Literature DB >> 24000314

The application of intraoperative transit time flow measurement to accurately assess anastomotic quality in sequential vein grafting.

Yang Yu1, Fan Zhang, Ming-Xin Gao, Hai-Tao Li, Jing-Xing Li, Wei Song, Xin-Sheng Huang, Cheng-Xiong Gu.   

Abstract

OBJECTIVES: Intraoperative transit time flow measurement (TTFM) is widely used to assess anastomotic quality in coronary artery bypass grafting (CABG). However, in sequential vein grafting, the flow characteristics collected by the conventional TTFM method are usually associated with total graft flow and might not accurately indicate the quality of every distal anastomosis in a sequential graft. The purpose of our study was to examine a new TTFM method that could assess the quality of each distal anastomosis in a sequential graft more reliably than the conventional TTFM approach.
METHODS: Two TTFM methods were tested in 84 patients who underwent sequential saphenous off-pump CABG in Beijing An Zhen Hospital between April and August 2012. In the conventional TTFM method, normal blood flow in the sequential graft was maintained during the measurement, and the flow probe was placed a few centimetres above the anastomosis to be evaluated. In the new method, blood flow in the sequential graft was temporarily reduced during the measurement by placing an atraumatic bulldog clamp at the graft a few centimetres distal to the anastomosis to be evaluated, while the position of the flow probe remained the same as in the conventional method. This new TTFM method was named the flow reduction TTFM. Graft flow parameters measured by both methods were compared.
RESULTS: Compared with the conventional TTFM, the flow reduction TTFM resulted in significantly lower mean graft blood flow (P < 0.05); in contrast, yielded significantly higher pulsatility index (P < 0.05). Diastolic filling was not significantly different between the two methods and was >50% in both cases. Interestingly, the flow reduction TTFM identified two defective middle distal anastomoses that the conventional TTFM failed to detect. Graft flows near the defective distal anastomoses were improved substantially after revision.
CONCLUSIONS: In this study, we found that temporary reduction of graft flow during TTFM seemed to enhance the sensitivity of TTFM to less-than-critical anastomotic defects in a sequential graft and to improve the overall accuracy of the intraoperative assessment of anastomotic quality in sequential vein grafting.

Entities:  

Keywords:  Anastomotic quality; Sequential saphenous coronary artery bypass grafting; Transit time flow measurement

Mesh:

Year:  2013        PMID: 24000314      PMCID: PMC3829511          DOI: 10.1093/icvts/ivt398

Source DB:  PubMed          Journal:  Interact Cardiovasc Thorac Surg        ISSN: 1569-9285


  25 in total

1.  Routine off-pump coronary artery bypass grafting is safe and feasible in high-risk patients with left main disease.

Authors:  Maximilian Y Emmert; Sacha P Salzberg; Burkhardt Seifert; Ulrich P Schurr; Simon P Hoerstrup; Oliver Reuthebuch; Michele Genoni
Journal:  Ann Thorac Surg       Date:  2010-04       Impact factor: 4.330

2.  Predictive value of intraoperative transit-time flow measurement for short-term graft patency in coronary surgery.

Authors:  Gabriele Di Giammarco; Marco Pano; Sergio Cirmeni; Piero Pelini; Giuseppe Vitolla; Michele Di Mauro
Journal:  J Thorac Cardiovasc Surg       Date:  2006-09       Impact factor: 5.209

3.  Right coronary system grafts: alone or together with left system grafts--angiographic results.

Authors:  Ozer Kandemir; Hilmi Tokmakoglu; Tevfik Tezcaner; Cem Yorgancioglu; Kaya Süzer; Yaman Zorlutuna
Journal:  Ann Thorac Cardiovasc Surg       Date:  2007-02       Impact factor: 1.520

4.  Single versus sequential saphenous vein grafting of the circumflex system: a flowmetric study.

Authors:  Francesco Onorati; Francesco Pezzo; Antonio Esposito; Barbara Impiombato; Maria Caterina Comi; Massimo Polistina; Attilio Renzulli
Journal:  Scand Cardiovasc J       Date:  2007-08       Impact factor: 1.589

5.  Neural network pattern recognition analysis of graft flow characteristics improves intra-operative anastomotic error detection in minimally invasive CABG.

Authors:  P B Cerrito; S C Koenig; D J VanHimbergen; S F Jaber; D L Ewert; P A Spence
Journal:  Eur J Cardiothorac Surg       Date:  1999-07       Impact factor: 4.191

6.  Evaluation of early coronary graft patency after coronary artery bypass graft surgery using multislice computed tomography angiography.

Authors:  Hosseinali Bassri; Farzad Salari; Fereidoun Noohi; Marzieh Motevali; Seifollah Abdi; Nader Givtaj; Kamal Raissi; Majid Haghjoo
Journal:  BMC Cardiovasc Disord       Date:  2009-12-08       Impact factor: 2.298

7.  Multi-detector computed tomography in coronary artery bypass graft assessment: a meta-analysis.

Authors:  Catherine M Jones; Thanos Athanasiou; Nicola Dunne; Joanne Kirby; Omer Aziz; Ahmed Haq; Christopher Rao; Vasilis Constantinides; Sanjay Purkayastha; Ara Darzi
Journal:  Ann Thorac Surg       Date:  2007-01       Impact factor: 4.330

8.  Flow in coronary bypass conduits on-pump and off-pump.

Authors:  Henrik K Kjaergard; Akhmad Irmukhamedov; John B Christensen; Thomas A Schmidt
Journal:  Ann Thorac Surg       Date:  2004-12       Impact factor: 4.330

9.  Transit-time blood flow measurements in sequential saphenous coronary artery bypass grafts.

Authors:  Håvard Nordgaard; Nicola Vitale; Rune Haaverstad
Journal:  Ann Thorac Surg       Date:  2009-05       Impact factor: 4.330

10.  Non-invasive assessment of coronary artery bypass graft patency using 16-slice computed tomography angiography.

Authors:  Emma S Houslay; Tristan Lawton; Anshuman Sengupta; Neal G Uren; Graham McKillop; David E Newby
Journal:  J Cardiothorac Surg       Date:  2007-06-05       Impact factor: 1.637

View more
  1 in total

Review 1.  Intraoperative graft assessment during coronary artery bypass surgery.

Authors:  Toshihiro Fukui
Journal:  Gen Thorac Cardiovasc Surg       Date:  2015-01-06
  1 in total