Clinical evaluation of the Canon TX-10 noncontact tonometer in healthy eyes.

Carlos García-Resúa1, María Jesús Giráldez Fernández, Eva Yebra-Pimentel, Silvia García-Montero.   

Abstract

PURPOSE: To assess the accuracy and repeatability of intraocular pressure (IOP) measurements obtained with the Canon TX-10 noncontact tonometer (NCT), using the Goldmann applanation tonometer (GAT) as the gold standard.
METHODS: Seventy-three young healthy subjects were enrolled in the study. Central corneal thickness (CCT) was obtained with Orbscan Topography System and then repeated IOP measurements were obtained with both tonometers. Canon TX-10 NCT was performed first, followed by GAT. Another study sample of 32 young healthy subjects was used to test the repeatability with both tonometers. Two sets of data were acquired separated by 1 week. Plotting the difference between the methods against the mean was done to compare the tonometers and the intersession variability. The hypothesis of zero bias was examined by a paired t test. The 95% limits of agreement (LoA) were also calculated.
RESULTS: No statistically significant differences were found when comparing the Canon TX-10 NCT with GAT, displaying close level of agreement with GAT as seen by the 95% LoA (-4.78 mmHg to +4.00 mmHg). CCT ranged from 419 microm to 585 microm and no relationship was found between CCT and IOP measurements. The coefficients of repeatability were 3.70 mmHg and 3.41 mmHg for GAT and TX-10 tonometers, respectively.
CONCLUSIONS: There was a good agreement between Canon TX-10 and GAT in healthy subjects. From the results obtained here, CCT appears to not affect IOP measurements when CCT is thinner than 570 microm.

Entities:  

Mesh:

Year:  2010        PMID: 20037897     DOI: 10.1177/112067211002000326

Source DB:  PubMed          Journal:  Eur J Ophthalmol        ISSN: 1120-6721            Impact factor:   2.597


  9 in total

1.  Repeatability of intra-ocular pressure and central corneal thickness measurements provided by a non-contact method of tonometry and pachymetry.

Authors:  Domenico Schiano Lomoriello; Marco Lombardo; Laura Tranchina; Francesco Oddone; Sebastiano Serrao; Pietro Ducoli
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-10-28       Impact factor: 3.117

2.  Observation seasonal variation of intraocular pressure in young healthy volunteers.

Authors:  Na Liao; Yan-Qian Xie; Guang-Yun Mao; Fan-Jun Bao; Zhong Lin; Hui-Lyu Jiang; Yuan-Bo Liang
Journal:  Int J Ophthalmol       Date:  2022-01-18       Impact factor: 1.779

3.  Performance of the PT100 noncontact tonometer in healthy eyes.

Authors:  Turki M AlMubrad
Journal:  Clin Ophthalmol       Date:  2011-05-19

4.  Two-position measurement of intraocular pressure by PT100 noncontact tonometry in comparison with Goldmann tonometry.

Authors:  Kelechi C Ogbuehi; John C Chijuka; Uchechukwu L Osuagwu
Journal:  Clin Ophthalmol       Date:  2011-09-06

5.  Comparison of intraocular pressure as measured by three different non-contact tonometers and goldmann applanation tonometer for non-glaucomatous subjects.

Authors:  Seung Pil Bang; Chong Eun Lee; Yu Cheol Kim
Journal:  BMC Ophthalmol       Date:  2017-11-02       Impact factor: 2.209

6.  Comparative Assessment of The Goldmann Applanation and Noncontact Tonometers in Intraocular Pressure Measurements in a Sample of Glaucoma Patients in the Cape Coast Metropolis, Ghana.

Authors:  Samuel Kyei; Cynthia Pakyennu Gboglu; Michael Agyemang Kwarteng; Frank Assiamah
Journal:  Niger Med J       Date:  2020-12-19

7.  Teaming-up nurses with ophthalmologists to expand the reach of eye care in a middle-income country: Validation of health data acquisition by nursing staff in a telemedicine strategy.

Authors:  Cassia Garcia Moraes Pagano; Tais de Campos Moreira; Daniel Sganzerla; Ana Maria Frölich Matzenbacher; Amanda Gomes Faria; Lucas Matturro; Felipe Cezar Cabral; Dimitris Rucks Varvaki Rados; Anelise Decavata Szortyka; Maicon Falavigna; Maria Eulalia Vinadé Chagas; Erno Harzheim; Marcelo Gonçalves; Roberto Umpierre; Aline Lutz de Araujo
Journal:  PLoS One       Date:  2021-11-30       Impact factor: 3.240

8.  Changes of intraocular pressure and refractive status in children following cycloplegic refraction with 1% cyclopentolate and 1% tropicamide.

Authors:  Kuo-Chi Hung; Hsiu-Mei Huang; Pei-Wen Lin
Journal:  Taiwan J Ophthalmol       Date:  2015-08-24

9.  The synergistic effect of inflammation and metabolic syndrome on intraocular pressure: A cross-sectional study.

Authors:  I-Te Lee; Jun-Sing Wang; Chia-Po Fu; Chia-Jen Chang; Wen-Jane Lee; Shih-Yi Lin; Wayne Huey-Herng Sheu
Journal:  Medicine (Baltimore)       Date:  2017-09       Impact factor: 1.889

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.