A post-beamforming 2-D pseudoinverse filter for coarsely sampled ultrasound arrays.

Yayun Wan1, Emad S Ebbini.   

Abstract

Beamforming artifacts due to coarse discretization of imaging apertures represent a significant barrier against the use of array probes in high-frequency applications. Nyquist sampling of array apertures dictates center-to-center spacing of lambda/2 for elimination of grating lobes in the array pattern. However, this requirement is hard to achieve using current transducer technologies, even at the lower end of high-frequency ultrasonic imaging (in the range 25-35 MHz). In this paper, we present a new design approach for 2-D regularized pseudoinverse (PIO) filters suitable for restoring imaging contrast in systems employing coarsely sampled arrays. The approach is based on a discretized 2-D imaging model for linear arrays assuming scattering from a Cartesian grid in the imaging field of view (FOV). We show that the discretized imaging operator can be represented with a block Toeplitz matrix with the blocks themselves being Toeplitz. With sufficiently large grid size in the axial and lateral directions, it is possible to replace this Toeplitz-block block Toeplitz (TBBT) operator with its circulant-block block circulant (CBBC) equivalent. This leads to a computationally efficient implementation of the regularized pseudoinverse filtering approach using the 2-D fast Fourier transform (FFT). The derivation of the filtering equation is shown in detail and the regularization procedure is fully described. Using FIELD, we present simulation data to show the 2-D point-spread functions (PSFs) for imaging systems employing linear arrays with fine and coarse sampling of the imaging aperture. PSFs are also computed for a coarsely sampled array with different levels of regularization to demonstrate the tradeoff between contrast and spatial resolution. These results demonstrate the well-behaved nature of the PSF with the variation in a single regularization parameter. Specifically, the 6 dB axial and lateral dimensions of the PSF increase gradually with increasing value of the regularization parameter. On the other hand, the peak grating lobe level decreases gradually with increasing value of the regularization parameter. The results are supported by image reconstructions from a simulated cyst phantom obtained using finely and coarsely sampled apertures with and without the application of the regularized 2-D PIO.

Entities:  

Mesh:

Year:  2009        PMID: 19811992      PMCID: PMC3150138          DOI: 10.1109/TUFFC.2009.1265

Source DB:  PubMed          Journal:  IEEE Trans Ultrason Ferroelectr Freq Control        ISSN: 0885-3010            Impact factor:   2.725


  21 in total

Review 1.  Intravascular ultrasound: novel pathophysiological insights and current clinical applications.

Authors:  S E Nissen; P Yock
Journal:  Circulation       Date:  2001-01-30       Impact factor: 29.690

Review 2.  Advances in ultrasound biomicroscopy.

Authors:  F S Foster; C J Pavlin; K A Harasiewicz; D A Christopher; D H Turnbull
Journal:  Ultrasound Med Biol       Date:  2000-01       Impact factor: 2.998

3.  Interdigital pair bonding for high frequency (20-50 MHz) ultrasonic composite transducers.

Authors:  R Liu; K A Harasiewicz; F S Foster
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2001-01       Impact factor: 2.725

4.  Clinical use of ultrasound biomicroscopy.

Authors:  C J Pavlin; K Harasiewicz; M D Sherar; F S Foster
Journal:  Ophthalmology       Date:  1991-03       Impact factor: 12.079

Review 5.  Use of modulated excitation signals in medical ultrasound. Part I: Basic concepts and expected benefits.

Authors:  Thanassis Misaridis; Jørgen Arendt Jensen
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2005-02       Impact factor: 2.725

6.  3-D ultrasound imaging using a forward-looking CMUT ring array for intravascular/intracardiac applications.

Authors:  David T Yeh; Omer Oralkan; Ira O Wygant; Matthew O'Donnell; Butrus T Khuri-Yakub
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2006-06       Impact factor: 2.725

7.  Pulse elongation and deconvolution filtering for medical ultrasonic imaging.

Authors:  B Haider; P A Lewin; K E Thomenius
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  1998       Impact factor: 2.725

8.  Arterial imaging with a new forward-viewing intravascular ultrasound catheter, I. Initial studies.

Authors:  J L Evans; K H Ng; M J Vonesh; B L Kramer; S N Meyers; T A Mills; B J Kane; W N Aldrich; Y T Jang; P G Yock
Journal:  Circulation       Date:  1994-02       Impact factor: 29.690

9.  Assessment of articular cartilage and subchondral bone: subtle and progressive changes in experimental osteoarthritis using 50 MHz echography in vitro.

Authors:  A Saïed; E Chérin; H Gaucher; P Laugier; P Gillet; J Floquet; P Netter; G Berger
Journal:  J Bone Miner Res       Date:  1997-09       Impact factor: 6.741

10.  Imaging of immature articular cartilage using ultrasound backscatter microscopy at 50 MHz.

Authors:  H K Kim; P S Babyn; K A Harasiewicz; H K Gahunia; K P Pritzker; F S Foster
Journal:  J Orthop Res       Date:  1995-11       Impact factor: 3.494

View more
  2 in total

1.  B-scan based acoustic source reconstruction for magnetoacoustic tomography with magnetic induction (MAT-MI).

Authors:  Leo Mariappan; Xu Li; Bin He
Journal:  IEEE Trans Biomed Eng       Date:  2010-11-22       Impact factor: 4.538

2.  Magnetoacoustic tomography with magnetic induction: bioimepedance reconstruction through vector source imaging.

Authors:  Leo Mariappan; Bin He
Journal:  IEEE Trans Med Imaging       Date:  2013-01-11       Impact factor: 10.048

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.