AIMS: Device-related infections (DRI) are not well understood in patients implanted with a cardiac resynchronization therapy (CRT) device. The aims of this study were: (i) to evaluate the prevalence of CRT DRI; (ii) to establish the factors predictive of CRT DRI. METHODS AND
RESULTS: Between January 2001 and May 2007, CRT implantation was performed in 303 patients (247 men, 82%). The mean follow-up was 31 +/- 19 months. Population characteristics were a mean age of 70 +/- 10 years old; 56 female; aetiology includes (202 dilated and 101 ischaemic cardiomyopathy); NYHA class 3.2 +/- 0.3; LVEF (26 +/- 6%), and a QRS width of 171 +/- 31 ms. Thirteen patients developed a DRI: endocarditis in four, pocket erosion in three, pocket abscess in five, and septicaemia in one. The prevalence of DRI was 4.3%. By univariate analysis, predictive factors of DRI were: procedure time (skin to skin: median of 85 vs. 57.5 min; P = 0.03), re-intervention (54 vs. 6.5%; P < 0.0001), haematoma (31 vs. 8.6% P = 0.01), lead dislodgement (23 vs. 6.2%; P = 0.03), dialysis (23.1 vs. 1.72%; P = 0.003), and procedure type [CRT-ICD (8.6%) vs. CRT PM (1.6%) or system up-grade (1.5%); P = 0.03]. Significant correlations were found between re-intervention and lead dislodgement (r = 0.8; P < 0.001), haematoma (r = 0.2; P < 0.001). Four independent predictive factors of DRI were identified as procedure time (P = 0.002); dialysis (P = 0.0001); re-intervention (P = 0.006), and procedure type (CRT-ICD vs. other procedures; P = 0.01).
CONCLUSION: This study found that the prevalence of CRT DRI is close to 4.3% at 2.6 years (1.7% per year incidence). Four independent predictive factors of infections were identified including re-intervention, procedure time, dialysis, and primo CRT-ICD implantation. These parameters should be part of the risk-benefit evaluation in patients selected for CRT implantation.

Entities:  

Mesh:

Year:  2009        PMID: 19875388     DOI: 10.1093/eurheartj/ehp421

Source DB:  PubMed          Journal:  Eur Heart J        ISSN: 0195-668X            Impact factor:   29.983


  36 in total

1.  A meta-analysis of antibacterial envelope use in prevention of cardiovascular implantable electronic device infection.

Authors:  Sajid Ali; Yousuf Kanjwal; Steven R Bruhl; Mohammed Alo; Mohammed Taleb; Syed S Ali; Ameer Kabour; Owais Khawaja
Journal:  Ther Adv Infect Dis       Date:  2017-05-01

Review 2.  18F-FDG-PET/CT Imaging to Diagnose Septic Emboli and Mycotic Aneurysms in Patients with Endocarditis and Cardiac Device Infections.

Authors:  Nidaa Mikail; Khadija Benali; Besma Mahida; Jonathan Vigne; Fabien Hyafil; François Rouzet; Dominique Le Guludec
Journal:  Curr Cardiol Rep       Date:  2018-03-06       Impact factor: 2.931

3.  Repeated procedures at the generator pocket are a determinant of implantable cardioverter-defibrillator infection.

Authors:  Eduardo Arana-Rueda; Alonso Pedrote; Manuel Frutos-López; Juan Acosta; Beatriz Jauregui; Lorena García-Riesco; Álvaro Arce-León; Federico Gómez-Pulido; Juan A Sánchez-Brotons; Encarnación Gutiérrez-Carretero; Arístides de Alarcón-González
Journal:  Clin Cardiol       Date:  2017-06-21       Impact factor: 2.882

4.  [Technical innovations and limitation in cardiac electrotherapy].

Authors:  H-H Minden
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2011-03

Review 5.  The changing landscape of cardiac pacing.

Authors:  S Serge Barold; Carsten W Israel
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2015-03

Review 6.  The role of biventricular pacing in the prevention and therapy of pacemaker-induced cardiomyopathy.

Authors:  Maya Guglin; S Serge Barold
Journal:  Ann Noninvasive Electrocardiol       Date:  2015-01-06       Impact factor: 1.468

7.  The Diagnosis and Treatment of Pacemaker-Associated Infection.

Authors:  Michael Döring; Sergio Richter; Gerhard Hindricks
Journal:  Dtsch Arztebl Int       Date:  2018-06-29       Impact factor: 5.594

8.  Cardiac Implantable Electronic Device Infection in Patients at Risk.

Authors:  Khaldoun G Tarakji; Christopher R Ellis; Pascal Defaye; Charles Kennergren
Journal:  Arrhythm Electrophysiol Rev       Date:  2016-05

Review 9.  Sex differences in device therapy for heart failure: utilization, outcomes, and adverse events.

Authors:  Naomi D Herz; Joseph Engeda; Robbert Zusterzeel; William E Sanders; Kathryn M O'Callaghan; David G Strauss; Samantha B Jacobs; Kimberly A Selzman; Ileana L Piña; Daniel A Caños
Journal:  J Womens Health (Larchmt)       Date:  2015-03-20       Impact factor: 2.681

10.  New Insights into Predictors of Cardiac Implantable Electronic Device Infection.

Authors:  Hossein Sadeghi; Abolfath Alizadehdiz; Amirfarjam Fazelifar; Zahra Emkanjoo; Majid Haghjoo
Journal:  Tex Heart Inst J       Date:  2018-06-01
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.