Abstract

OBJECTIVE: Both the Medical Technology and Innovations (MTI) and plusoptiX photoscreeners are used to objectively screen for amblyogenic risk factors in children. The MTI has been extensively studied, but the limited availability of film may render it obsolete. We compared the MTI with the plusoptiX, a newer digital photoscreener, for the ability to detect amblyogenic factors when compared with a comprehensive pediatric ophthalmic examination. We believe our results will help to guide community-based vision screening programs.
METHODS: One hundred fifty-one children were examined consecutively in our office. Each patient was screened with the MTI and plusoptiX devices on the same day as part of a comprehensive pediatric ophthalmic examination. Results via MTI were evaluated by an expert masked examiner (R.W.A.), and the plusoptiX results were interpreted by the incorporated software.
RESULTS: Sixty-five percent of patients were found to have amblyopia or amblyogenic risk factors during the pediatric ophthalmic examination conducted via the American Association of Pediatric Ophthalmology and Strabismus referral criteria. We found the MTI photoscreener to have a sensitivity of 83.6%, specificity of 90.5%, false- positive rate of 9.4%, false-negative rate of 16.3%, and positive predictive value of 94.2%. The plusoptiX demonstrated a sensitivity of 98.9%, specificity of 96.1%, false- positive rate of 3.7%, false-negative rate of 1.0%, and positive predictive value of 97.9%.
CONCLUSION: The MTI and plusoptiX photoscreeners proved to be effective when compared with a comprehensive cycloplegic pediatric ophthalmic examination. The plusoptiX, however, was found to have a higher sensitivity and specificity than the MTI.

Entities:  

Mesh:

Year:  2009        PMID: 20008713     DOI: 10.1001/archophthalmol.2009.294

Source DB:  PubMed          Journal:  Arch Ophthalmol        ISSN: 0003-9950


  10 in total

1.  Performance of Plusoptix A09 Photo Screener in Refractive Error Screening in School Children Aged between 5 and 15 Years in the Southern Part of India.

Authors:  Avinash V Prabhu; Jyothi Thomas; Ramesh S Ve; Sayantan Biswas
Journal:  J Curr Ophthalmol       Date:  2020-07-04

2.  A comparison of the PlusoptiX S04 and A09 photoscreeners.

Authors:  Eric Singman; Noelle Matta; Jing Tian; Abby Brubaker; David Silbert
Journal:  Strabismus       Date:  2013-06

3.  Performance of the Plusoptix A09 photoscreener in detecting amblyopia risk factors in Chinese children attending an eye clinic.

Authors:  Xiao-Ran Yan; Wan-Zhen Jiao; Zhi-Wei Li; Wen-Wen Xu; Feng-Jiao Li; Li-Hua Wang
Journal:  PLoS One       Date:  2015-06-01       Impact factor: 3.240

4.  Accuracy of PlusOptix A09 distance refraction in pediatric myopia and hyperopia.

Authors:  Arnaud Payerols; Claudie Eliaou; Véronique Trezeguet; Max Villain; Vincent Daien
Journal:  BMC Ophthalmol       Date:  2016-06-01       Impact factor: 2.209

5.  Sensitivity and Specificity of Preschool Vision Screening in Iran.

Authors:  Hassan Hashemi; Abbasali Yekta; Ebrahim Jafarzadehpur; Hadi Ostadimoghaddam; Amir Asharlous; Payam Nabovati; Mehdi Khabazkhoob
Journal:  Iran J Public Health       Date:  2017-02       Impact factor: 1.429

6.  Evaluation of a smartphone photoscreening app to detect refractive amblyopia risk factors in children aged 1-6 years.

Authors:  Robert W Arnold; James W O'Neil; Kim L Cooper; David I Silbert; Sean P Donahue
Journal:  Clin Ophthalmol       Date:  2018-08-23

Review 7.  Scope and costs of autorefraction and photoscreening for childhood amblyopia-a systematic narrative review in relation to the EUSCREEN project data.

Authors:  Anna M Horwood; Helen J Griffiths; Jill Carlton; Paolo Mazzone; Arinder Channa; Mandy Nordmann; Huibert J Simonsz
Journal:  Eye (Lond)       Date:  2020-11-30       Impact factor: 3.775

Review 8.  A Systematic Review of Current Teleophthalmology Services in New Zealand Compared to the Four Comparable Countries of the United Kingdom, Australia, United States of America (USA) and Canada.

Authors:  Liam Walsh; Sheng Chiong Hong; Renoh Johnson Chalakkal; Kelechi C Ogbuehi
Journal:  Clin Ophthalmol       Date:  2021-10-04

9.  Comparative Validation of PlusoptiX and AI-Optic Photoscreeners in Children with High Amblyopia Risk Factor Prevalence.

Authors:  Robert W Arnold
Journal:  Clin Ophthalmol       Date:  2022-08-16

10.  Pediatric vision screening using the plusoptiX A12C photoscreener in Chinese preschool children aged 3 to 4 years.

Authors:  Dan Huang; Xuejuan Chen; Xiaohan Zhang; Yue Wang; Hui Zhu; Hui Ding; Jing Bai; Ji Chen; Zhujun Fu; Zijin Wang; Hu Liu
Journal:  Sci Rep       Date:  2017-05-17       Impact factor: 4.379