Literature DB >> 19880185

A 27-gauge instrument system for transconjunctival sutureless microincision vitrectomy surgery.

Yusuke Oshima1, Taku Wakabayashi, Tatsuhiko Sato, Masahito Ohji, Yasuo Tano.   

Abstract

OBJECTIVES: To evaluate the efficiency, preliminary safety, and feasibility of a 27-gauge instrument system for transconjunctival microincision vitrectomy surgery (MIVS) in a variety of vitreoretinal diseases.
DESIGN: Experimental, interventional case series. PARTICIPANTS: Thirty-one eyes (31 patients) underwent a variety of vitreoretinal procedures using the 27-gauge transconjunctival MIVS system to treat epiretinal membrane (n = 10), idiopathic macular holes (n = 7), diabetic vitreous hemorrhage (n = 5), vitreous opacity with suspicion of intraocular lymphoma (n = 4), focal diabetic traction retinal detachment (n = 3), macular traction syndrome (n = 1), and macular edema secondary to central retinal vein occlusion (n = 1).
METHODS: We developed a 27-gauge instrument system that includes an infusion line, a high-speed vitreous cutter, an illumination system, and a variety of vitreoretinal instruments, such as membrane forceps and sharp-tipped endophotocoagulation probes. The duty cycle of the 27- and 25-gauge cutters was measured for several cut rates using a high-speed imaging camera. Infusion and aspiration rates were measured using balanced saline solution (BSS) and porcine vitreous with different aspiration levels. Surgical outcomes, including anatomic success, visual outcomes, operating times, and intraoperative and postoperative complications, were evaluated. MAIN OUTCOME MEASURES: Duty cycle of cutters, infusion and aspiration rates, and surgical results of 27-gauge vitrectomy.
RESULTS: Although the infusion and aspiration rates of the 27-gauge system measured in BSS were reduced to an average of 62% and 80%, respectively, compared with those of the 25-gauge system, the duty cycle of the 27-gauge cutter, 61% at 1000 cpm and 38% at 1500 cpm, was equal to or better than those of the 25-gauge cutter (62% and 28%, respectively). Analysis of the fluid dynamics showed that vented gas-forced infusion can be set to range from 20 to 30 mmHg to control intraocular pressure (IOP) during 27-gauge vitrectomy. Anatomic success was achieved in all study eyes (100%); 20 eyes (65%) had visual improvement of 3 lines or more. No eyes required conversion to larger gauge instrument. All sclerotomies self-sealed without hypotony (IOP < or = 7 mmHg) from 1 day postoperatively.
CONCLUSIONS: Although the fluid dynamics and cutting efficiency of 27-gauge instruments are lower compared with 25-gauge MIVS, the 27-gauge system is feasible and may reduce concerns about wound sealing-related complications in selected cases. FINANCIAL DISCLOSURE(S): Proprietary or commercial disclosure may be found after the references. Copyright 2010 American Academy of Ophthalmology. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2009        PMID: 19880185     DOI: 10.1016/j.ophtha.2009.06.043

Source DB:  PubMed          Journal:  Ophthalmology        ISSN: 0161-6420            Impact factor:   12.079


  76 in total

1.  Ocular biocompatibility of nitinol intraocular clips.

Authors:  Jeffrey L Olson; Raul Velez-Montoya; Michael Erlanger
Journal:  Invest Ophthalmol Vis Sci       Date:  2012-01-25       Impact factor: 4.799

2.  Microincision cataract surgery combined with vitrectomy: a case series.

Authors:  A Jalil; L Steeples; S Subramani; M S Bindra; F Dhawahir-Scala; N Patton
Journal:  Eye (Lond)       Date:  2014-01-10       Impact factor: 3.775

3.  One-port vs. three-port diagnostic vitrectomy for posterior segment diseases of unknown origin.

Authors:  Marco Coassin; Lorenzo Braconi; Giancarlo Sborgia; Giuseppe Mangano; Valentina Mastrofilippo; Antonio Di Zazzo; Luigi Fontana; Luca Cimino
Journal:  Int Ophthalmol       Date:  2020-07-09       Impact factor: 2.031

4.  A Comparative Study of 23-Gauge and 27-Gauge Vitrectomy for Puckers or Floaters, Including Evaluation of the Effect of Combined Phaco-Vitrectomy Surgery on Postoperative Outcome.

Authors:  Peter Stalmans
Journal:  Ophthalmologica       Date:  2021-02-09       Impact factor: 3.250

5.  Endophthalmitis associated with Purpureocillium lilacinum during infliximab treatment for surgically induced necrotizing scleritis, successfully treated with 27-gauge vitrectomy.

Authors:  Masaaki Yoshida; Shunji Yokokura; Hiroshi Kunikata; Naoko Takada; Kazuichi Maruyama; Masahiro Toyokawa; Kazushi Kashio; Mitsuo Kaku; Toru Nakazawa
Journal:  Int Ophthalmol       Date:  2017-04-20       Impact factor: 2.031

6.  Long-term outcomes of sutureless 25-G+ pars-plana vitrectomy for the management of diabetic tractional retinal detachment.

Authors:  Mikel Mikhail; Andre Ali-Ridha; Sarah Chorfi; Michael A Kapusta
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2016-08-02       Impact factor: 3.117

7.  Multicenter survey of sutureless 27-gauge vitrectomy for primary rhegmatogenous retinal detachment: a consecutive series of 410 cases.

Authors:  Yoichiro Shinkai; Yusuke Oshima; Kazuhito Yoneda; Jiro Kogo; Hisanori Imai; Akira Watanabe; Yoshitsugu Matsui; Kotaro Suzuki; Chie Sotozono
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2019-09-02       Impact factor: 3.117

8.  Outcomes of transconjunctival sutureless 27-gauge vitrectomy for vitreoretinal diseases.

Authors:  Jie Li; San-Mei Liu; Wen-Tao Dong; Fang Li; Cai-Hong Zhou; Xiao-Dan Xu; Jie Zhong
Journal:  Int J Ophthalmol       Date:  2018-03-18       Impact factor: 1.779

9.  Comparative study of 27-gauge and 25-gauge vitrectomy performed as day surgery.

Authors:  Ryusaburo Mori; Saigen Naruse; Hiroyuki Shimada
Journal:  Int Ophthalmol       Date:  2017-07-03       Impact factor: 2.031

10.  Outcomes of transconjunctival sutureless 27-gauge vitrectomy with silicone oil infusion.

Authors:  Okan Toygar; Cindy W Mi; Daniel M Miller; Christopher D Riemann
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2016-04-19       Impact factor: 3.117

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.