Samer Sourial1, Maritha Marcusson-Ståhl, Karin Cederbrant.
Abstract
The sensitivity of different renal regions to xenobiotics requires the development of a multiplex immunoassay for the simultaneous analysis of kidney biomarkers. Calbindin D28K is a distal tubule-specific protein that can be detected in urine under pathological conditions. In this study, a pair of anti-calbindin D28K antibodies was used in an immunoassay for the detection of calbindin D28K expression in rat and human kidney and urine. Comparative analysis of the immunoassay was performed on the Meso Scale Development (MSD) and Luminex platforms. Analysis on both platforms detected calbindin D28K concentrations between 100 ng/mL and 100 pg/mL. Luminex detected 10-fold the amount of calbindin D28K in samples analyzed as compared to MSD, whereas calbindin D28K level in rat and human urine was below detection limit in both platforms. The application of the immunoassays described herein may be useful in toxicological and pathological studies of distal tubular damage in rats and human.Entities:
Mesh:
Substances:
Year: 2009 PMID: 19826628 PMCID: PMC2760132 DOI: 10.1155/2009/187426
Source DB: PubMed Journal: J Biomed Biotechnol ISSN: 1110-7243
Figure 1(a) Calbindin D28K standard curve analysis on MSD. (b) Luminex analysis of calbindin D28K; with standard curve indicating the fluorescence intensity (FI) at the different concentrations.
Linearity analysis of rat cortical kidney homgenate dilutions.
| MSD | Luminex | |||||
|---|---|---|---|---|---|---|
| Dilution factor | Measured pg/mL | CV%+ | %Mean* | Measured pg/mL | CV%+ | %Mean* |
| 100 | 5554 | 1.2 | 131 | 49762 | 0.1 | 128 |
| 200 | 2410 | 1.9 | 114 | 22802 | 5.1 | 117 |
| 400 | 1233 | 11 | 116 | 11568 | 1.2 | 119 |
| 800 | 424 | 12 | 80 | 5421 | 8.0 | 112 |
| 1600 | 247 | 6.7 | 93 | 2709 | 5.2 | 111 |
| 3200 | 110 | 16 | 83 | 1120 | 3.7 | 92 |
| 6400 | 55 | 14 | 83 | 422 | 12 | 69 |
| 12800 | — | 155 | 6.2 | 51 | ||
*Calculated concentration (dilution factor × measured concentration) as a percentage of the average of the calculated concentration from all dilutions. The average calculated concentration of rat cortical kidney homogenate analyzed by MSD and Luminex was 424 ng/mL and 3.9 μg/mL, respectively.
+Coefficient of variation.
Precision analysis using rat cortical kidney homogenate.
| MSD | Luminex | |||
|---|---|---|---|---|
| Dilution | 1 : 400 | 1 : 4000 | 1 : 400 | 1 : 4000 |
| Mean | 945 | 87 | 14167 | 1041 |
| SD* | 97 | 17 | 723 | 103 |
| CV+ | 10% | 20% | 5% | 10% |
*Standard of deviation.
+Coefficient of variation.
Interassay variation analysis of rat cortical kidney homogenate.
| Assay | MSD (ng/mL) | Luminex ( |
|---|---|---|
| 1 | 424 | 3.9 |
| 2 | 407 | 3.1 |
| 3 | 283 | 3.3 |
| SD* | 76 | 0.4 |
*Standard of deviation.
MSD specificity analysis of human kidney homogenate.
| Region | Concentration (ng/mL) |
|---|---|
| Cortex | 64 |
| Medulla | 3 |
| Papilla | — |