Thrombus size and Doppler embolic signal intensity.

Matthew J Martin1, Emma M L Chung, Kumar V Ramnarine, Alison H Goodall, A Ross Naylor, David H Evans.   

Abstract

BACKGROUND: Migration of thrombus through the cerebral arteries is a common cause of stroke. Thrombus emboli can be detected non-invasively using Doppler ultrasound, but even where the embolus composition is known, there is currently no method for estimating the size of an embolus based on the returned ultrasound signal. Here we report the results of in vitro experiments investigating the relationship between size and embolic signal intensity for fresh thrombus emboli with a view to estimating the sizes of thrombi detected following carotid surgery.
METHOD: Thrombi were formed from whole blood using the 'Chandler loop' method under flow conditions similar to those associated with arterial thrombus formation in vivo. A total of 390 Doppler embolic signals were then measured from 37 pieces of thrombus circulated in a pulsatile closed-flow circuit. The dimensions of each of the thrombi were measured before and after circulation using an optical microscope. Relationships between thrombus size and embolic signal properties were then investigated using standard statistical methods with a view to size estimation of thrombi during clinical monitoring.
RESULTS: Although embolic signals generally became more intense with increasing thrombus size, strong oscillations due to resonance effects were observed. Pearson tests revealed strong positive correlations between embolus diameter, signal intensity and duration (r > 0.8, p < or = 0.01).
CONCLUSIONS: This study provides experimental evidence supporting theoretical predictions relating Doppler embolic signal intensity to thrombus size. In our discussion, we tentatively suggest how this information might be used to size emboli in clinical practice. Copyright 2009 S. Karger AG, Basel.

Entities:  

Mesh:

Year:  2009        PMID: 19713699      PMCID: PMC2756770          DOI: 10.1159/000235627

Source DB:  PubMed          Journal:  Cerebrovasc Dis        ISSN: 1015-9770            Impact factor:   2.762


  23 in total

1.  Experimental demonstration of noninvasive transskull adaptive focusing based on prior computed tomography scans.

Authors:  J F Aubry; M Tanter; M Pernot; J L Thomas; M Fink
Journal:  J Acoust Soc Am       Date:  2003-01       Impact factor: 1.840

Review 2.  Perioperative stroke.

Authors:  Magdy Selim
Journal:  N Engl J Med       Date:  2007-02-15       Impact factor: 91.245

3.  An ultrasonic detector for microgasemboli in a bloodflow line.

Authors:  J Lubbers; J W Van Den Berg
Journal:  Ultrasound Med Biol       Date:  1977       Impact factor: 2.998

4.  Characteristics of Doppler embolic signals observed following carotid endarterectomy.

Authors:  Emma M L Chung; Lingke Fan; A Ross Naylor; David H Evans
Journal:  Ultrasound Med Biol       Date:  2006-07       Impact factor: 2.998

5.  Pulse Doppler ultrasound detection, characterization and size estimation of emboli in flowing blood.

Authors:  M A Moehring; J R Klepper
Journal:  IEEE Trans Biomed Eng       Date:  1994-01       Impact factor: 4.538

6.  Sizing emboli in blood using pulse Doppler ultrasound--I: Verification of the EBR model.

Authors:  M A Moehring; J A Ritcey
Journal:  IEEE Trans Biomed Eng       Date:  1996-06       Impact factor: 4.538

7.  Timing of clinically significant microembolism after carotid endarterectomy.

Authors:  A L Abbott; C R Levi; J L Stork; G A Donnan; B R Chambers
Journal:  Cerebrovasc Dis       Date:  2007-01-30       Impact factor: 2.762

8.  Impact of clinical environment on embolus detection: a comparison of automated and manual detection of Doppler embolic signals.

Authors:  Emma M L Chung; Lingke Fan; A Ross Naylor; David H Evans
Journal:  Cerebrovasc Dis       Date:  2006-11-16       Impact factor: 2.762

9.  Enhanced detection of thromboemboli with the use of targeted microbubbles.

Authors:  Matthew J Martin; Emma M L Chung; Alison H Goodall; Alberto Della Martina; Kumar V Ramnarine; Lingke Fan; Sarah V Hainsworth; A Ross Naylor; David H Evans
Journal:  Stroke       Date:  2007-09-06       Impact factor: 7.914

10.  Limits of uncertainty in measured values of embolus-to-blood ratio due to Doppler sample volume shape and location.

Authors:  Emma L Angell; David H Evans
Journal:  Ultrasound Med Biol       Date:  2003-07       Impact factor: 2.998

View more
  3 in total

1.  Sizing gaseous emboli using Doppler embolic signal intensity.

Authors:  Caroline Banahan; James P Hague; David H Evans; Rizwan Patel; Kumar V Ramnarine; Emma M L Chung
Journal:  Ultrasound Med Biol       Date:  2012-03-06       Impact factor: 2.998

2.  Numerical prediction of thrombus risk in an anatomically dilated left ventricle: the effect of inflow cannula designs.

Authors:  Sam Liao; Benjamin Simpson; Michael Neidlin; Tim A S Kaufmann; Zhiyong Li; Maria A Woodruff; Shaun D Gregory
Journal:  Biomed Eng Online       Date:  2016-12-28       Impact factor: 2.819

3.  An in vitro comparison of embolus differentiation techniques for clinically significant macroemboli: dual-frequency technique versus frequency modulation method.

Authors:  Caroline Banahan; Zach Rogerson; Clément Rousseau; Kumar V Ramnarine; David H Evans; Emma M L Chung
Journal:  Ultrasound Med Biol       Date:  2014-09-11       Impact factor: 2.998

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.