases Innov Tech       Date:  2021-05-21
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.