Agreement between glucose trends derived from three simultaneously worn continuous glucose sensors.

Allen B King1, Dana Armstrong, Caroline Chu, Manoj Sharma, David Price.   

Abstract

BACKGROUND: Sensors detect the rate and direction of glucose trend. They need to be accurate and reproducible as could be evidenced by strong agreement between multiple sensors. We evaluated this relationship through simultaneously worn glucose sensors using several methods of slope analysis.
METHODS: Ten type 1 diabetic, insulin pump-treated subjects were studied while simultaneously wearing three CGMS Gold sensors each. Sensors were placed in the right abdomen (reference), left abdomen, and left upper arm. Sensors were calibrated and chronologically aligned. Data were only interpreted and included if there were 24 hours of data simultaneously obtained from all three sensors.
RESULTS: Using a two-point derived slope, increasing the duration of the trend from 5 to 60 minutes improved agreement between sensors. Using a 20-minute rolling average trend (using every 5-minute glucose value during the 20 minutes) improved the agreement to 94.3%. Finally, using whichever of the two comparator sensor rolling average trends was closest to the reference (better of two), the agreement improved to 98.2%. However, for these trend analysis methods, when the absolute reference rate of change was more than 1 mg/dl/min, the agreement decreased. Even with the best analysis approach, at an absolute reference sensor rate of change of >2 mg/dl/min, the agreement between sensors was only 40.0%.
CONCLUSION: Despite several methods of analysis, trend agreement from multiple sensors diminishes as the absolute rate of change of reference glucose increases.

Entities:  

Keywords:  closed loop; continuous glucose monitoring; glucose trends; trend analysis

Year:  2008        PMID: 19885270      PMCID: PMC2769805          DOI: 10.1177/193229680800200515

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  9 in total

Review 1.  Continuous glucose monitoring: roadmap for 21st century diabetes therapy.

Authors:  David C Klonoff
Journal:  Diabetes Care       Date:  2005-05       Impact factor: 19.112

Review 2.  Continuous glucose monitoring and closed-loop systems.

Authors:  R Hovorka
Journal:  Diabet Med       Date:  2006-01       Impact factor: 4.359

3.  Accuracy of the 5-day FreeStyle Navigator Continuous Glucose Monitoring System: comparison with frequent laboratory reference measurements.

Authors:  Richard L Weinstein; Sherwyn L Schwartz; Ronald L Brazg; Jolyon R Bugler; Thomas A Peyser; Geoffrey V McGarraugh
Journal:  Diabetes Care       Date:  2007-03-02       Impact factor: 19.112

4.  Algorithms for care in adults using continuous glucose monitoring.

Authors:  Irl B Hirsch
Journal:  J Diabetes Sci Technol       Date:  2007-01

5.  The accuracy of the FreeStyle Navigator continuous glucose monitoring system in children with type 1 diabetes.

Authors:  Darrell M Wilson; Roy W Beck; William V Tamborlane; Mariya J Dontchev; Craig Kollman; Peter Chase; Larry A Fox; Katrina J Ruedy; Eva Tsalikian; Stuart A Weinzimer
Journal:  Diabetes Care       Date:  2007-01       Impact factor: 19.112

Review 6.  Development of the implantable glucose sensor. What are the prospects and why is it taking so long?

Authors:  D A Gough; J C Armour
Journal:  Diabetes       Date:  1995-09       Impact factor: 9.461

7.  The use of two continuous glucose sensors during and after surgery.

Authors:  T M Vriesendorp; J H DeVries; F Holleman; M Dzoljic; J B L Hoekstra
Journal:  Diabetes Technol Ther       Date:  2005-04       Impact factor: 6.118

8.  A fully implantable subcutaneous glucose sensor array: enhanced accuracy from multiple sensing units and a median-based algorithm.

Authors:  W Kenneth Ward; Heather M Casey; Matthew J Quinn; Isaac F Federiuk; Michael D Wood
Journal:  Diabetes Technol Ther       Date:  2003       Impact factor: 6.118

9.  Reproducibility of glucose measurements using the glucose sensor.

Authors:  Muriel Metzger; Gil Leibowitz; Julio Wainstein; Benjamin Glaser; Itamar Raz
Journal:  Diabetes Care       Date:  2002-07       Impact factor: 19.112

  9 in total
  1 in total

1.  Consistency of Continuous Ambulatory Interstitial Glucose Monitoring Sensors.

Authors:  Pei T Wu; David E Segovia; Cathy C Lee; Kim-Lien Nguyen
Journal:  Biosensors (Basel)       Date:  2018-05-16
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.