Ocular biocompatibility of nitinol intraocular clips.

Jeffrey L Olson1, Raul Velez-Montoya, Michael Erlanger.   

Abstract

PURPOSE: To evaluate the tolerance and biocompatibility of a preformed nitinol intraocular clip in an animal model after anterior segment surgery.
METHODS: Yucatan mini-pigs were used. A 30-gauge prototype injector was used to attach a shape memory nitinol clip to the iris of five pigs. Another five eyes received conventional polypropylene suture with a modified Seipser slip knot. The authors compared the surgical time of each technique. All eyes underwent standard full-field electroretinogram at baseline and 8 weeks after surgery. The animals were euthanized and eyes collected for histologic analysis after 70 days (10 weeks) postsurgery. The corneal thickness, corneal endothelial cell counts, specular microscopy parameters, retina cell counts, and electroretinogram parameters were compared between the groups. A two sample t-test for means and a P value of 0.05 were use for assessing statistical differences between measurements.
RESULTS: The injection of the nitinol clip was 15 times faster than conventional suturing. There were no statistical differences between the groups for corneal thickness, endothelial cell counts, specular microscopy parameters, retina cell counts, and electroretinogram measurements.
CONCLUSIONS: The nitinol clip prototype is well tolerated and showed no evidence of toxicity in the short-term. The injectable delivery system was faster and technically less challenging than conventional suture techniques.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22064995      PMCID: PMC3292371          DOI: 10.1167/iovs.11-8496

Source DB:  PubMed          Journal:  Invest Ophthalmol Vis Sci        ISSN: 0146-0404            Impact factor:   4.799


  39 in total

1.  Passivation of nitinol wire for vascular implants--a demonstration of the benefits.

Authors:  B O'Brien; W M Carroll; M J Kelly
Journal:  Biomaterials       Date:  2002-04       Impact factor: 12.479

Review 2.  Surface, corrosion and biocompatibility aspects of Nitinol as an implant material.

Authors:  Svetlana A Shabalovskaya
Journal:  Biomed Mater Eng       Date:  2002       Impact factor: 1.300

3.  Small-gauge vitrectomy in traumatic retinal detachment.

Authors:  Rita Ehrlich; Philip Polkinghorne
Journal:  Clin Exp Ophthalmol       Date:  2011-07       Impact factor: 4.207

4.  Surface conditions of Nitinol wires, tubing, and as-cast alloys. The effect of chemical etching, aging in boiling water, and heat treatment.

Authors:  S A Shabalovskaya; J Anderegg; F Laab; P A Thiel; G Rondelli
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2003-04-15       Impact factor: 3.368

5.  Facilitated coronary anastomosis using a nitinol U-Clip device: bovine model.

Authors:  A C Hill; T P Maroney; R Virmani
Journal:  J Thorac Cardiovasc Surg       Date:  2001-05       Impact factor: 5.209

6.  Combined 23-gauge microincisonal vitrectomy surgery and phacoemulsification with AcrySof toric intraocular lens implantation: a comparative study.

Authors:  D H Park; J P Shin; S Y Kim
Journal:  Eye (Lond)       Date:  2011-07-15       Impact factor: 3.775

7.  Minimally invasive combined aortic and mitral valve replacement.

Authors:  Pragnesh Joshi; Chirag Doshi; Mahesh Vinchurkar; Rajesh Thosani; Prashant Sagar; Vijay Mahajan
Journal:  Heart Lung Circ       Date:  2010-12-10       Impact factor: 2.975

8.  Repeatability and reproducibility of anterior chamber volume measurements using 3-dimensional corneal and anterior segment optical coherence tomography.

Authors:  Shinichi Fukuda; Keisuke Kawana; Yoshiaki Yasuno; Tetsuro Oshika
Journal:  J Cataract Refract Surg       Date:  2011-03       Impact factor: 3.351

9.  Operating microscope light-induced phototoxic maculopathy after transscleral sutured posterior chamber intraocular lens implantation.

Authors:  Eui Yong Kweon; Min Ahn; Dong Wook Lee; In Cheon You; Min Jung Kim; Nam Chun Cho
Journal:  Retina       Date:  2009 Nov-Dec       Impact factor: 4.256

10.  Comparison of bimanual and micro-coaxial phacoemulsification with torsional ultrasound.

Authors:  Yujuan Wang; Yuanling Xia; Xialin Liu; Danying Zheng; Lixia Luo; Yizhi Liu
Journal:  Acta Ophthalmol       Date:  2010-02-16       Impact factor: 3.761

View more
  3 in total

1.  Chronic Granulomatous Inflammation after CyPass® Implantation.

Authors:  Kristina Joana Schoelles; Alexandra Anton; Claudia Auw-Haedrich
Journal:  Ocul Oncol Pathol       Date:  2020-02-11

2.  Biocompatibility and feasibility of VisiPlate, a novel ultrathin, multichannel glaucoma drainage device.

Authors:  Brandon W Kao; Elana Meer; Thomas A Barbolt; Richard A Lewis; Iqbal Ike Ahmed; Vivian Lee; Samuel M Nicaise; Georgia Griggs; Eydie G Miller-Ellis
Journal:  J Mater Sci Mater Med       Date:  2021-11-24       Impact factor: 3.896

Review 3.  Hydrus microstent implantation for surgical management of glaucoma: a review of design, efficacy and safety.

Authors:  Saba Samet; Jeb A Ong; Iqbal Ike K Ahmed
Journal:  Eye Vis (Lond)       Date:  2019-10-22
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.