| Literature DB >> 20336452 |
Oluwatosin O Dada1, Lauren M Ramsay, Jane A Dickerson, Nathan Cermak, Rong Jiang, Cuiru Zhu, Norman J Dovichi.
Abstract
We report a multiplexed capillary electrophoresis system employing an array of 32 capillaries with a micromachined sheath-flow cuvette as the detection chamber. The sample streams were simultaneously excited with a 473-nm laser beam, and the fluorescence emission was imaged on a CCD camera with a pair of doublet achromat lens. The instrument produced mass detection limits of 380 +/- 120 yoctomoles for fluorescein in zone electrophoresis. Capillary isoelectric focusing of fluorescent standards produced peaks with an average width of 0.0029 +/- 0.0008 pH. Capillary coating stability limits the reproducibility of the analysis.Entities:
Mesh:
Substances:
Year: 2010 PMID: 20336452 PMCID: PMC3252051 DOI: 10.1007/s00216-010-3595-x
Source DB: PubMed Journal: Anal Bioanal Chem ISSN: 1618-2642 Impact factor: 4.142