Pulsatility index variations using two different transit-time flowmeters in coronary artery bypass surgery.

Håvard B Nordgaard1, Nicola Vitale, Rafael Astudillo, Attilio Renzulli, Pål Romundstad, Rune Haaverstad.   

Abstract

OBJECTIVE: Transit-time flow measurement is widely accepted as an intra-operative assessment in coronary artery bypass grafting (CABG). However, the two most commonly applied flowmeters, manufactured by MediStim ASA and Transonic Inc., have different default filter settings of 20 and 10 Hz, respectively. This may cause different flow measurements, which will influence the reported results. The aim was to compare pulsatility index (PI) values recorded by the MediStim and Transonic flowmeters in two different clinical settings: (1) analysis of the flow patterns recorded simultaneously by both flowmeters in the same CABGs; and (2) evaluation of flow patterns under different levels of filter settings in the same grafts.
METHODS: Graft flow and PI were measured using the two different flowmeters simultaneously in 19 bypass grafts. Finally, eight grafts were assessed under different digital filter settings at 5, 10, 20, 30, 50 and 100 Hz.
RESULTS: The Transonic flowmeter provided substantially lower PI as compared with the MediStim flowmeter. By increasing the filter setting in the flowmeter, PI increased considerably.
CONCLUSIONS: The Transonic flowmeter displayed a lower PI than the MediStim, due to a lower filter setting. In the Transonic,flow signals are filtered at a lower level, rendering a 'smoother' pattern of flow curves. Because different filter settings determine different PIs, caution must be taken when flow values and flowmeters are compared. The type of flowmeter should be indicated whenever graft flow measurements and derived indexes are provided [corrected]. Copyright 2009 European Association for Cardio-Thoracic Surgery. All rights reserved.

Entities:  

Mesh:

Year:  2009        PMID: 20031439     DOI: 10.1016/j.ejcts.2009.11.030

Source DB:  PubMed          Journal:  Eur J Cardiothorac Surg        ISSN: 1010-7940            Impact factor:   4.191


  4 in total

1.  The application of intraoperative transit time flow measurement to accurately assess anastomotic quality in sequential vein grafting.

Authors:  Yang Yu; Fan Zhang; Ming-Xin Gao; Hai-Tao Li; Jing-Xing Li; Wei Song; Xin-Sheng Huang; Cheng-Xiong Gu
Journal:  Interact Cardiovasc Thorac Surg       Date:  2013-09-01

2.  Transit Time Flowmetry in Coronary Artery Bypass Grafting-experience at Queen Alia Heart Institute, Jordan.

Authors:  Basel Harahsheh
Journal:  Oman Med J       Date:  2012-11

Review 3.  Physiological insights of recent clinical diagnostic and therapeutic technologies for cardiovascular diseases.

Authors:  Kenji Shigemi; Soichiro Fuke; Dai Une; Keita Saku; Shuji Shimizu; Toru Kawada; Toshiaki Shishido; Kenji Sunagawa; Masaru Sugimachi
Journal:  J Physiol Sci       Date:  2017-07-05       Impact factor: 2.781

4.  A Single-opening&closing Valve Tester for Direct Measurement of Closing Volume of the Heart Valve.

Authors:  Hao Wang; Zhan Cui; Zhongxi Zhou; Zhaoming He
Journal:  Cardiovasc Eng Technol       Date:  2021-06-25       Impact factor: 2.495

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.