Literature DB >> 19822467

An online self-tunable method to denoise CGM sensor data.

Andrea Facchinetti1, Giovanni Sparacino, Claudio Cobelli.   

Abstract

Continuous glucose monitoring (CGM) devices can be very useful in diabetes management. Unfortunately, their use in online applications, e.g., for hypo/hyperalert generation, is made difficult by random noise measurement. Remarkably, the SNR of CGM data varies with the sensor and with the individual. As a consequence, approaches in which filter parameters are not allowed to adapt to the current SNR are likely to be suboptimal. In this paper, we present a new online methodology to reduce noise in CGM signals by a Kalman filter (KF), whose unknown parameters are adjusted in a given individual by a stochastically based smoothing criterion exploiting data of a burn-in interval. The performance of the new KF approach is quantitatively assessed on Monte Carlo simulations and 24 real CGM datasets. Our results are compared with those obtained by a moving-average (MA) filtering approach with fixed parameters currently in use in likely all commercial CGM devices. Results show that the new KF approach performs much better than MA. For instance, on real data, for comparable signal denoising, the delay introduced by KF is about 35% less than that obtained by MA.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19822467     DOI: 10.1109/TBME.2009.2033264

Source DB:  PubMed          Journal:  IEEE Trans Biomed Eng        ISSN: 0018-9294            Impact factor:   4.538


  18 in total

1.  AP@home: a novel European approach to bring the artificial pancreas home.

Authors:  Lutz Heinemann; Carsten Benesch; J Hans DeVries
Journal:  J Diabetes Sci Technol       Date:  2011-11-01

2.  Continuous glucose monitoring: real-time algorithms for calibration, filtering, and alarms.

Authors:  B Wayne Bequette
Journal:  J Diabetes Sci Technol       Date:  2010-03-01

3.  Physical activity measured by physical activity monitoring system correlates with glucose trends reconstructed from continuous glucose monitoring.

Authors:  Chiara Zecchin; Andrea Facchinetti; Giovanni Sparacino; Chiara Dalla Man; Chinmay Manohar; James A Levine; Ananda Basu; Yogish C Kudva; Claudio Cobelli
Journal:  Diabetes Technol Ther       Date:  2013-08-14       Impact factor: 6.118

4.  Signal processing algorithms implementing the "smart sensor" concept to improve continuous glucose monitoring in diabetes.

Authors:  Andrea Facchinetti; Giovanni Sparacino; Claudio Cobelli
Journal:  J Diabetes Sci Technol       Date:  2013-09-01

5.  Hypoglycemia early alarm systems based on recursive autoregressive partial least squares models.

Authors:  Elif Seyma Bayrak; Kamuran Turksoy; Ali Cinar; Lauretta Quinn; Elizabeth Littlejohn; Derrick Rollins
Journal:  J Diabetes Sci Technol       Date:  2013-01-01

6.  Inpatient studies of a Kalman-filter-based predictive pump shutoff algorithm.

Authors:  Fraser Cameron; Darrell M Wilson; Bruce A Buckingham; Hasmik Arzumanyan; Paula Clinton; H Peter Chase; John Lum; David M Maahs; Peter M Calhoun; B Wayne Bequette
Journal:  J Diabetes Sci Technol       Date:  2012-09-01

7.  Real-time state estimation and long-term model adaptation: a two-sided approach toward personalized diagnosis of glucose and insulin levels.

Authors:  Claudia Eberle; Christoph Ament
Journal:  J Diabetes Sci Technol       Date:  2012-09-01

8.  Processing Diabetes Mellitus Composite Events in MAGPIE.

Authors:  Albert Brugués; Stefano Bromuri; Michael Barry; Óscar Jiménez Del Toro; Maciej R Mazurkiewicz; Przemyslaw Kardas; Josep Pegueroles; Michael Schumacher
Journal:  J Med Syst       Date:  2015-11-21       Impact factor: 4.460

9.  Comparison between one-point calibration and two-point calibration approaches in a continuous glucose monitoring algorithm.

Authors:  Zeinab Mahmoudi; Mette Dencker Johansen; Jens Sandahl Christiansen; Ole Hejlesen
Journal:  J Diabetes Sci Technol       Date:  2014-04-21

10.  Personalized State-space Modeling of Glucose Dynamics for Type 1 Diabetes Using Continuously Monitored Glucose, Insulin Dose, and Meal Intake: An Extended Kalman Filter Approach.

Authors:  Qian Wang; Peter Molenaar; Saurabh Harsh; Kenneth Freeman; Jinyu Xie; Carol Gold; Mike Rovine; Jan Ulbrecht
Journal:  J Diabetes Sci Technol       Date:  2014-03-24
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.