Validation of the Lloyds pharmacy BP11 oscillometric blood pressure monitor according to the international protocol of the European society of hypertension.

Andrew Coleman1, Stephen Steel, Annemarie deGreeff, Andrew Shennan.   

Abstract

BACKGROUND: The BP11 is currently one of the lowest-cost blood pressure (BP) monitors available in the UK and has sold approximately 1.5 million units since its launch in 2006. As access to home-use BP monitoring increases, it is important that such devices are validated so that consumers and clinicians can be confident of their accuracy.
METHODS: The device was validated using the European Society of Hypertension International Protocol. Data collection and analysis were carried out under the laboratory standard ISO17025. Adherence to the standard and protocol was audited on-site by the United Kingdom Accreditation Service. Nine sequential same-arm measurements were taken alternating between trained observers and the test device. Minor variations from the protocol are reported.
RESULTS: The device passed both phases of the European Society of Hypertension International Protocol. The mean (standard deviation) of the difference between the observers and the device measurements was -1.2 (6.4) mmHg for systolic and 2.1 (6.8) mmHg for diastolic pressures, respectively.
CONCLUSION: The BP11 BP monitor can be recommended for home and professional use in an adult population.

Entities:  

Mesh:

Year:  2010        PMID: 20110808     DOI: 10.1097/MBP.0b013e32833713cf

Source DB:  PubMed          Journal:  Blood Press Monit        ISSN: 1359-5237            Impact factor:   1.444


  3 in total

1.  Neuronal expression of the mitochondrial protein prohibitin confers profound neuroprotection in a mouse model of focal cerebral ischemia.

Authors:  Anja Kahl; Corey J Anderson; Liping Qian; Henning Voss; Giovanni Manfredi; Costantino Iadecola; Ping Zhou
Journal:  J Cereb Blood Flow Metab       Date:  2017-07-17       Impact factor: 6.200

2.  Accuracy of monitors used for blood pressure checks in English retail pharmacies: a cross-sectional observational study.

Authors:  James Hodgkinson; Constantinos Koshiaris; Una Martin; Jonathan Mant; Carl Heneghan; Fd Richard Hobbs; Richard J McManus
Journal:  Br J Gen Pract       Date:  2016-03-29       Impact factor: 5.386

3.  Arm circumference: its importance for dialysis patients in the obesity era.

Authors:  Tekin Akpolat; Coşkun Kaya; Cengiz Utaş; Turgay Arinsoy; Hülya Taşkapan; Emre Erdem; M Emin Yilmaz; Rezzan Ataman; Semra Bozfakioğlu; Çetin Özener; Ibrahim Karayaylali; Rümeyza Kazancioğlu; Taner Çamsari; Mahmut Yavuz; Fevzi Ersoy; Soner Duman; Kenan Ateş
Journal:  Int Urol Nephrol       Date:  2012-06-30       Impact factor: 2.370

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.