Accuracy of two stereolithographic guide systems for computer-aided implant placement: a computed tomography-based clinical comparative study.

Volkan Arisan1, Z Cuneyt Karabuda, Tayfun Ozdemir.   

Abstract

BACKGROUND: Stereolithographic surgical guides provide significant benefits during the simultaneous placement of multiple implants with regard to the final prosthetic plan. However, deviation from the planning poses a significant risk. Deviations of implants that were placed by bone-, tooth-, and mucosa-supported stereolithographic surgical guides were examined in this study.
METHODS: After enrolling 54 eligible patients, 294 implants were planned on cone-beam computerized tomography CB(CT)-derived images. Sixty guides, both single- and multiple-type, were produced using two commercial systems. Mucosa-supported guides were fixed with osteosynthesis screws. Implants were inserted, and at the end of osseointegration period, a new CB(CT) scan was performed. Preoperative planning was merged with the new CB(CT) data to identify the deviations between the planned and placed implants for each support type and manufacturer. The Kruskal-Wallis and Mann-Whitney U tests were used for comparison (P <0.05).
RESULTS: There were no damage-related complications in any critical anatomy. Implants that were placed by bone-supported guides had the highest mean deviations (5.0 degrees +/- 1.66 degrees angular, and 1.70 +/- 0.52 mm and 1.99 +/- 0.64 mm for implant shoulder and tip, respectively), whereas the lowest deviations were measured in implants that were placed by mucosa-supported guides (2.9 degrees +/- 0.39 degrees angular, and 0.7 +/- 0.13 mm and 0.76 +/- 0.15 mm for implant shoulder and tip, respectively).
CONCLUSIONS: Computer-aided planning and manufacturing surgical guides in accordance with CB(CT) images may help clinicians place implants. Rigid screw fixation of a single guide incorporating metal sleeves and a special drill kit further minimizes deviations.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20059416     DOI: 10.1902/jop.2009.090348

Source DB:  PubMed          Journal:  J Periodontol        ISSN: 0022-3492            Impact factor:   6.993


  27 in total

1.  The influence of exposure parameters on jawbone model accuracy using cone beam CT and multislice CT.

Authors:  B Vandenberghe; S Luchsinger; J Hostens; E Dhoore; R Jacobs
Journal:  Dentomaxillofac Radiol       Date:  2012-01-26       Impact factor: 2.419

2.  CT dataset anisotropy management for oral implantology planning software.

Authors:  Tommaso Chiarelli; Evelina Lamma; Tommaso Sansoni
Journal:  Int J Comput Assist Radiol Surg       Date:  2012-06-17       Impact factor: 2.924

3.  Accuracy of different types of computer-aided design/computer-aided manufacturing surgical guides for dental implant placement.

Authors:  Wei Geng; Changying Liu; Yucheng Su; Jun Li; Yanmin Zhou
Journal:  Int J Clin Exp Med       Date:  2015-06-15

4.  Implant Bed Preparation with an Erbium, Chromium Doped Yttrium Scandium Gallium Garnet (Er,Cr: YSGG) Laser Using Stereolithographic Surgical Guide.

Authors:  Gülin Seymen; Zeynep Turgut; Gizem Berk; Ayşen Bodur
Journal:  J Lasers Med Sci       Date:  2013

5.  Evaluation of accuracy in implant site preparation performed in single- or multi-step drilling procedures.

Authors:  Nadine Marheineke; Uta Scherer; Martin Rücker; Constantin von See; Björn Rahlf; Nils-Claudius Gellrich; Marcus Stoetzer
Journal:  Clin Oral Investig       Date:  2017-12-17       Impact factor: 3.573

6.  3D accuracy of implant positions in template-guided implant placement as a function of the remaining teeth and the surgical procedure: a retrospective study.

Authors:  Sigmar Schnutenhaus; Cornelia Edelmann; Heike Rudolph; Jens Dreyhaupt; Ralph G Luthardt
Journal:  Clin Oral Investig       Date:  2018-01-22       Impact factor: 3.573

Review 7.  3D imaging, 3D printing and 3D virtual planning in endodontics.

Authors:  Pratik Shah; B S Chong
Journal:  Clin Oral Investig       Date:  2018-01-12       Impact factor: 3.573

Review 8.  Computerized implant-dentistry: Advances toward automation.

Authors:  Minkle Gulati; Vishal Anand; Sanjeev Kumar Salaria; Nikil Jain; Shilpi Gupta
Journal:  J Indian Soc Periodontol       Date:  2015 Jan-Feb

9.  An assessment of template-guided implant surgery in terms of accuracy and related factors.

Authors:  Jee-Ho Lee; Ji-Man Park; Soung-Min Kim; Myung-Joo Kim; Jong-Ho Lee; Myung-Jin Kim
Journal:  J Adv Prosthodont       Date:  2013-11-28       Impact factor: 1.904

10.  Computer-designed selective laser sintering surgical guide and immediate loading dental implants with definitive prosthesis in edentulous patient: A preliminary method.

Authors:  Giovanni Di Giacomo; Jorge Silva; Rodrigo Martines; Sergio Ajzen
Journal:  Eur J Dent       Date:  2014-01