In vitro evaluation of relative perforating potential of central venous catheters: comparison of materials, selected models, number of lumens, and angles of incidence to simulated membrane.

N Gravenstein1, R H Blackshear.   

Abstract

Perforation of the vena cava or atrium is a serious complication of monitoring with a central venous catheter. We designed an in vitro model with a pulsating simulated membrane to evaluate a number of variables that could affect relative perforating potential of different types of central venous catheters. To determine the perforating potential of central venous catheters, we studied the effects of (1) the angle of incidence (n = 6) between catheter and simulated membrane; (2) catheter material (polyurethane and polyethylene); (3) make (manufacturer and model) (n = 6), with 3 catheters of each make tested; (4) design (n = 3 each: silicone rubber, open-ended, blunt-ended, and polyurethane pigtail); and (5) number of lumens (single, double, or triple). Each trial was repeated five times with each catheter that was tested. Perforation was significantly more likely when the angle of incidence between catheter and pulsating simulated membrane was greater than 40 degrees than when it was 40 degrees (P less than 0.05). Perforation was less likely with single-lumen than comparable French-sized double- and triple-lumen catheters; among single-lumen catheters, perforation required many more pulsations with a polyurethane than a polyethylene catheter (P less than 0.001). Perforation potential differed significantly among 6 makes of 7-French triple-lumen catheters (P less than 0.05). Compared with other materials or designs, silicone rubber or a pigtail tip decreased the perforation potential of catheters (P less than 0.001). These data offer additional objective information to consider when choosing and positioning central venous catheters.

Entities:  

Mesh:

Substances:

Year:  1991        PMID: 1999693     DOI: 10.1007/bf01617891

Source DB:  PubMed          Journal:  J Clin Monit        ISSN: 0748-1977


  22 in total

1.  Cardiac tamponade from central venous catheters.

Authors:  R J Defalque; C Campbell
Journal:  Anesthesiology       Date:  1979-03       Impact factor: 7.892

2.  The aberrant central venous catheter and its complications.

Authors:  C S Langston
Journal:  Radiology       Date:  1971-07       Impact factor: 11.105

3.  Appraisal of the dislocation of central venous catheter tips using subclavian and arm veins.

Authors:  B van Dijk; P M Bakker
Journal:  Anaesthesist       Date:  1977-03       Impact factor: 1.041

4.  Hydrothorax as a late complication of central venous indwelling catheters.

Authors:  T J Iberti; L B Katz; M A Reiner; T Brownie; K B Kwun
Journal:  Surgery       Date:  1983-11       Impact factor: 3.982

5.  Cardiac tamponade from central venous catheterization: two cases in premature infants with survival.

Authors:  J C Opitz; W Toyama
Journal:  Pediatrics       Date:  1982-07       Impact factor: 7.124

6.  Cardiac tamponade associated with a multilumen central venous catheter.

Authors:  S P Maschke; H J Rogove
Journal:  Crit Care Med       Date:  1984-07       Impact factor: 7.598

Review 7.  Cardiac tamponade from central venous catheters.

Authors:  H Edwards; T C King
Journal:  Arch Surg       Date:  1982-07

8.  Material thrombogenicity in central venous catheterization III. A comparison between soft polyvinylchloride and soft polyurethane elastomer, long, antebrachial catheters.

Authors:  I Curelaru; B Gustavsson; E Hultman; E Jondmundsson; L E Linder; T Stefánsson; O Stenqvist
Journal:  Acta Anaesthesiol Scand       Date:  1984-04       Impact factor: 2.105

9.  Stiffness of central venous catheters.

Authors:  O Stenqvist; I Curelaru; L E Linder; B Gustavsson
Journal:  Acta Anaesthesiol Scand       Date:  1983-04       Impact factor: 2.105

10.  Catheter-induced lesions of the right side of the heart. A one-year prospective study of 141 autopsies.

Authors:  B S Ducatman; J C McMichan; W D Edwards
Journal:  JAMA       Date:  1985-02-08       Impact factor: 56.272

View more
  17 in total

1.  A better landmark for positioning a central venous catheter.

Authors:  Lawrence J Caruso; Nikolaus Gravenstein; A Joseph Layon; Keith Peters; Andrea Gabrielli
Journal:  J Clin Monit Comput       Date:  2002-08       Impact factor: 2.502

2.  Variables decreasing tip movement of peripherally inserted central catheters in pediatric patients.

Authors:  Ralph Gnannt; Bairbre L Connolly; Dimitri A Parra; Joao Amaral; Rahim Moineddin; Avnesh S Thakor
Journal:  Pediatr Radiol       Date:  2016-06-07

3.  [Anaesthesia procedures and invasive vascular access in severely injured patients at trauma room admission in Germany : An online survey].

Authors:  M F Struck; P Hilbert-Carius; B Hossfeld; J Hinkelbein; M Bernhard; T Wurmb
Journal:  Anaesthesist       Date:  2017-01-11       Impact factor: 1.041

4.  The pericardial reflection and the tip of the central venous catheter - topographical analysis in stillborn babies.

Authors:  Frank Eifinger; Anne Vierzig; Bernhard Roth; Martin Scaal; Friederike Koerber
Journal:  Pediatr Radiol       Date:  2016-06-27

5.  [Modified ECG-guidance for optimal central venous catheter tip positioning. A transesophageal echocardiography controlled study].

Authors:  W Schummer; C Schummer; C Schelenz; P Schmidt; R Fröber; E Hüttemann
Journal:  Anaesthesist       Date:  2005-10       Impact factor: 1.041

6.  Fatal cardiac perforation caused by the dilator of a central venous catheterization kit.

Authors:  M R Mitchell; D G Wood; M Naraghi; J M Riopelle
Journal:  J Clin Monit       Date:  1993-09

7.  Venobronchial fistula: an unusual early complication after central venous catheterization.

Authors:  V Guariento; G B Luciani; B Dal Corso; A Carletti; A Mazzucco; A Luzzani
Journal:  Intensive Care Med       Date:  1997-11       Impact factor: 17.440

8.  Central venous catheter-associated pericardial tamponade in a 6-day old: a case report.

Authors:  Swati O Arya; Gurumurthy M Hiremath; Kingsley C Okonkwo; Michael D Pettersen
Journal:  Int J Pediatr       Date:  2010-02-09

9.  Central venous catheter tip migration due to tracheal extubation: a prospective randomized study.

Authors:  Manuel F Struck; Theresa Jünemann; Konrad Reinhart; Wolfram Schummer
Journal:  J Clin Monit Comput       Date:  2016-07-28       Impact factor: 2.502

10.  Perforations associated with peripherally inserted central catheters in a neonatal population.

Authors:  Andrew J Sertic; Bairbre L Connolly; Michael J Temple; Dimitri A Parra; Joao G Amaral; Kyong-Soon Lee
Journal:  Pediatr Radiol       Date:  2017-10-06
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.