Influence of image metrics when assessing image quality from a test object in cardiac X-ray systems.

Eliseo Vano1, Carlos Ubeda, Bernhard Geiger, Luis C Martinez, Stephen Balter.   

Abstract

Modern fluoroscopic systems used for invasive cardiology typically acquire digital images in a 1,024 × 1,024 × 12 bits. These images are maintained in the original format while they remain on the imaging system itself. However, images are usually stored using a reduced 512 × 512 × 8-bits format. This paper presents a method for digital analysis of test objects images. The results obtained using image-intensifier and flat-detector systems are given for the original and reduced matrices. Images were acquired using a test object (TO) and a range of polymethyl methacrylate (PMMA) thicknesses from 4 to 28 cm. Adult patient protocols were evaluated for 16-28 cm of PMMA using the image-intensifier system. Pediatric protocols were evaluated for 4-16 cm of PMMA using the flat-detector system. The TO contains disks of various thicknesses to evaluate low contrast sensitivity and a bar pattern to evaluate high-contrast spatial resolution (HCSR). All available fluoroscopic and cine modes were evaluated. Entrance surface air kerma was also measured. Signal-to-noise ratio (SNR) was evaluated using regions of interest (ROI). HCSR was evaluated by comparing the statistical analysis of a ROI placed over the image of the bar pattern against a reference ROI. For both systems, an improvement of approximately 20% was observed for the SNR on the reduced matrices. However, the HCSR parameter was substantially lower in the reduced metrics. Cardiologists should consider the clinical influence of reduced spatial resolution when using the archived images.

Entities:  

Mesh:

Year:  2011        PMID: 20127268      PMCID: PMC3056969          DOI: 10.1007/s10278-009-9268-7

Source DB:  PubMed          Journal:  J Digit Imaging        ISSN: 0897-1889            Impact factor:   4.056


  10 in total

1.  American College of Cardiology/European Society of Cardiology International Study of Angiographic Data Compression Phase I: The effect of lossy data compression on recognition of diagnostic features in digital coronary angiography.

Authors:  R A Kerensky; J T Cusma; P Kubilis; R Simon; T M Bashore; J W Hirshfeld; D R Holmes; C J Pepine; S E Nissen
Journal:  J Am Coll Cardiol       Date:  2000-04       Impact factor: 24.094

2.  Introduction and background: the International Angiographic Compression study.

Authors:  S E Nissen; J W Hirshfeld; R Simon
Journal:  J Am Coll Cardiol       Date:  2000-04       Impact factor: 24.094

3.  American College of Cardiology/European Society of Cardiology International Study of Angiographic Data Compression Phase III: measurement of image quality differences at varying levels of data compression.

Authors:  R Brennecke; U Bürgel; R Simon; G Rippin; H P Fritsch; T Becker; S E Nissen
Journal:  J Am Coll Cardiol       Date:  2000-04       Impact factor: 24.094

4.  Patient doses from fluoroscopically guided cardiac procedures in pediatrics.

Authors:  L C Martinez; E Vano; F Gutierrez; C Rodriguez; R Gilarranz; M J Manzanas
Journal:  Phys Med Biol       Date:  2007-07-24       Impact factor: 3.609

5.  Signal detectability in digital radiography: spatial domain figures of merit.

Authors:  Robert M Gagne; Jonathan S Boswell; Kyle J Myers
Journal:  Med Phys       Date:  2003-08       Impact factor: 4.071

6.  Patient dosimetry for x rays used in medical imaging.

Authors: 
Journal:  J ICRU       Date:  2005-12

7.  Dynamic flat panel detector versus image intensifier in cardiac imaging: dose and image quality.

Authors:  E Vano; B Geiger; A Schreiner; C Back; J Beissel
Journal:  Phys Med Biol       Date:  2005-11-23       Impact factor: 3.609

8.  Impact of various compression rates on interpretation of digital coronary angiograms.

Authors:  S Silber; R Dörr; G Zindler; H Mühling; T Diebel
Journal:  Int J Cardiol       Date:  1997-07-25       Impact factor: 4.164

9.  Effective doses to patients from paediatric cardiac catheterization.

Authors:  J Rassow; A A Schmaltz; F Hentrich; C Streffer
Journal:  Br J Radiol       Date:  2000-02       Impact factor: 3.039

10.  Application of European protocol in the evaluation of contrast-to-noise ratio and mean glandular dose for two digital mammography systems.

Authors:  W E Muhogora; A Devetti; R Padovani; P Msaki; F Bonutti
Journal:  Radiat Prot Dosimetry       Date:  2008-02-18       Impact factor: 0.972

  10 in total
  2 in total

1.  Influence of image metrics when assessing image quality from a test object in cardiac X-ray systems: Part II.

Authors:  Roberto Sanchez; Eliseo Vano; Carlos Ubeda; Jose M Fernandez; Stephen Balter; Bart Hoornaert
Journal:  J Digit Imaging       Date:  2012-08       Impact factor: 4.056

2.  Biplane interventional pediatric system with cone-beam CT: dose and image quality characterization for the default protocols.

Authors:  Eva Corredoira; Eliseo Vañó; Luis Alejo; Carlos Ubeda; Federico Gutiérrez-Larraya; Julia Garayoa
Journal:  J Appl Clin Med Phys       Date:  2016-07-08       Impact factor: 2.102

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.