Kai Sun1, Nobuko Suzuki, Zheyu Li, Ryoko Araki, Kosei Ueno, Saulius Juodkazis, Masumi Abe, Sumihare Noji, Hiroaki Misawa.
Abstract
For the fractionation of fragments of interest from selective PCR products generated by high coverage gene expression profiling (HiCEP) analysis, high-resolution with the ability to discriminate and fractionate fragments differing by one base (base pair) in size is highly required. We report here on a new 4-inch diameter spiral-channel chip device for automatic high-fidelity fractionation. Overlapping DNA fragments of 180, 181 and 182 bases, with only one-base difference in size, were successfully fractionated. The collected fragments were PCR amplified, and then evaluated by size checking analysis, DNA sequencing, and homolog search. The high-resolution fractionation has been achieved because of the combined contributions of (i) the high-resolution separation using a 30 cm long spiral channel, (ii) a blocking technique to avoid contamination from unselected fragments during CE, and (iii) precise micro-scale target extraction. Contaminations due to unselected fractions have been greatly decreased to a negligible level by optimization of the extraction position and extraction time corresponding to the targeted segment only. This technique can be adapted to a wide range of applications, such as protein or cell collections where requirements for the high purity are more important than the amount of recovered fractionated material.Mesh:
Substances:
Year: 2009 PMID: 20013913 DOI: 10.1002/elps.200900455
Source DB: PubMed Journal: Electrophoresis ISSN: 0173-0835 Impact factor: 3.535