Assessment of toric intraocular lens alignment by a refractive power/corneal analyzer system and slitlamp observation.

Paul J Carey1, Antonio Leccisotti, Victoria E McGilligan, Ed A Goodall, C B Tara Moore.   

Abstract

PURPOSE: To assess the validity of an internal optical path difference map of a refractive power/corneal analyzer system in determining the alignment of toric intraocular lenses (IOLs). SETTINGS: Private practices, Spring Hill, Brisbane, and Chermside, Australia.
METHODS: This retrospective study comprised patients with more than 1.5 diopters of preexisting corneal astigmatism who had phacoemulsification and AcrySof toric IOL implantation. Preoperatively, the surgical eye was marked at the slitlamp microscope using a 4-point technique. The desired IOL orientation was marked with a Mendez marker based on the steep corneal axis. The toric IOL axis was measured 3 weeks postoperatively by rotating the slitlamp beam to align with the IOL axis indicator marks and using the Internal OPD Map on the Nidek OPD-Scan system. Uncorrected (UDVA) and corrected (CDVA) distance visual acuities, residual refractive sphere, and residual keratometric and refractive cylinders were also measured at 3 weeks.
RESULTS: Postoperatively, the mean UDVA was 0.17 logMAR +/- 0.18 (SD) and the mean CDVA, -0.01 +/- 0.12 logMAR; 88.2% of eyes had a UDVA of 0.3 or better, and no eye lost lines of visual acuity. There was an 82.33% reduction in defocus equivalent and a 64.62% reduction in refractive cylinder. The mean IOL misalignment measured by slitlamp was 2.55 +/- 2.76 degrees and by the internal map, 2.65 +/- 1.98 degrees. The correlation between the 2 methods was highly significant (r = 0.99, P<.001).
CONCLUSIONS: Both refractive power/corneal analyzer system and slitlamp observation were reliable and predictable methods of assessing IOL alignment. The 4-point preoperative marking technique yielded clinically acceptable, accurate toric IOL alignment. Copyright 2010 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2010        PMID: 20152601     DOI: 10.1016/j.jcrs.2009.08.033

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  12 in total

1.  Optical performance of toric intraocular lenses in the presence of decentration.

Authors:  Bin Zhang; Jin-Xue Ma; Dan-Yan Liu; Ying-Hua Du; Cong-Rong Guo; Yue-Xian Cui
Journal:  Int J Ophthalmol       Date:  2015-08-18       Impact factor: 1.779

2.  Comparison of clinical outcomes among 3 marking methods for toric intraocular lens implantation.

Authors:  Hideyuki Onishi; Hidemasa Torii; Kazuhiro Watanabe; Kazuo Tsubota; Kazuno Negishi
Journal:  Jpn J Ophthalmol       Date:  2016-03-09       Impact factor: 2.447

3.  Analysis of internal astigmatism and higher order aberrations in eyes implanted with a new diffractive multifocal toric intraocular lens.

Authors:  Peter Mojzis; David P Piñero; Vera Ctvrteckova; Iva Rydlova
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2012-05-17       Impact factor: 3.117

4.  A comparison of three different corneal marking methods used to determine cyclotorsion in the horizontal meridian.

Authors:  Hung-Yuan Lin; Yi-Ting Fang; Ya-Jung Chuang; Justin N Karlin; Hsin-Yang Chen; Szu-Yuan Lin; Pi-Jung Lin; Ming Chen
Journal:  Clin Ophthalmol       Date:  2017-02-08

5.  Verion digital marking versus smartphone-assisted manual marking and isolated manual marking in toric intraocular lens implantation.

Authors:  Zain Irfan Khatib; Suhas S Haldipurkar; Vijay Shetty
Journal:  Indian J Ophthalmol       Date:  2020-03       Impact factor: 1.848

6.  Visual outcome and rotational stability of open loop toric intraocular lens implantation in Indian eyes.

Authors:  Arvind Venkataraman
Journal:  Indian J Ophthalmol       Date:  2013-11       Impact factor: 1.848

7.  Management of moderate and severe corneal astigmatism with AcrySof® toric intraocular lens implantation - Our experience.

Authors:  Javed Hussain Farooqui; Archana Koul; Ranjan Dutta; Noshir Minoo Shroff
Journal:  Saudi J Ophthalmol       Date:  2015-07-30

8.  Evaluation of Scheimpflug imaging system as an added tool in improving the accuracy of reference marking (as compared to the slit lamp marking system) for toric intraocular lens implantation.

Authors:  Ajay Sharma; Akanksha Batra
Journal:  Indian J Ophthalmol       Date:  2020-04       Impact factor: 1.848

9.  Mobile App for Pre- and Intraoperative Assessment of Astigmatism Correction with Toric Intraocular Lenses in Cataract Surgery.

Authors:  Francisco Aecio Fernandes Dias; Vinicius Jose Fernandes Dias; Barbara de Araujo Lima Dutra; Anna Christina Siqueira Marques; Edgar Marçal; Abrahao da Rocha Lucena; Joao Crispim Ribeiro
Journal:  J Ophthalmol       Date:  2020-03-15       Impact factor: 1.909

10.  Comparison Between Digital and Manual Marking for Toric Intraocular Lenses: A Randomized Trial.

Authors:  Abdel Hamid Elhofi; Hany Ahmed Helaly
Journal:  Medicine (Baltimore)       Date:  2015-09       Impact factor: 1.817

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.