Numerical simulation of the effect of rate of change of glucose on measurement error of continuous glucose monitors.

Marc B Taub1, Thomas A Peyser, J Erik Rosenquist.   

Abstract

BACKGROUND: A 5-day in-patient study designed to assess the accuracy of the FreeStyle Navigator Continuous Glucose Monitoring System revealed that the level of accuracy of the continuous sensor measurements was dependent on the rate of glucose change. When the absolute rate of change was less than 1 mg*dl(-1)*min(-1) (75% of the time), the median absolute relative difference (ARD) was 8.5%, with 85% of all points falling within the A zone of the Clarke error grid. When the absolute rate of change was greater than 2 mg*dl(-1)*min(-1) (8% of the time), the median ARD was 17.5%, with 59% of all points falling within the Clarke A zone.
METHOD: Numerical simulations were performed to investigate effects of the rate of change of glucose on sensor measurement error. This approach enabled physiologically relevant distributions of glucose values to be reordered to explore the effect of different glucose rate-of-change distributions on apparent sensor accuracy.
RESULTS: The physiological lag between blood and interstitial fluid glucose levels is sufficient to account for the observed difference in sensor accuracy between periods of stable glucose and periods of rapidly changing glucose.
CONCLUSIONS: The role of physiological lag on the apparent decrease in sensor accuracy at high glucose rates of change has implications for clinical study design, regulatory review of continuous glucose sensors, and development of performance standards for this new technology. This work demonstrates the difficulty in comparing accuracy measures between different clinical studies and highlights the need for studies to include both relevant glucose distributions and relevant glucose rate-of-change distributions.

Entities:  

Keywords:  Navigator; continuous glucose monitoring system; glucose; glucose monitoring; glucose sensors

Year:  2007        PMID: 19885136      PMCID: PMC2769653          DOI: 10.1177/193229680700100512

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  24 in total

Review 1.  Glucose sensors: toward closed loop insulin delivery.

Authors:  Chee W Chia; Christopher D Saudek
Journal:  Endocrinol Metab Clin North Am       Date:  2004-03       Impact factor: 4.741

Review 2.  Continuous glucose monitoring: roadmap for 21st century diabetes therapy.

Authors:  David C Klonoff
Journal:  Diabetes Care       Date:  2005-05       Impact factor: 19.112

3.  Accuracy of the 5-day FreeStyle Navigator Continuous Glucose Monitoring System: comparison with frequent laboratory reference measurements.

Authors:  Richard L Weinstein; Sherwyn L Schwartz; Ronald L Brazg; Jolyon R Bugler; Thomas A Peyser; Geoffrey V McGarraugh
Journal:  Diabetes Care       Date:  2007-03-02       Impact factor: 19.112

Review 4.  Can interstitial glucose assessment replace blood glucose measurements?

Authors:  K Rebrin; G M Steil
Journal:  Diabetes Technol Ther       Date:  2000       Impact factor: 6.118

Review 5.  Should minimal blood glucose variability become the gold standard of glycemic control?

Authors:  Irl B Hirsch; Michael Brownlee
Journal:  J Diabetes Complications       Date:  2005 May-Jun       Impact factor: 2.852

6.  Accuracy of newer-generation home blood glucose meters in a Diabetes Research in Children Network (DirecNet) inpatient exercise study.

Authors:  Stuart A Weinzimer; Roy W Beck; H Peter Chase; Larry A Fox; Bruce A Buckingham; William V Tamborlane; Craig Kollman; Julie Coffey; Dongyuan Xing; Katrina J Ruedy
Journal:  Diabetes Technol Ther       Date:  2005-10       Impact factor: 6.118

7.  Quantifying temporal glucose variability in diabetes via continuous glucose monitoring: mathematical methods and clinical application.

Authors:  Boris P Kovatchev; William L Clarke; Marc Breton; Kenneth Brayman; Anthony McCall
Journal:  Diabetes Technol Ther       Date:  2005-12       Impact factor: 6.118

Review 8.  Hypoglycemia warning signal and glucose sensors: requirements and concepts.

Authors:  Tim Heise; Theodor Koschinsky; Lutz Heinemann; Volker Lodwig
Journal:  Diabetes Technol Ther       Date:  2003       Impact factor: 6.118

9.  The effect of intensive treatment of diabetes on the development and progression of long-term complications in insulin-dependent diabetes mellitus.

Authors:  D M Nathan; S Genuth; J Lachin; P Cleary; O Crofford; M Davis; L Rand; C Siebert
Journal:  N Engl J Med       Date:  1993-09-30       Impact factor: 91.245

Review 10.  Hypoglycaemia: the limiting factor in the glycaemic management of Type I and Type II diabetes.

Authors:  P E Cryer
Journal:  Diabetologia       Date:  2002-04-26       Impact factor: 10.122

View more
  2 in total

Review 1.  Continuous glucose monitoring: a review for behavioral researchers.

Authors:  Julie Wagner; Howard Tennen; Howard Wolpert
Journal:  Psychosom Med       Date:  2012-05       Impact factor: 4.312

2.  A head-to-head comparison of personal and professional continuous glucose monitoring systems in people with type 1 diabetes: Hypoglycaemia remains the weak spot.

Authors:  Othmar Moser; Marlene Pandis; Felix Aberer; Harald Kojzar; Daniel Hochfellner; Hesham Elsayed; Melanie Motschnig; Thomas Augustin; Philipp Kreuzer; Thomas R Pieber; Harald Sourij; Julia K Mader
Journal:  Diabetes Obes Metab       Date:  2018-12-25       Impact factor: 6.577

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.