Aerodynamic characteristics of a dry powder inhaler at low inhalation flows using a mixing inlet with an Andersen Cascade Impactor.

Dinesh Kumar Nadarassan1, Khaled H Assi, Henry Chrystyn.   

Abstract

The aerodynamic characteristics of the dose emitted from a dry powder inhaler (DPI) are inhalation flow dependent but have not been determined for low flows. We have designed novel methodology to measure these at <28.3 lmin(-1). The original Andersen Cascade Impactor (ACI) designed for use at 60 lmin(-1) was adapted to include a mixing inlet (MIXINLET) which allows inhalation flows through the DPI from 5 to 60 lmin(-1). The mean fine particle dose (FPD) from a formoterol Turbuhaler using the MIXINLET method at 10, 20, 28.3, 40 and 60 lmin(-1) was 0.55, 1.39, 1.80, 2.88 and 5.86 microg and the mass median aerodynamic diameter (MMAD) was 6.6, 6.0, 5.4, 5.1 and 2.8 microm. Similarly, the FPD using the ACI method was 0.13, 0.69, 1.50, 2.48 and 5.42 microg and MMADs were 12.2, 7.4, 5.5, 4.8 and 2.7 microm. The accuracy of the original ACI <28.3 lmin(-1) is unknown. The ACI with the mixing inlet allows the determination of the in vitro dose emission properties of DPIs at flows <28.3 lmin(-1) whilst maintaining a constant flow through the ACI. This methodology, therefore, can help to focus attention to the lowest inhalation flow required for a DPI. Copyright 2010 Elsevier B.V. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20093182     DOI: 10.1016/j.ejps.2010.01.002

Source DB:  PubMed          Journal:  Eur J Pharm Sci        ISSN: 0928-0987            Impact factor:   4.384


  2 in total

1.  Investigation of dry powder inhaler (DPI) resistance and aerosol dispersion timing on emitted aerosol aerodynamic particle sizing by multistage cascade impactor when sampled volume is reduced from compendial value of 4 L.

Authors:  Hlack Mohammed; Jan Arp; Frank Chambers; Mark Copley; Volker Glaab; Mark Hammond; Derek Solomon; Kerry Bradford; Theresa Russell; Yvonne Sizer; Steven C Nichols; Daryl L Roberts; Christopher Shelton; Roland Greguletz; Jolyon P Mitchell
Journal:  AAPS PharmSciTech       Date:  2014-05-29       Impact factor: 3.246

2.  The comparison of fluid dynamics parameters in an Andersen cascade impactor equipped with and without a preseparator.

Authors:  Janwit Dechraksa; Tan Suwandecha; Kittinan Maliwan; Teerapol Srichana
Journal:  AAPS PharmSciTech       Date:  2014-04-24       Impact factor: 3.246

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.