| Literature DB >> 19936754 |
Frank M Zijta1, Shandra Bipat, Jaap Stoker.
Abstract
OBJECTIVE: To determine the diagnostic accuracy of MR-colonography for the detection of colorectal lesions.Entities:
Mesh:
Substances:
Year: 2009 PMID: 19936754 PMCID: PMC2850516 DOI: 10.1007/s00330-009-1663-4
Source DB: PubMed Journal: Eur Radiol ISSN: 0938-7994 Impact factor: 5.315
Fig. 1Flow chart indicating selection of articles included for analysis (and potentially relevant studies that were excluded by reviewers [13–36])
Quality assessment of included studies using relevant items of the QUADAS tool
| Study | Representative patient population | Selection criteria specified | Accurate verification | Time interval specifieda | Complete sample size verificationb | Reproducible description index test | Blinded interpretation index test | Blinded interpretation reference test | Intermediate index test results reported | Description of withdrawals |
|---|---|---|---|---|---|---|---|---|---|---|
| Achiam (2008) [ | Yes | No | Yes | Unclear | No | Yes | Yes | Noc | Yes | Yes |
| Florie (2007) [ | Yes | Yes | Yes | Yes | Yes | Yes | Yes | Yesd | Yes | Yes |
| Goehde (2005) [ | Yes | Yes | Yes | Yes | Yes | Yes | Yes | Yes | Yes | Yes |
| Hartmann (2006) [ | Yes | Yes | Yes | Yes | Yes | Yes | Yes | Yes | No | Yes |
| Kerker (2008) [ | Yes | Yes | Yes | Yes | No | Yes | Unclear | Yes | Unclear | Yes |
| Kuehle (2007) [ | Yes | Yes | Yes | Yes | No | Yes | Yes | Noc | Yes | Yes |
| Lauenstein (2002) [ | Yes | Yes | Yes | Yes | Yes | Yes | Yes | Yes | Unclear | Unclear |
| Lauenstein (2005) [ | Yes | Yes | Yes | Yes | Yes | Yes | Yes | Yes | Yes | Yes |
| Leung (2004) [ | Yes | Yes | Yes | Yes | Yes | Yes | Yes | Noc | Yes | Yes |
| Luboldt (1998) [ | Yes | No | Yes | Yes | Yes | Yes | Yes | Unclear | Yes | Yes |
| Luboldt (2000) [ | Yes | Unclear | Yes | Yes | Yes | Yes | Yes | Yes | Yes | Yes |
| Saar (2007) [ | Yes | Yes | Yes | Yes | Yes | Yes | Yes | Unclear | Unclear | Yes |
| Saar (2008) [ | Yes | Yes | Yes | Yes | Yes | Yes | Yes | Noc | Unclear | Yes |
aMaximum time interval was defined as 4 weeks
bDefined as at least 90% receiving the reference test
cSegmental unblinding
dReassessment in new session if needed
Study characteristics of included studies
| Study | Number of subjects for analysis | Consecutively recruited | Indications for colonoscopya | Patients with CRC or polyps >10 mm | Prevalence of patients with polyps >10 mm (%) | Mean age ±SD; range (years) | Sex ratio (male/female) | Study design |
|---|---|---|---|---|---|---|---|---|
| Achiam (2008) [ | 47 | Yes | NA | 12/47 | 25 | NA | NA | Prospective |
| Florie (2007) [ | 200 | Yes | 5–8 | 12/200 | 6 | 58 ± 12 | 128/72 | Prospective |
| Goehde (2005) [ | 42 | Yes | 1–6 | NA | NA | NA | 18/24 | Prospective |
| Hartmann (2006) [ | 92 | Yes | 1,2,3,9,10 | 17/92 | 18 | 61.5 ± 14.5 | 52/40 | Prospective |
| Kerker (2008) [ | 80 | Unclear | 1,3,4,10 | NA | NA | NA | NA | Prospective |
| Kuehle (2007) [ | 315 | NA | 1 | 20/315 | 6.3 | NA | NA | Prospective |
| Lauenstein (2002) [ | 24 | NA | 2,4,9 | 13/24 | 25 | 57.4; 33–78 | 12/12 | Prospective |
| Lauenstein (2005) [ | 37 | NA | 2,4,5,9 | NA | NA | NA | NA | Prospective |
| Leung (2004) [ | 156 | NA | 1,2,4,7,10 | 11/156 | 4.5 | 55.2 ± 9.1 | 74/82 | Prospective |
| Luboldt (1998) [ | 21 | NA | NA | 4/21 | 19 | NA | NA | Prospective |
| Luboldt (2000) [ | 117 | NA | NA | 14/117 | 12 | NA | NA | Prospective |
| Saar (2007) [ | 120 | Yes | 6,9 | NA | NA | 69; 22–87 | 56/64 | Prospective |
| Saar (2008) [ | 34 | NA | 2,4,6,7,9 | NA | NA | 60b; 21–88 | 21/13 | Prospective |
NA not available
a 1 screening, 2 faecal blood, 3 abdominal complaints, 4 change in bowel habits, 5 personal history of CRC, 6 personal history of polyps, 7 family history of CRC, 8 family history of polyps, 9 faecal occult blood testing positive (FOBT+), 10 anaemia
bMedian age
MR imaging characteristics of included studies
| Study | Type of bowel preparation/faecal tagging | Dietary restrictions | Technical method | Type of enema used | Volume (L) | Enema pressure | Spasmolytic drugs (mg) | Intravenous contrast agent (mmol/kg) |
|---|---|---|---|---|---|---|---|---|
| Achiam (2008) [ | 4 × 200 cc ferumoxsil/barium sulphate | Low-fibre diet | Dark-lumen | Water-based | 2–2.5 | 1.5 m H2O | Buscopan® (40) | Dotarem® (0.2) |
| Florie (2007) [ | 6 × 10 cc Magnevist® + lactulose | Low-fibre diet | Bright-lumen | Mixture water/gadolinium-based | 1.9 | 0.8 m H2O | Buscopan® (20) | None |
| Goehde (2005) [ | 6 × 150 cc barium sulphate | Low-fibre diet | Dark-lumen | Water-based | 2–2.5 | NA | Buscopan® (40) | Multihance® (0.2) |
| Hartmann (2006) [ | Bowel cleansing (4 L PEG–electr) | No | Dark-lumen | Water-based | 2–2.5 | NA | Buscopan® (40) | Multihance® |
| Kerker (2008) [ | Bowel cleansing (4 L PEG solution) | No | Dark-lumen | Water-based | 2 | NA | Buscopan® (40) | Magnevist® |
| Kuehle (2007) [ | 6 × 150 cc solution (5% Gastrografin/1% barium/0.2% bean gum) | No | Dark-lumen | Water-based | 2 | 1–1.5 | Buscopan® (40) | Dotarem® (0.2) |
| Lauenstein (2002) [ | 4/5 × 200 cc barium sulphate | Low-fibre diet | Dark-lumen | Water-based | 1.5–2.5 | 1.0–1.5 m H2O | Buscopan® (20) | Multihance® (0.2) |
| Lauenstein (2005) [ | Bowel cleansing (3 L electrolyte solution) | No | Dark-lumen | Water-based | 2.5 | 1.0 m H2O | Buscopan® (20) | Multihance® (0.2) |
| Leung (2004) [ | Bowel cleansing (phospho soda) | No | Dark-lumen | Air-based | 30–40 puffsa | NA | Hyoscine (20) | None |
| Luboldt (1998) [ | Bowel cleansing (3 L PEG solution) | No | Bright-lumen | Mixture water/gadolinium-based | 1.5–2.0 | 1–2 m H2O | Buscopan® (20) | Magnevist® (0.1) |
| Luboldt (2000) [ | Bowel cleansing (3 L bowel preparation solution) | No | Bright-lumen | Mixture water/gadolinium-based | 1.8–3.0 | 1.0 m H2O | Buscopan® (20) | None |
| Saar (2007) [ | Bowel cleansing (4–6 L PEG–electr) | No | Bright-lumen | Mixture water/gadolinium-based | 1.5–2.5 | NA | Buscopan® (20) | None |
| Saar (2008) [ | Bowel cleansing (3–4 L electrolytic solution) | No | Dark-lumen | Water-based | 2.5 | ±1 m H2O | Buscopan® (40) | Multihance® (0.2) |
PEG polyethylene glycol, electr electrolyte
aManual air-inflation
MRI technical parameters of included studies
| Study | Type of MRI scanner | Field strength (T) | Sequence | TR (ms)/TE(ms) | Matrix | Slice thickness (mm) | FOV (mm) | Prone (P)/ supine (S) | Imaging plane | Breath hold (s) | Examination time (min) |
|---|---|---|---|---|---|---|---|---|---|---|---|
| Achiam (2008) [ | Philips | 1.5 | 3D T1 | 3.2/1.6 | NA | 1.7 | 420 | NA | Coronal | 22 | NA |
| Florie (2007) [ | GE | 1.5 | 3D T1 | 5.4/1.6 | 256 × 192 | 2 | 480 × 336 | P/S | Coronal | 10–20 | 45 |
| 2D T2 | 1,354/64 | 320 × 224 | 4 | 460 × 322 | P/S | Axial | |||||
| 2D T2 | 1,050/62 | 256 × 160 | 5 | 400 × 280 | P/S | Coronal | |||||
| Philips | 1.5 | 3D T1 | 5.6/1.9 | 512 × 304 | 2.5 | 480 × 432 | P/S | Coronal | 10–20 | 45 | |
| 2D T2 | 676/165 | 256 × 256 | 5 | 400 × 320 | P/S | Axial | |||||
| 2D T2 | 867/195 | 256 × 256 | 5 | 480 × 432 | P/S | Coronal | |||||
| Philips | 3.0 | 3D T1 | 5.4/2.0 | 192 × 163 | 2.5 | 480 × 384 | P/S | Coronal | 10–20 | 45 | |
| 2D T2 | 1,002/60 | 455 × 455 | 5 | 455 × 455 | P/S | Axial | |||||
| 2D T2 | 1,002/60 | 304 × 274 | 5 | 485 × 485 | P/S | Coronal | |||||
| Goehde (2005) [ | Siemens | 1.5 | 3D T1 | 1.64/0.6 | 512 × 460 | 1.5–2.0 | 450 × 450 | P | Coronal | 22 | NA |
| Hartmann (2006) [ | Siemens | 1.5 | 3D T1 | 3.1/1.17 | NA | 1.5–2.0 | 400 × 400 | P | Coronal | 22 | NA |
| Kerker (2008) [ | Siemens | 1.5 | 3DT1 | 1.92/0.8 | NA | NA | 400 | NA | Coronal | 20–25 | NA |
| 2DT1 | 522/22 | NA | NA | 240 | NA | Sagittal | |||||
| Kuehle (2007) [ | Siemens | 1.5 | 3D T1 | 3.08/1.13 | 168 × 256 | 1.8–2.4 | 500 (z) | P | Coronal | <20 | NA |
| 2DT1 | 125/1.83 | 168 × 256 | NA | 500 | P | Axial | |||||
| 2D T1/T2a | 3.79/1.9 | 205 × 256 | NA | 400 | P | Coronal | |||||
| Lauenstein (2002) [ | Siemens | 1.5 | 3D T1 | 1.64/0.60 | 460 × 512 | 1.57 | 450 | P | Coronal | 22 | <20 |
| Lauenstein (2005) [ | Siemens | 1.5 | 3D T1 | 3.1/1.1 | 180 × 256 | 1.8b | 400 | S | Coronal | 22 | NA |
| 2D T1/T2a | 4.5/2.2 | 402 × 512 | 3 | 400 | P/S | Coronal | 21 | ||||
| Leung (2004) [ | Siemens | 1.5 | 2D T2 | ∞/56 | 256 × 134 | 4 | 340–400 | P/S | Coronal/axial | 24 | 20.6 ± 2.7 |
| 3D T1 | 3.1/1.1 | 256 × 165 | 2.5 | 380–400 | P/S | Coronal | 22 | ||||
| Luboldt (1998) [ | GE | 1.5 | 3D T1 | 4.7/2 | 256 × 160 | 2.2–3.6 | 380–460 | P/S | NA | 28 | NA |
| 2D T1 | 6/1.3 | 256 × 160 | 10 | 360–400 | NA | NA | 28 | ||||
| Luboldt (2000) [ | GE | 1.5 | 3D T1 | 3.8/2.5 | 384 × 192 | 2–3 | 340–420 | P/S | Coronal | <30 | NA |
| 2D T2 | ∞/65 | 256 × 160 | 6 | 340–420 | P/S | Coronal | <30 | ||||
| Saar (2007) [ | Philips | 1.5 | 2D T1 | 4/1.7 | 128 × 128 | 150 | 450 × 450 | P/S | Coronal | 20–23 | 21.5 |
| 3D T1 | 3.3/1.3 | 256 × 256 | 1.5 | 450 × 450 | P/S | Coronal | 20–23 | ||||
| Saar (2008) [ | Siemens | 3.0 | 3D T1 | 2.53/1.03 | 384 × 384 | 1.6 × 1.0 × 1.5c | 400 × 400 | S | Coronal | NA | NA |
| 3D T1 | 3.14/1.51 | 256 × 256 | 2.4 × 1.x1.5c | 400 × 400 | S | Coronal |
NA not available
aBalanced gradient echo (GRE)
bUsing zero-filling interpolation
cVoxel size
Image analysis characteristics
| Study | Number of observers | Experience of observers definedb | Data analysis methodc | Consensus reading | Review time (min) | Image quality assessmente |
|---|---|---|---|---|---|---|
| Achiam (2008) [ | 2 | 50 cases | 2D | Nod | NA | Yes |
| Florie (2007) [ | 2 | 40 cases | 2D | Nod | NA | Yes |
| Goehde (2005) [ | 2 | No | 2D | Yes | NA | Yes |
| Hartmann (2006) [ | 5a | No | 2D/3D | Yes | NA | No |
| Kerker (2008) [ | 2 | No | NA | NA | NA | No |
| Kuehle (2007) [ | 2 | No | 2D/3D | Yes | NA | Yes |
| Lauenstein (2002) [ | 2 | No | 2D/3D | Yes | NA | Yes |
| Lauenstein (2005) [ | 2 | No | 2D/3D | Yes | NA | Yes |
| Leung (2004) [ | 2 | No | 2D/3D | Yes | NA | No |
| Luboldt (1998) [ | 2 | No | 2D/3D | Yes | 35–55 | Yes |
| Luboldt (2000) [ | 2 | No | 2D/3D | Yes | NA | Yes |
| Saar (2007) [ | 2 | >50 cases | 2D/3D | Yes | 23.5 (13.5–48) | Yes |
| Saar (2008) [ | 2 | >50 cases | 2D | Yes | NA | Yes |
NA not available
a2 radiologists and 3 gastroenterologists
bDefined as exact number of validated MR-colonography training cases
c 2D multiplanar reformation (MPR), 3D virtual endoscopy
dQuality assessment includes assessment of colonic distension, artefacts, lumen homogeneity or SNR/CNR
eInterobserver agreement
Fig. 2a Forest plot of per patient sensitivity and specificity, including sensitivity and specificity estimates, for all polyps. FN false-negative, FP false-positive, TN true-negative, TP true-positive values. Lauenstein (2005) compared two different sequences in the same study population [44]; results of both sequences are used for calculating sensitivity and specificity estimates. Heterogeneity (I 2) between study results for sensitivities was 86% (CI 79–91%) and for specificities 58% (CI 28–76%). b Forest plot of per patient sensitivity and specificity, including pooled sensitivity and specificity, for polyps 10 mm or larger. Heterogeneity (I 2) between study results for sensitivities was 37% (CI 10–63%) and for specificities 60% (CI 17–80%)
Fig. 3Forest plot of per polyp sensitivity, including pooled per polyp sensitivity, for polyps smaller than 6 mm (a), polyps 6–9 mm (b) and polyps 10 mm or larger (c). FP false-positive, TN true-negative values. Heterogeneity (I 2) among study results for sensitivities for polyps smaller than 6 mm was 81% (CI 65–90%); polyps 6–9 mm, 80% (CI 62–89%); and polyps 10 mm or larger, 51% (CI 8–74%)
Fig. 4Forest plot of sensitivity of MR-colonography in the detection of CRC. Lauenstein (2005) and Saar (2008) compared two different sequences in the same study population [44, 49]; results of both sequences are outlined in this forest plot
Characteristics of the search strategy
| Search terms | Number of hits | |
|---|---|---|
| PUBMED | ||
| #1 | MR-colonography [keyword] | 122 |
| #2 | Magnetic resonance colonography [keyword] | 175 |
| #3 | Virtual colonoscopy [keyword] | 1,230 |
| #4 | Magnetic resonance imaging [MeSH} | 225,890 |
| #5 | #3 AND #4 | 115 |
| #6 | #1 OR #2 OR #5 | 209 |
| Limits | ||
| Human and publication date: from 1997 to 2009 | 191 | |
| EMBASE | ||
| #1 | MR-colonography [keyword] | 99 |
| #2 | Magnetic resonance colonography [keyword] | 78 |
| #3 | Exp Nuclear Magnetic Resonance Imaging [MeSH] | 232,502 |
| #4 | Virtual colonoscopy [keyword] | 365 |
| #5 | #3 AND #4 | 51 |
| #6 | #1 OR #2 OR #5 | 175 |
| Limits | 156 | |
| Human and publication date: from 1997 to 2009 | ||
| Cochrane | ||
| #1 | MR-colonography [Title, Abstract, Keyword] | 3 |
| #2 | Magnetic resonance colonography [Title, Abstract, Keyword] | 6 |
| #3 | Magnetic resonance imaging [Title, Abstract, Keyword] | 4,008 |
| #4 | Virtual colonoscopy [Title, Abstract, Keyword] | 32 |
| #5 | #3 AND #4 | 1 |
| #6 | #1 OR #2 OR #5 | 6 |