Increases in whole blood glucose measurements using optically based self-monitoring of blood glucose analyzers due to extreme Canadian winters.

George C Cembrowski1, Barbara Smith, Ellen M O'Malley.   

Abstract

BACKGROUND: Temperature and humidity have been reported to influence the results of whole blood glucose (WBG) measurements.
METHODS: To determine whether patient WBG values were affected by seasonal variation, we conducted a retrospective analysis of 3 years' worth of weekly averages of patient WBG in five Edmonton hospitals.
RESULTS: In all five hospitals, the winter WBG averages were consistently higher than the summer WBG averages, with the differences varying between 5% and 9%. Whole blood glucose averages were negatively correlated with the outside temperature. This seasonal variation was not observed in weekly patient averages of specimens run in a central hospital laboratory.
INTERPRETATION: It is probable that the seasonal variation of WBG arises from the very low indoor humidities that are associated with external subzero temperatures. These increases in WBG in cold weather may be due to limitations in the WBG measuring systems when operated in decreased humidities and/or increased evaporation of the blood sample during the blood glucose measurement process. The implications of this seasonal variation are significant in that it (1) introduces increased variability in patient WBG, (2) may result in increased glucose-lowering therapy during periods of external cold and low indoor humidity, and (3) confounds evaluations of WBG meter technology in geographic regions of subzero temperature and low indoor humidity. To mitigate the risk of diagnosing and treating factitious hyperglycemia, the humidity of patient care areas must be strictly controlled. Copyright 2009 Diabetes Technology Society.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 20144309      PMCID: PMC2769945          DOI: 10.1177/193229680900300407

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  19 in total

1.  Standards of medical care in diabetes.

Authors: 
Journal:  Diabetes Care       Date:  2005-01       Impact factor: 19.112

2.  Performance of glucose dehydrogenase-and glucose oxidase-based blood glucose meters at high altitude and low temperature.

Authors:  Daniel Oberg; Claes-Göran Ostenson
Journal:  Diabetes Care       Date:  2005-05       Impact factor: 19.112

3.  Identifying variables associated with inaccurate self-monitoring of blood glucose: proposed guidelines to improve accuracy.

Authors:  R Bergenstal; J Pearson; G S Cembrowski; D Bina; J Davidson; S List
Journal:  Diabetes Educ       Date:  2000 Nov-Dec       Impact factor: 2.140

4.  Performance of seven blood glucose testing systems at high altitude.

Authors:  B P Giordano; W Thrash; L Hollenbaugh; W P Dube; C Hodges; A Swain; C R Banion; G J Klingensmith
Journal:  Diabetes Educ       Date:  1989 Sep-Oct       Impact factor: 2.140

5.  Consensus statement on self-monitoring of blood glucose.

Authors: 
Journal:  Diabetes Care       Date:  1987 Jan-Feb       Impact factor: 19.112

6.  Effect of ambient temperature and humidity on performance of blood glucose meters.

Authors:  J M King; C A Eigenmann; S Colagiuri
Journal:  Diabet Med       Date:  1995-04       Impact factor: 4.359

7.  Assessment of "Average of Normals" quality control procedures and guidelines for implementation.

Authors:  G S Cembrowski; E P Chandler; J O Westgard
Journal:  Am J Clin Pathol       Date:  1984-04       Impact factor: 2.493

8.  Thermal environment and subjective responses of patients and staff in a hospital during winter.

Authors:  Nobuko Hashiguchi; Megumi Hirakawa; Yutaka Tochihara; Yumi Kaji; Chitake Karaki
Journal:  J Physiol Anthropol Appl Human Sci       Date:  2005-01

9.  Outcome benefit of intensive insulin therapy in the critically ill: Insulin dose versus glycemic control.

Authors:  Greet Van den Berghe; Pieter J Wouters; Roger Bouillon; Frank Weekers; Charles Verwaest; Miet Schetz; Dirk Vlasselaers; Patrick Ferdinande; Peter Lauwers
Journal:  Crit Care Med       Date:  2003-02       Impact factor: 7.598

10.  Patient and physician analytic goals for self-monitoring blood glucose instruments.

Authors:  S L Weiss; G S Cembrowski; R S Mazze
Journal:  Am J Clin Pathol       Date:  1994-11       Impact factor: 2.493

View more
  3 in total

Review 1.  Technologies for continuous glucose monitoring: current problems and future promises.

Authors:  Santhisagar Vaddiraju; Diane J Burgess; Ioannis Tomazos; Faquir C Jain; Fotios Papadimitrakopoulos
Journal:  J Diabetes Sci Technol       Date:  2010-11-01

Review 2.  Geoenvironmental diabetology.

Authors:  Curtiss B Cook; Kay E Wellik; Margaret Fowke
Journal:  J Diabetes Sci Technol       Date:  2011-07-01

3.  The Effects of Temperature and Relative Humidity on Point-of-Care Glucose Measurements in Hospital Practice in a Tropical Clinical Setting.

Authors:  Busadee Pratumvinit; Nattakom Charoenkoop; Soamsiri Niwattisaiwong; Gerald J Kost; Panutsaya Tientadakul
Journal:  J Diabetes Sci Technol       Date:  2016-08-22