Literature DB >> 20031837

Detection and elimination of microemboli related to cardiopulmonary bypass.

Robert C Groom1, Reed D Quinn, Paul Lennon, Desmond J Donegan, John H Braxton, Robert S Kramer, Paul W Weldner, Louis Russo, Seth D Blank, Angus A Christie, Andreas H Taenzer, Richard J Forest, Cantwell Clark, Janine Welch, Cathy S Ross, Gerald T O'Connor, Donald S Likosky.   

Abstract

BACKGROUND: Neurobehavioral impairment is a common complication of coronary bypass surgery. Cerebral microemboli during cardiopulmonary bypass (CPB) are a principal mechanism of cognitive injury. The aim of this work was to study the occurrence of cerebral embolism during CPB and to evaluate the effectiveness of evidence-based CPB circuit component and process changes on the exposure of the patient to emboli. METHODS AND
RESULTS: M-Mode Doppler was used to detect emboli in the inflow and outflow of cardiopulmonary circuit and in the right and left middle cerebral arteries. Doppler signals were merged into a single display to allow real-time associations between discrete clinical techniques and emboli detection. One hundred sixty-nine isolated coronary artery bypass grafting (CABG) patients were studied between 2002 and 2008. There was no statistical difference in median microemboli detected in the inflow of the CPB circuit, (Phase I, 931; Phase II, 1214; Phase III, 1253; Phase IV, 1125; F [3,158]=0.8, P=0.96). Significant changes occurred in median microemboli detected in the outflow of the CPB circuit across phases, (Phase I, 702; Phase II, 572; Phase III, 596; Phase IV, 85; F [3,157]=13.1, P<0.001). Significant changes also occurred in median microemboli detected in the brain across phases, (Phase I, 604; Phase II, 429; Phase III, 407; Phase IV, 138; F [3,153]=14.4, P<0.001). Changes in the cardiopulmonary bypass circuit were associated with an 87.9% (702 versus 85) reduction in median microemboli in the outflow of the CPB circuit (P<0.001), and a 77.2% (604 versus 146) reduction in microemboli in the brain (P<0.001).
CONCLUSIONS: Changes in CPB techniques and circuit components, including filter size and type of pump, resulted in a reduction in more than 75% of cerebral microemboli.

Entities:  

Mesh:

Year:  2009        PMID: 20031837     DOI: 10.1161/CIRCOUTCOMES.108.803163

Source DB:  PubMed          Journal:  Circ Cardiovasc Qual Outcomes        ISSN: 1941-7713


  18 in total

1.  Microbubble generation and transmission of medtronic's affinity hardshell venous reservoir and collapsible venous reservoir bag: an in-vitro comparison.

Authors:  Kieron C Potger; Darryl McMillan; Mark Ambrose
Journal:  J Extra Corpor Technol       Date:  2011-09

2.  In vitro evaluation of gaseous microemboli handling of cardiopulmonary bypass circuits with and without integrated arterial line filters.

Authors:  Saifei Liu; Richard F Newland; Phillip J Tully; Sigrid C Tuble; Robert A Baker
Journal:  J Extra Corpor Technol       Date:  2011-09

3.  Vacuum-Assisted Venous Drainage: A 2014 Safety Survey.

Authors:  Rachel Gambino; Bruce Searles; Edward M Darling
Journal:  J Extra Corpor Technol       Date:  2015-09

4.  Vacuum assist: angel or demon CON.

Authors:  Timothy W Willcox
Journal:  J Extra Corpor Technol       Date:  2013-06

Review 5.  Brain protection during cardiac surgery: circa 2012.

Authors:  John W Hammon
Journal:  J Extra Corpor Technol       Date:  2013-06

6.  Microbubble transmission during cardiotomy infusion of a hardshell venous reservoir with integrated cardiotomy versus a softshell venous reservoir with separated cardiotomy: an in vitro comparison.

Authors:  Kieron C Potger; Darryl McMillan; Mark Ambrose
Journal:  J Extra Corpor Technol       Date:  2013-06

7.  Oxygen or nitrogen: which is the lesser of two evils?

Authors:  Gary Grist
Journal:  J Extra Corpor Technol       Date:  2013-03

8.  Collaborative Quality Improvement Reduces Postoperative Pneumonia After Isolated Coronary Artery Bypass Grafting Surgery.

Authors:  Donald S Likosky; Steven D Harrington; Lourdes Cabrera; Alphonse DeLucia; Carol E Chenoweth; Sarah L Krein; Dylan Thibault; Min Zhang; Roland A Matsouaka; Raymond J Strobel; Richard L Prager
Journal:  Circ Cardiovasc Qual Outcomes       Date:  2018-11

Review 9.  Clinical microsystems: a critical framework for crossing the quality chasm.

Authors:  Donald S Likosky
Journal:  J Extra Corpor Technol       Date:  2014-03

Review 10.  Hyperoxia: a review of the risks and benefits in adult cardiac surgery.

Authors:  Robert W Young
Journal:  J Extra Corpor Technol       Date:  2012-12
View more