| Literature DB >> 19923569 |
Jerry W Pickering1, Matthew T Larson, Thomas B Martins, Susan S Copple, Harry R Hill.
Abstract
In our 14-valent Luminex assay for pneumococcal antibodies, we identified two groups of sera that caused false-positive results. The first group bound nonspecifically to the Luminex microspheres. The second group reacted specifically with bovine serum albumin (BSA). We describe here methods that eliminated the false-positive reactivity of both groups.Entities:
Mesh:
Substances:
Year: 2009 PMID: 19923569 PMCID: PMC2812081 DOI: 10.1128/CVI.00329-09
Source DB: PubMed Journal: Clin Vaccine Immunol ISSN: 1556-679X