Literature DB >> 19838116

Using evoked compound action potentials to assess activation of electrodes and predict C-levels in the Tempo+ cochlear implant speech processor.

Isaac Alvarez1, Angel de la Torre, Manuel Sainz, Cristina Roldán, Hansjoerg Schoesser, Philipp Spitzer.   

Abstract

OBJECTIVE: In this study, we analyze how electrically evoked compound action potential (ECAP) responses can be used to assess whether electrodes should be activated in the map and to estimate C levels in the Med-El Tempo+ Cochlear Implant Speech Processor.
DESIGN: ECAP thresholds were measured using the ECAP Recording System of the Pulsar CI implant. Twenty-one postlingually and 28 prelingually deafened patients participated in this study. The relationship between ECAP responses and the activation of electrodes was analyzed. Because an error in the estimation of T levels (behavioral thresholds) has less effect on hearing quality than an error in the estimation of C levels in the Tempo+ cochlear implant speech processor (maximum comfort levels), correlation and regression analyses were performed between ECAP thresholds and C levels.
RESULTS: The observation of an evoked potential generally implied that the electrode was activated because only 3.5% of electrodes that yielded measurable evoked responses were deactivated, because of collateral stimulations or an unpleasant hearing sensation. In contrast, the absence of an evoked potential did not imply that an electrode should be deactivated, because 20% of these electrodes provided a useful auditory sensation. ECAP responses did not predict the absolute behavioral comfort levels because of the excessive error between behavioral C levels and those derived from ECAP thresholds (the mean relative error is 43.78%). However, by applying a normalization procedure, ECAP measurements allowed the C-level profile to be predicted with a mean relative error of 6%; that is, they provided useful data to determine the C level of each electrode relative to the average C level of the patient.
CONCLUSIONS: ECAP is a reliable and an useful objective measurement that can assist in the fitting of the Tempo+ cochlear implant speech processor. From results presented in this work, a protocol is proposed for fitting this cochlear implant system. This protocol facilitates appropriate cochlear implant fitting, particularly for children or uncooperative patients.

Entities:  

Mesh:

Year:  2010        PMID: 19838116     DOI: 10.1097/AUD.0b013e3181bdb88f

Source DB:  PubMed          Journal:  Ear Hear        ISSN: 0196-0202            Impact factor:   3.570


  8 in total

Review 1.  [Intra- and postoperative electrophysiological diagnostics].

Authors:  T Wesarg; S Arndt; A Aschendorff; R Laszig; R Beck; L Jung; S Zirn
Journal:  HNO       Date:  2017-04       Impact factor: 1.284

2.  Maps created using a new objective procedure (C-NRT) correlate with behavioral, loudness-balanced maps: a study in adult cochlear implant users.

Authors:  Alessandro Scorpecci; Alessandra D'Elia; Paolo Malerba; Italo Cantore; Patrizia Consolino; Franco Trabalzini; Gaetano Paludetti; Nicola Quaranta
Journal:  Eur Arch Otorhinolaryngol       Date:  2016-05-30       Impact factor: 2.503

3.  The transtympanic promontory stimulation test in patients with auditory deprivation: correlations with electrical dynamics of cochlear implant and speech perception.

Authors:  Mohammad Alfelasi; Jean Pierre Piron; Caroline Mathiolon; Nadjmah Lenel; Michel Mondain; Alain Uziel; Frederic Venail
Journal:  Eur Arch Otorhinolaryngol       Date:  2012-09-30       Impact factor: 2.503

4.  [Objective measures for setting the processors of cochlear implant systems : Use of discrimination functions and consideration of electrode profiles].

Authors:  S Hoth; I Herisanu; M Praetorius
Journal:  HNO       Date:  2016-12       Impact factor: 1.284

5.  Long-term evolution of the electrical stimulation levels for cochlear implant patients.

Authors:  Jose Luis Vargas; Manuel Sainz; Cristina Roldan; Isaac Alvarez; Angel de la Torre
Journal:  Clin Exp Otorhinolaryngol       Date:  2012-11-13       Impact factor: 3.372

6.  Fine-grain recordings of the electrically evoked compound action potential amplitude growth function in cochlear implant recipients.

Authors:  Lutz Gärtner; Thomas Lenarz; Andreas Büchner
Journal:  Biomed Eng Online       Date:  2018-10-19       Impact factor: 2.819

7.  ARTFit-A Quick and Reliable Tool for Performing Initial Fittings in Users of MED-EL Cochlear Implants.

Authors:  Lutz Gärtner; Timo Bräcker; Mathias Kals; Richard T Penninger; Mareike Billinger-Finke; Thomas Lenarz; Andreas Büchner
Journal:  Life (Basel)       Date:  2022-02-11

8.  Intracorporeal Cortical Telemetry as a Step to Automatic Closed-Loop EEG-Based CI Fitting: A Proof of Concept.

Authors:  Andy J Beynon; Bart M Luijten; Emmanuel A M Mylanus
Journal:  Audiol Res       Date:  2021-12-13
  8 in total