Intraocular pressure values obtained by ocular response analyzer, dynamic contour tonometry, and goldmann tonometry in keratokonic corneas.

Atilla Bayer1, Afsun Sahin, Volkan Hürmeriç, Gökhan Ozge.   

Abstract

PURPOSE: To determine the agreement between dynamic contour tonometer (DCT), Goldmann applanation tonometer (GAT), and Ocular Response Analyzer (ORA) in keratoconic corneas and to find out the effect of corneal biomechanics on intraocular pressure (IOP) measurements obtained by these devices. SUBJECTS AND METHODS: IOP was measured with the ORA, DCT, and GAT in random order in 120 eyes of 61 keratoconus patients. Central corneal thickness (CCT) and keratometry were measured after all IOP determinations had been made. The mean IOP measurement by the ORA and DCT was compared with the measurement by the GAT, using Student t test. Bland-Altman analysis was performed to assess the clinical agreement between these methods. The effect of corneal hysteresis (CH), corneal resistance factor (CRF), and CCT on measured IOP was explored by multiple backward stepwise linear regression analysis.
RESULTS: The mean±SD patient age was 30.6±11.2 years. The mean±SD IOP measurement obtained with GAT, ORA Goldmann-correlated IOP (IOPg), ORA corneal-compensated IOP (IOPcc), and DCT was 10.96±2.8, 10.23±3.5, 14.65±2.8, and 15.42±2.7 mm Hg, respectively. The mean±SD CCT was 464.08±58.4 microns. The mean difference between IOPcc and GAT (P<0.0001), IOPcc and DCT (P<0.001), GAT and DCT (P<0.0001), IOPg and GAT (P<0.002), and IOPg and DCT (P<0.0001), was highly statistically significant. In multivariable regression analysis, DCT IOP and GAT IOP measurements were significantly associated with CH and CRF (P<0.0001 for both).
CONCLUSIONS: DCT seemed to be affected by CH and CRF, and the IOP values tended to be higher when compared with GAT. ORA-measured IOPcc was found to be independent of CCT and suitable in comparison to the DCT in keratoconic eyes.

Entities:  

Mesh:

Year:  2010        PMID: 20164793     DOI: 10.1097/IJG.0b013e3181ca7aeb

Source DB:  PubMed          Journal:  J Glaucoma        ISSN: 1057-0829            Impact factor:   2.503


  8 in total

1.  [New aspects on biomechanics of the cornea in keratoconus].

Authors:  Z Gatzioufas; B Seitz
Journal:  Ophthalmologe       Date:  2013-09       Impact factor: 1.059

2.  The influence of corneal geometrical and biomechanical properties on tonometry readings in keratoconic eyes.

Authors:  Mustafa Değer Bilgeç; Eray Atalay; Ömer Sözer; Hüseyin Gürsoy; Muzaffer Bilgin; Nilgün Yıldırım
Journal:  Int Ophthalmol       Date:  2019-12-02       Impact factor: 2.031

3.  Corneal properties in children with congenital isolated growth hormone deficiency.

Authors:  Pinar Nalcacioglu-Yuksekkaya; Emine Sen; Ufuk Elgin; Mumin Hocaoglu; Faruk Ozturk; Sebahat Agladıoglu Yilmaz; Havva Nur Kendirci; Semra Cetinkaya; Zehra Aycan
Journal:  Int J Ophthalmol       Date:  2014-04-18       Impact factor: 1.779

4.  Which tonometry in eyes with keratoconus?

Authors:  H Altinkaynak; C Kocasarac; H Dundar; N Sayin; N Kara; E Bozkurt; N Duru
Journal:  Eye (Lond)       Date:  2015-12-04       Impact factor: 3.775

5.  Evaluation of Goldmann applanation tonometry, rebound tonometry and dynamic contour tonometry in keratoconus.

Authors:  Fatih Özcura; Nilgün Yıldırım; Emre Tambova; Afsun Şahin
Journal:  J Optom       Date:  2016-07-09

6.  Reply to comment by De Bernardo and Rosa on "Evaluation of Goldmann applanation tonometry, rebound tonometry and dynamic contour tonometry in keratoconus".

Authors:  Fatih Özcura; Nilgün Yıldırım
Journal:  J Optom       Date:  2017-12-15

7.  Distribution of central corneal thickness and its association with ocular parameters in a large central European cohort: the Gutenberg health study.

Authors:  Esther M Hoffmann; Julia Lamparter; Alireza Mirshahi; Heike Elflein; René Hoehn; Christian Wolfram; Katrin Lorenz; Max Adler; Philipp S Wild; Andreas Schulz; Barbara Mathes; Maria Blettner; Norbert Pfeiffer
Journal:  PLoS One       Date:  2013-08-01       Impact factor: 3.240

8.  Inaccuracy of intraocular pressure measurement in congenital corneal opacity: three case reports.

Authors:  Byeong Soo Kang; Jin Wook Jeoung; Joo Youn Oh
Journal:  BMC Ophthalmol       Date:  2020-01-02       Impact factor: 2.209

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.