Literature DB >> 10428231

Prodromes of failure in total knee arthroplasty.

J H Lonner1, J M Siliski, R D Scott.   

Abstract

A total of 102 revision total knee arthroplasties (TKAs) were reviewed to determine the prodromal symptoms and radiographic findings associated with failure. Presenting symptoms included pain (84%); swelling (76%); progressive varus or valgus deformity (19%); instability (17%); stiffness (17%); clicking or grinding (7%); catching (4%); and patellar pain, subluxation, or clicking (4%). Radiographs were diagnostic in 91% of cases, demonstrating complete radiolucencies (80%), polyethylene wear (43%), component breakage (5%), metallic debris (3%), patellar subluxation or dislocation (4%), and osteolysis (4%). Gross intraoperative findings included polyethylene wear (72%), osteolysis (22%), metal-wear synovitis (9%), component breakage (6%), patellar wear and dissociation (4%), and occult sepsis (5%). The average duration of symptoms was 13 months; the interval between orthopedic evaluations averaged 23 months. Based on this information, we recommend that an annual questionnaire and weight-bearing radiographs be used to ensure adequate surveillance of TKA patients and avoid complications associated with delay in diagnosis of polyethylene wear or implant loosening.

Entities:  

Mesh:

Year:  1999        PMID: 10428231     DOI: 10.1016/s0883-5403(99)90106-7

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  14 in total

1.  Knee wear measured in retrievals: a polished tray reduces insert wear.

Authors:  Daniel J Berry; John H Currier; Michael B Mayor; John P Collier
Journal:  Clin Orthop Relat Res       Date:  2012-07       Impact factor: 4.176

Review 2.  Imaging near orthopedic hardware.

Authors:  Matthew F Koff; Alissa J Burge; Kevin M Koch; Hollis G Potter
Journal:  J Magn Reson Imaging       Date:  2017-02-02       Impact factor: 4.813

3.  Highly crosslinked polyethylene is safe for use in total knee arthroplasty.

Authors:  Jeffrey T Hodrick; Erik P Severson; Deborah S McAlister; Brian Dahl; Aaron A Hofmann
Journal:  Clin Orthop Relat Res       Date:  2008-09-10       Impact factor: 4.176

Review 4.  Causes of failure and etiology of painful primary total knee arthroplasty.

Authors:  Romain Seil; Dietrich Pape
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2011-08-11       Impact factor: 4.342

5.  How are those "lost to follow-up" patients really doing? A compliance comparison in arthroplasty patients.

Authors:  Jung Keun Choi; Jeffrey A Geller; David A Patrick; Wenbao Wang; William Macaulay
Journal:  World J Orthop       Date:  2015-01-18

6.  Two stage fracture of a polyethylene post in a 9-year-old posterior-stabilized knee prosthesis: a case report.

Authors:  Fabio D'Angelo; Daniele Marcolli; Paolo Bulgheroni; Luigi Murena; Terenzio Congiu; Paolo Cherubino
Journal:  J Med Case Rep       Date:  2010-02-23

7.  Weight of polyethylene wear particles is similar in TKAs with oxidized zirconium and cobalt-chrome prostheses.

Authors:  Young-Hoo Kim; Jun-Shik Kim; Wansoo Huh; Kwang-Hoon Lee
Journal:  Clin Orthop Relat Res       Date:  2009-12-01       Impact factor: 4.176

8.  Polyethylene wear is influenced by manufacturing technique in modular TKA.

Authors:  Adolph V Lombardi; Bradley S Ellison; Keith R Berend
Journal:  Clin Orthop Relat Res       Date:  2008-09-13       Impact factor: 4.176

9.  No clinical benefit of titanium nitride coating in cementless mobile-bearing total knee arthroplasty.

Authors:  Ruud P van Hove; Richard M Brohet; Barend J van Royen; Peter A Nolte
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2014-10-05       Impact factor: 4.342

10.  Dislocation following total knee arthroplasty: A report of six cases.

Authors:  Manuel Villanueva; Antonio Ríos-Luna; Javier Pereiro; Homid Fahandez-Saddi; Antonio Pérez-Caballer
Journal:  Indian J Orthop       Date:  2010-10       Impact factor: 1.251

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.