Evaluation of the Perkins handheld applanation tonometer in the measurement of intraocular pressure in dogs and cats.

Silvia Franco Andrade1, Tatiana Cremonezi, Cristiane Aparecida Miranda Zachi, Caroline Ferreira Lonchiati, Juliana Dalarrosa Amatuzzi, Keila Priscilla Sakamoto, Paulo Augusto de Arruda Mello.   

Abstract

OBJECTIVE: To evaluate and to validate the accuracy of the Perkins handheld applanation tonometer in the measurement of IOP in dogs and cats. ANIMALS: Twenty eyes from 10 dogs and 10 cats immediately after sacrifice were used for the postmortem study and 20 eyes from 10 clinically normal and anesthetized dogs and cats were used for the in vivo study. Both eyes of 20 conscious dogs and cats were also evaluated. PROCEDURE: Readings of IOP postmortem and in vivo were taken using manometry (measured with a mercury column manometer) and tonometry (measured with a Perkins handheld applanation tonometer). The IOP measurement with Perkins tonometer in anesthetized and conscious dogs and cats was accomplished by instillation of proxymetacaine 0.5% and of 1% fluorescein eye drops.
RESULTS: The correlation coefficient (r(2)) between the manometry and the Perkins tonometer were 0.982 (dogs) and 0.988 (cats), and the corresponding linear regression equation were y = 0.0893x + 0.1105 (dogs) and y = 0.0899x + 0.1145 (cats) in the postmortem study. The mean IOP readings with the Perkins tonometer after calibration curve correction were 14.9 +/- 1.6 mmHg (range 12.2-17.2 mmHg) in conscious dogs, and were 15.1 +/- 1.7 mmHg (range 12.1-18.7 mmHg) in conscious cats.
CONCLUSION: There was an excellent correlation between the IOP values obtained from direct ocular manometry and the Perkinstonometer in dogs and cats. The Perkins handheld tonometer could be in the future a new alternative for the diagnosis of glaucoma in veterinary ophthalmology.

Entities:  

Mesh:

Year:  2009        PMID: 19751486     DOI: 10.1111/j.1463-5224.2009.00702.x

Source DB:  PubMed          Journal:  Vet Ophthalmol        ISSN: 1463-5216            Impact factor:   1.644


  5 in total

Review 1.  Feline glaucoma--a comprehensive review.

Authors:  Gillian J McLellan; Paul E Miller
Journal:  Vet Ophthalmol       Date:  2011-09       Impact factor: 1.644

2.  Intraocular pressure measurements in cattle, sheep, and goats with 2 different types of tonometers.

Authors:  Nina Peche; Johanna Corinna Eule
Journal:  Can J Vet Res       Date:  2018-07       Impact factor: 1.310

3.  Evaluation of the Perkins handheld applanation tonometer in horses and cattle.

Authors:  Silvia Franco Andrade; Daniel Silva Kupper; Luiz Fernando Rodrigues Pinho; Elizabeth Cunha Franco; Marcus Vinícius Felix Fabri Prataviera; Rodrigo Rolim Duarte; José Ricardo Cecílio Junqueira
Journal:  J Vet Sci       Date:  2011-06       Impact factor: 1.672

4.  Comparison among TonoVet, TonoVet Plus, Tono-Pen Avia Vet, and Kowa HA-2 portable tonometers for measuring intraocular pressure in dogs.

Authors:  João Victor Goulart Consoni Passareli; Felipe Franco Nascimento; Giovana José Garcia Estanho; Claudia Lizandra Ricci; Glaucia Prada Kanashiro; Rogério Giuffrida; Silvia Franco Andrade
Journal:  Vet World       Date:  2021-09-21

5.  Applanation tonometry: a comparison of the Perkins handheld and Goldmann slit lamp-mounted methods.

Authors:  R Arora; H Bellamy; Mw Austin
Journal:  Clin Ophthalmol       Date:  2014-03-26
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.