Accuracy of methemoglobin detection by pulse CO-oximetry during hypoxia.

John R Feiner1, Philip E Bickler, Paul D Mannheimer.   

Abstract

BACKGROUND: Methemoglobin in the blood cannot be detected by conventional pulse oximetry, although it can bias the oximeter's estimate (Spo2) of the true arterial functional oxygen saturation (Sao2). A recently introduced "Pulse CO-Oximeter" (Masimo Rainbow SET(R) Radical-7 Pulse CO-Oximeter, Masimo Corp., Irvine, CA) is intended to additionally monitor noninvasively the fractional carboxyhemoglobin and methemoglobin content in blood. The purpose of our study was to determine whether hypoxia affects the new device's estimated methemoglobin reading accuracy, and whether the presence of methemoglobin impairs the ability of the Radical-7 and a conventional pulse oximeter (Nonin 9700, Nonin Medical Inc., Plymouth, MN) to detect decreases in Sao2.
METHODS: Eight and 6 healthy adults were included in 2 study groups, respectively, each fitted with multiple sensors and a radial arterial catheter for blood sampling. In the first group, IV administration of approximately 300 mg sodium nitrite increased subjects' methemoglobin level to a 7% to 8% target and hypoxia was induced to different levels of Sao2 (70%-100%) by varying fractional inspired oxygen. In the second group, 15% methemoglobin at room air and 80% Sao2 were targeted. Pulse CO-oximeter readings were compared with arterial blood values measured using a Radiometer multiwavelength hemoximeter. Pulse CO-oximeter methemoglobin reading performance was analyzed by observing the incidence of meaningful reading errors at the various hypoxia levels. This was used to determine the impact on predictive values for detecting methemoglobinemia. Spo2 reading bias, precision, and root mean square error were evaluated during conditions of elevated methemoglobin.
RESULTS: Observations spanned 66.2% to 99% Sao2 and 0.6% to 14.4% methemoglobin over the 2 groups (170 blood draws). Masimo methemoglobin reading bias and precision over the full Sao2 span was 7.7% +/- 13.0%. Best accuracy was found in the 95% to 100% Sao2 range (1.9% +/- 2.5%), progressing to its worst in the 70% to 80% range (24.8% +/- 15.6%). Occurrence of methemoglobin readings in error >5% increased over each 5-point decrease in Sao2 (P < 0.05). Masimo Spo2 readings were biased -6.3% +/- 3.0% in the 95% to 100% Sao2 range with 4% to 8.3% methemoglobin. Both the Radical-7 and Nonin 9700 pulse oximeters accurately detected decreases in Sao(2) <90% with 4% to 15% methemoglobin, despite displaying low Spo2 readings when Sao2 was >95%.
CONCLUSIONS: The Radical-7's methemoglobin readings become progressively more inaccurate as Sao2 decreases <95%, at times overestimating true values by 10% to 40%. Elevated methemoglobin causes the Spo2 readings to underestimate Sao2 similar to conventional 2-wavelength pulse oximeters at high saturation. Spo2 readings from both types of instruments continue to trend downward during the development of hypoxemia (Sao2 <90%) with methemoglobin levels up to 15%.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 20007731     DOI: 10.1213/ANE.0b013e3181c91bb6

Source DB:  PubMed          Journal:  Anesth Analg        ISSN: 0003-2999            Impact factor:   5.108


  13 in total

1.  Minocycline toxicity: case files of the University of Massachusetts medical toxicology fellowship.

Authors:  Matthew D Zuckerman; Katherine L Boyle; Christopher D Rosenbaum
Journal:  J Med Toxicol       Date:  2012-09

2.  Evaluation of multiwave pulse total-hemoglobinometer during general anesthesia.

Authors:  Daisuke Toyoda; Rie Yasumura; Mitsue Fukuda; Ryoichi Ochiai; Yoshifumi Kotake
Journal:  J Anesth       Date:  2013-10-22       Impact factor: 2.078

3.  Case files of the medical toxicology fellowship at Banner Good Samaritan Medical Center in Phoenix, AZ: methemoglobinemia following dapsone exposure.

Authors:  Joshua Canning; Michael Levine
Journal:  J Med Toxicol       Date:  2011-06

4.  Methemoglobinemia and arterial blood gas.

Authors:  Sim Sai Tin; Viroj Wiwanitkit
Journal:  Indian Heart J       Date:  2015-02-25

5.  Congenital methaemoglobinaemia diagnosed in an adolescent boy.

Authors:  Amal Paul; Sujith Thomas Chacko
Journal:  BMJ Case Rep       Date:  2019-03-31

6.  The Mysterious Story of Brown Blood and High Saturation Gap.

Authors:  Hans Raj Pahadiya; Manoj Lakhotia; Ronak Gandhi; Gopal Raj Prajapati
Journal:  Indian J Hematol Blood Transfus       Date:  2016-05-27       Impact factor: 0.900

7.  Methemoglobin levels in generally anesthetized pediatric dental patients receiving prilocaine versus lidocaine.

Authors:  Lauren L Gutenberg; Jung-Wei Chen; Larry Trapp
Journal:  Anesth Prog       Date:  2013

Review 8.  Recommendations for diagnosis and treatment of methemoglobinemia.

Authors:  Achille Iolascon; Paola Bianchi; Immacolata Andolfo; Roberta Russo; Wilma Barcellini; Elisa Fermo; Gergely Toldi; Stefano Ghirardello; Davis Rees; Richard Van Wijk; Antonis Kattamis; Patrick G Gallagher; Noemi Roy; Ali Taher; Razan Mohty; Andreas Kulozik; Lucia De Franceschi; Antonella Gambale; Mariane De Montalembert; Gian Luca Forni; Cornelis L Harteveld; Josef Prchal
Journal:  Am J Hematol       Date:  2021-09-23       Impact factor: 13.265

9.  Accuracy of carboxyhemoglobin detection by pulse CO-oximetry during hypoxemia.

Authors:  John R Feiner; Mark D Rollins; Jeffrey W Sall; Helge Eilers; Paul Au; Philip E Bickler
Journal:  Anesth Analg       Date:  2013-03-11       Impact factor: 5.108

10.  Methaemoglobin and COHb in patients with malaria.

Authors:  Thomas Hänscheid; Tom Gresnigt; Sascha Löhr; Arnaud Flamen; Thomas Zoller; José Melo-Cristino; Martin P Grobusch
Journal:  Malar J       Date:  2014-07-23       Impact factor: 2.979

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.