| Literature DB >> 20212919 |
Renata Raina1, Michele L Etter.
Abstract
A new liquid chromatography (LC)-negative ion electrospray ionization (ESI(-))-tandem mass spectrometry (MS/MS) method with post-column addition of ammonia in methanol has been developed for the analysis of acid herbicides: 2,4-dichlorophenoxy acetic acid, 4-chloro-o-tolyloxyacetic acid, 2-(2-methyl-4-chlorophenoxy)butyric acid, mecoprop, dichlorprop, 4-(2,4-dichlorophenoxy) butyric acid, 2,4,5-trichlorophenoxy propionic acid, dicamba and bromoxynil, along with their degradation products: 4-chloro-2-methylphenol, 2,4-dichlorophenol, 2,4,5-trichlorophenol and 3,5-dibromo-4-hydroxybenzoic acid. The samples were extracted from the surface water matrix using solid-phase extraction (SPE) with a polymeric sorbent and analyzed with LC ESI(-) with selected reaction monitoring (SRM) using a three-point confirmation approach. Chromatography was performed on a Zorbax Eclipse XDB-C18 (50 x 4.6 mm i.d., 1.8 mum) with a gradient elution using water-methanol with 2 mM ammonium acetate mobile phase at a flow rate of 0.15 mL/min. Ammonia in methanol (0.8 M) was added post-column at a flow rate of 0.05 mL/min to enhance ionization of the degradation products in the MS source. One SRM transition was used for quantitative analysis while the second SRM along with the ratio of SRM1/SRM2 within the relative standard deviation determined by standards for each individual pesticide and retention time match were used for confirmation. The standard deviation of ratio of SRM1/SRM2 obtained from standards run on the day of analysis for different phenoxyacid herbicides ranged from 3.9 to 18.5%. Limits of detection (LOD) were between 1 and 15 ng L(-1) and method detection limits (MDL) with strict criteria requiring <25% deviation of peak area from best-fit line for both SRM1 and SRM2 ranged from 5 to 10 ng L(-1) for acid ingredients (except dicamba at 30 ng L(-1)) and from 2 to 30 ng L(-1) for degradation products. The SPE-LC-ESI(-) MS/MS method permitted low nanogram-per-liter determination of pesticides and degradation products for surface water samples.Entities:
Keywords: liquid chromatography-tandem mass spectrometry; pesticides; phenoxyacid herbicides; post-column reagent addition; water analysis
Year: 2010 PMID: 20212919 PMCID: PMC2832344 DOI: 10.4137/aci.s3148
Source DB: PubMed Journal: Anal Chem Insights ISSN: 1177-3901
Liquid chromatography-tandem mass spectrometry (LC/MS/MS) retention time (tR), and selected reaction monitoring (SRM) transitions using electrospray ionization (ESI) in negative ion mode for phenoxyacid herbicides, their degradation products, and surrogate and internal standard.
| Acid herbicides | ||
| 1. Dicamba | 6.85 | |
| 2. Bromoxynil | 7.82 | |
| 3. MCPA | 9.20 | |
| 4. 2,4-D | 9.34 | |
| 5. Mecoprop | 10.72 | |
| 6. Dichlorprop | 11.13 | |
| 7. 2,4-DB | 14.58 | |
| 8. MCPB | 14.72 | |
| 9. 2,4,5-TP | 15.55 | |
| Degradation products | ||
| 10. DBHBA | 6.16 | |
| 11. CMP | 23.20 | |
| 12. DCP | 24.99 | |
| 13. TCP | 32.45 | |
| Internal/Surrogate standards | ||
| 15. d5-2,4-D | 9.20 | |
| 16. 13C6-2,4-D | 9.20 | |
| 17. 13C6-2,4,5-T | 12.10 | |
First SRM transition (in bold): quantitative SRM, second SRM: qualitative SRM.
Recoveries (%) obtained after solid-phase extraction with ChromP as a function of sample pH.
| Dicamba | 100 | 86 | 73 | 24 | 28 | 13 | 81 |
| 2,4-D | 88 | 80 | 79 | 86 | 79 | 85 | 93 |
| 2,4-DB | 92 | 78 | 79 | 77 | 75 | 76 | 85 |
| Dichlorprop | 89 | 79 | 82 | 82 | 72 | 82 | 100 |
| MCPA | 86 | 79 | 81 | 84 | 80 | 83 | 96 |
| 2,4,5-TP | 93 | 82 | 79 | 81 | 79 | 85 | 100 |
| Mecoprop | 89 | 79 | 79 | 85 | 78 | 88 | 100 |
| Bromoxynil | 78 | 71 | 69 | 79 | 69 | 71 | 96 |
| MCPB | 84 | 79 | 77 | 77 | 80 | 75 | 92 |
SPE conditions: 1 L of water spiked at 75 ng L−1 and eluted with 40/60 v/v% methanol/ethylacetate (F1 fraction); optimized SPE method conditions: 1 L of water spiked at 75 ng L−1 and eluted with 8 mL of 60/40 v/v% methanol/ethyl acetate (F1 fraction).
Figure 1.Percentage response of degradation products relative to response when no post-column reagent system is used as a function of molarity of ammonia in methanol. Degradation Products: DBHBA measured at 295.0 > 250.3; DCP at 160.9 > 124.7; CMP at 140.9 > 105.2; and TCP at 194.6 > 158.9. Percentage response is response of post-column system/response no post-column system multiplied by 100. LC-ESI−-SRM response shown is average peak height of three replicates. Sample injection 5 μL degradation product at 750 μg L−1 with post-column flow of 50 μL min−1 and column flow of 150 μL min−1.
Correlation coefficient (R2) for Calibration Curves and Method Detection Limits (MDLs) of phenoxyacid herbicides and their degradation products at two different ammonia concentrations in methanol for the post-column reagent. Flow rate of post-column reagent held constant at 0.50 mL min−1, and mobile phase flow rate 0.15 mL min−1. Standards prepared at 1, 2, 5,10, 20, 30, 50, 70, 90, 110, 130, and 150 ng L−1.
| Dicamba | 0.978/0.904 | 0.976/0.816 | 70/70 | |
| Bromoxynil | 0.985/0.985 | 0.988/0.994 | 10/10 | |
| MCPA | 0.996/0.992 | 0.993/0.980 | 5/10 | |
| 2,4-D | 0.994/0.993 | 0.995/0.985 | 10/15 | |
| Mecoprop | 0.998/0.981 | 0.998/0.977 | 5/20 | |
| Dichlorprop | 0.993/0.994 | 2/5 | 0.996/0.998 | |
| 2,4-DB | 0.979/0.991 | 0.975/0.994 | 20/2 | |
| MCPB | 0.995/0.991 | 0.996/0.976 | 5/30 | |
| 2,4,5-TP | 0.995/0.993 | 10/10 | 0.998/0.997 | |
| DBHBA | 0.988/0.969 | 0.994/0.889 | 50/30 | |
| CMP | 0.987/0.970 | 0.959/0.942 | 20/20 | |
| DCP | 0.977/0.977 | 10/10 | 0.986/0.995 | |
| TCP | 0.995/0.994 | 0.996/0.995 | ||
Note: Underlined are the best overall conditions for MDL.
Figure 2.Selected Reaction Monitoring (SRM) chromatograms of phenoxyacid herbicides and degradation products. Shown are quantitative SRMs with SRM transition labelled for each chromatogram. Note SRM at 140.9 > 105.2 and 160.9 > 124.7 are also the qualitative SRM transition for some analytes. Peak numbering for phenoxyacid herbicides are as follows: 1. dicamba; 2. bromoxynil; 3. MCPA; 4. 2,4-D; 5. mecoprop; 6. dichlorprop; 7. 2,4-DB; 8. MCPB; 9. 2,4,5-TP; and degradation products: 10. DBHBA; 11. CMP; 12. DCP; 13. TCP. See Table 1 for retention times. Standard 75 ng L−1.
Limit of Detection (LOD), MDL and ratio of response of SRM1/SRM2 for phenoxyacid herbicides and their degradation products at optimal ammonia in Methanol Concentration. LOD and MDL highest reported level for both SRM1 and SRM2 (see MDL from Table 3); SRM1/SRM2 ± relative standard deviation (RSD) is a response of peak areas obtained on the day of analysis provided here for analysis of samples shown in Figure 1 and 2. Standard spike level is 1 ng L−1, except for bromoxynil and MCPB at 5 ng L−1, DBHBA, CMP, DCP at 15 ng L−1, and dicamba at 20 ng L−1. Ratios determined over linear dynamic range of MDL-100 ng L−1. Ammonia concentration of 0.80 M in methanol.
| Dicamba | 13 | 30 | 3.76 ± 18.5% |
| Bromoxynil | 5 | 5 | 1.07 ± 6.7% |
| MCPA | 1 | 5 | 41.4 ± 6.2% |
| 2,4-D | 1 | 10 | 4.29 ± 5.3% |
| Mecoprop | 1 | 10 | 49.2 ± 8.7% |
| Dichlorprop | 1 | 5 | 1.69 ± 3.9% |
| 2,4-DB | 5 | 5 | 0.70 ± 8.2% |
| MCPB | 5 | 10 | 5.60 ± 10.1% |
| 2,4,5-TP | 1 | 10 | 7.59 ± 7.2% |
| DBHBA | 15 | 30 | 1.92 ± 9.1% |
| CMP | 15 | 20 | 3.00 ± 13.4% |
| DCP | 15 | 10 | 3.80 ± 12.9% |
| TCP | 1 | 2 | 1.90 ± 5.0% |
Figure 3.Selected Monitoring Chromatograms (SRMs) for Sample WP216 containing 2,4-D and its degradation product DCP, and mecoprop and its degradation product CMP. Shown are both the quantitative and qualitative SRM transitions (SRM1 and SRM2) for analytes with expanded x-scale scale for the degradation products DCP and CMP. Quantitative and Qualitative SRM Transitions: 2,4-D are 219.0 > 174.9 and 174.9 > 144.9; DCP are 160.9 > 124.7 and 163.0 > 124.7; mecoprop are 213.0 > 140.9 and 140.9 > 105.2; and CMP are 140.9 > 105.2 and 142.8 > 105.2 (see Table 1). Analytes identified: 4. 2,4-D; 5. mecoprop; 11. CMP; and 12. DCP. Sample WP216 diluted at 1:2 with addition of IS for 5 μL injection.
Levels of phenoxyacid herbicides and their degradation products in surface water samples collected from two storm water ponds in the city of Regina. Average level determined for samples above detection limits. Shown only for pesticides detected. Total number of samples is 86.
| MCPA | 82 | 379 | 11 | 61 |
| 2,4-D | 626 | 3267 | 36 | 86 |
| Mecoprop | 730 | 2403 | 15 | 86 |
| Dichlorprop | 261 | 1260 | 47 | 12 |
| CMP | 45 | 101 | 22 | 11 |