Literature DB >> 19763217

The Changes in Range of Motion after a Lumbar Spinal Arthroplasty with Charité in the Human Cadaveric Spine under Physiologic Compressive Follower Preload : A Comparative Study between Load Control Protocol and Hybrid Protocol.

Se-Hoon Kim1, Ung-Kyu Chang, Jae-Chil Chang, Kwon-Soo Chun, T Jesse Lim, Daniel H Kim.   

Abstract

OBJECTIVE: To compare two testing protocols for evaluating range of motion (ROM) changes in the preloaded cadaveric spines implanted with a mobile core type Charité lumbar artificial disc.
METHODS: Using five human cadaveric lumbosacral spines (L2-S2), baseline ROMs were measured with a bending moment of 8 Nm for all motion modes (flexion/extension, lateral bending, and axial rotation) in intact spine. The ROM was tracked using a video-based motion-capturing system. After the Charité disc was implanted at the L4-L5 level, the measurement was repeated using two different methods : 1) loading up to 8 Nm with the compressive follower preload as in testing the intact spine (Load control protocol), 2) loading in displacement control until the total ROM of L2-S2 matches that when the intact spine was loaded under load control (Hybrid protocol). The comparison between the data of each protocol was performed.
RESULTS: The ROMs of the L4-L5 arthroplasty level were increased in all test modalities (p < 0.05 in bending and rotation) under both load and hybrid protocols. At the adjacent segments, the ROMs were increased in all modes except flexion under load control protocol. Under hybrid protocol, the adjacent segments demonstrated decreased ROMs in all modalities except extension at the inferior segment. Statistical significance between load and hybrid protocols was observed during bending and rotation at the operative and adjacent levels (p < 0.05).
CONCLUSION: In hybrid protocol, the Charité disc provided a relatively better restoration of ROM, than in the load control protocol, reproducing clinical observations in terms of motion following surgery.

Entities:  

Keywords:  Charité™; Follower preload; Hybrid protocol; Load control protocol; Lumbar spinal arthroplasty; Range of motion

Year:  2009        PMID: 19763217      PMCID: PMC2744024          DOI: 10.3340/jkns.2009.46.2.144

Source DB:  PubMed          Journal:  J Korean Neurosurg Soc        ISSN: 1225-8245


  37 in total

1.  Effects of charité artificial disc on the implanted and adjacent spinal segments mechanics using a hybrid testing protocol.

Authors:  Vijay K Goel; Jonathan N Grauer; Tushar Ch Patel; Ashok Biyani; Koichi Sairyo; Srilakshmi Vishnubhotla; Aaron Matyas; Ian Cowgill; Miranda Shaw; Rebecca Long; David Dick; Manohar M Panjabi; Hassan Serhan
Journal:  Spine (Phila Pa 1976)       Date:  2005-12-15       Impact factor: 3.468

2.  Biomechanical study of lumbar spinal arthroplasty with a semi-constrained artificial disc (activ L) in the human cadaveric spine.

Authors:  Sung-Kon Ha; Se-Hoon Kim; Daniel H Kim; Jung-Yul Park; Dong-Jun Lim; Sang-Kook Lee
Journal:  J Korean Neurosurg Soc       Date:  2009-03-31

3.  Design of an intervertebral disc prosthesis.

Authors:  T P Hedman; J P Kostuik; G R Fernie; W G Hellier
Journal:  Spine (Phila Pa 1976)       Date:  1991-06       Impact factor: 3.468

4.  Intradiscal pressure measurements above an instrumented fusion. A cadaveric study.

Authors:  S L Weinhoffer; R D Guyer; M Herbert; S L Griffith
Journal:  Spine (Phila Pa 1976)       Date:  1995-03-01       Impact factor: 3.468

5.  Risk factors for adjacent-segment failure following lumbar fixation with rigid instrumentation for degenerative instability.

Authors:  S Etebar; D W Cahill
Journal:  J Neurosurg       Date:  1999-04       Impact factor: 5.115

6.  Artificial disc replacement with the modular type SB Charité III: 2-year results in 50 prospectively studied patients.

Authors:  W S Zeegers; L M Bohnen; M Laaper; M J Verhaegen
Journal:  Eur Spine J       Date:  1999       Impact factor: 3.134

7.  The transition zone above a lumbosacral fusion.

Authors:  M F Hambly; L L Wiltse; N Raghavan; G Schneiderman; C Koenig
Journal:  Spine (Phila Pa 1976)       Date:  1998-08-15       Impact factor: 3.468

8.  Neurological complications of lumbar artificial disc replacement and comparison of clinical results with those related to lumbar arthrodesis in the literature: results of a multicenter, prospective, randomized investigational device exemption study of Charité intervertebral disc. Invited submission from the Joint Section Meeting on Disorders of the Spine and Peripheral Nerves, March 2004.

Authors:  Fred H Geisler; Scott L Blumenthal; Richard D Guyer; Paul C McAfee; John J Regan; J Patrick Johnson; Bradford Mullin
Journal:  J Neurosurg Spine       Date:  2004-09

9.  L5-S1 segment survivorship and clinical outcome analysis after L4-L5 isolated fusion.

Authors:  Gary Ghiselli; Jeffrey C Wang; Wellington K Hsu; Edgar G Dawson
Journal:  Spine (Phila Pa 1976)       Date:  2003-06-15       Impact factor: 3.468

10.  Centrode characteristics of the lumbar spine as a function of segmental instability.

Authors:  S D Gertzbein; J Seligman; R Holtby; K W Chan; N Ogston; A Kapasouri; M Tile
Journal:  Clin Orthop Relat Res       Date:  1986-07       Impact factor: 4.176

View more
  5 in total

1.  Cervical disc prosthesis versus arthrodesis using one-level, hybrid and two-level constructs: an in vitro investigation.

Authors:  Cédric Barrey; Sophie Campana; Sylvain Persohn; Gilles Perrin; Wafa Skalli
Journal:  Eur Spine J       Date:  2011-08-11       Impact factor: 3.134

2.  Relevance of using a compressive preload in the cervical spine: an experimental and numerical simulating investigation.

Authors:  Cédric Barrey; Marc-Antoine Rousseau; Sylvain Persohn; Sophie Campana; Gilles Perrin; Wafa Skalli
Journal:  Eur J Orthop Surg Traumatol       Date:  2015-04-07

Review 3.  Meta-analyses comparing spine simulators with cadavers and finite element models by analysing range-of-motion data before and after lumbar total disc replacement.

Authors:  Tobias Bohn; Susanne A J Lang; Stephanie Roll; Helene Schrader; Matthias Pumberger; Karin Büttner-Janz
Journal:  J Adv Res       Date:  2020-06-23       Impact factor: 10.479

4.  We Need to Talk about Lumbar Total Disc Replacement.

Authors:  Stephen Beatty
Journal:  Int J Spine Surg       Date:  2018-08-03

5.  The current testing protocols for biomechanical evaluation of lumbar spinal implants in laboratory setting: a review of the literature.

Authors:  Sabrina A Gonzalez-Blohm; James J Doulgeris; William E Lee; Thomas M Shea; Kamran Aghayev; Frank D Vrionis
Journal:  Biomed Res Int       Date:  2015-02-15       Impact factor: 3.411

  5 in total