Wayne A Hing1, Keith Rome, Alyse Fm Cameron.
Abstract
BACKGROUND: Diagnostic ultrasound provides a method of analysing soft tissue structures of the musculoskeletal system effectively and reliably. The aim of this study was to evaluate within and between session reliability of measuring muscle dorso-plantar thickness, medio-lateral length and cross-sectional area, of the abductor hallucis muscle using two different ultrasound machines, a higher end Philips HD11 Ultrasound machine and clinically orientated Chison 8300 Deluxe Digital Portable Ultrasound System.Entities:
Year: 2009 PMID: 19917127 PMCID: PMC2784754 DOI: 10.1186/1757-1146-2-33
Source DB: PubMed Journal: J Foot Ankle Res ISSN: 1757-1146 Impact factor: 2.303
Figure 1Ultrasound image of abductor hallucis muscle with dorso-plantar thickness and medio-lateral width points marked from Phillips HD11(longitudinal view).
Figure 2Ultrasound image of abductor hallucis muscle from Chison 8300 (longitudinal view).
Descriptive statistics of abductor hallucis muscle parameters
| Phillips HD11 | Chison 8300 | ||
|---|---|---|---|
| Day | Mean ± SD | Mean ± SD | |
| Dorso-Plantar thickness (mm) | 1 | 11.56 ± 1.07 | 11.54 ± 1.13 |
| 2 | 11.54 ± 1.02 | 11.60 ± 1.07 | |
| Medio-lateral width (mm) | 1 | 28.98 ± 2.77 | 28.81 ± 2.75 |
| 2 | 29.03 ± 2.60 | 28.97 ± 2.62 | |
| Cross-sectional area (mm2) | 1 | 269.33 ± 35.99 | 262.35 ± 38.60 |
| 2 | 275.99 ± 36.04 | 270.26 ± 34.31 |
Within session reliability
| ICC [3.1] | 95% CI | SEM | SRD | |
|---|---|---|---|---|
| Dorso-plantar thickness (mm) | 0.97 | (0.95 to 0.99) | 0.12 | 0.34 |
| Medio-lateral width (mm) | 0.96 | (0.95 to 0.98) | 0.87 | 2.47 |
| Cross-sectional area (mm2) | 0.98 | (0.96 to 0.98) | 9.36 | 26.59 |
| Dorso-planatr thickness (mm) | 0.99 | (0.99 to 0.99) | 0.09 | 0.26 |
| Medio-lateral length (mm) | 0.95 | (0.92 to 0.97) | 1.06 | 3.01 |
| Cross-sectional area (mm2) | 0.99 | (0.98 to 0.99) | 6.73 | 19.12 |
Between session reliability: Single measures
| ICC [3.1] | 95% CI | SEM | SRD | |
|---|---|---|---|---|
| Dorso-plantar thickness (mm) | 0.95 | (0.91 to 0.97) | 0.46 | 1.31 |
| Medio-lateral width (mm) | 0.88 | (0.81 to 0.93) | 0.28 | 0.79 |
| Cross-sectional area (mm2) | 0.65 | (0.48 to 0.78) | 40.10 | 113.93 |
| Dorso-plantar thickness (mm) | 0.92 | (0.87 to 0.95) | 0.13 | 0.37 |
| Medio-lateral width (mm) | 0.78 | (0.66 to 0.87) | 0.08 | 0.22 |
| Cross-sectional area (mm2) | 0.64 | (0.47 to 0.77) | 35.30 | 100.29 |
Between session reliability: average measures
| ICC [3.1] | 95% CI | SEM | SRD | |
|---|---|---|---|---|
| Dorso-plantar thickness (mm) | 0.97 | (0.95 to 0.98) | 0.25 | 0.71 |
| Medio-lateral width (mm) | 0.94 | (0.90 to 0.96) | 0.90 | 2.56 |
| Cross-sectional area (mm2) | 0.79 | (0.65 to 0.88) | 20.15 | 57.24 |
| Dorso-plantar thickness (mm) | 0.96 | (0.93 to 0.97) | 0.32 | 0.91 |
| Medio-lateral width (mm) | 0.88 | (0.80 to 0.93) | 1.20 | 3.41 |
| Cross-sectional area (mm2) | 0.78 | (0.64 to 0.87) | 21.78 | 61.88 |
Figure 3Bland & Altman plot for Phillips HD11 and Chison 8300 results for abductor hallucis Dorso-plantar thickness.
Figure 4Bland & Altman plot for Phillips HD11 and Chison 8300 results for abductor hallucis medio-lateral width.
Figure 5Bland & Altman plot for Phillips HD11 and Chison 8300 results for abductor hallucis cross-sectional area.