Literature DB >> 19713648

Predictors and prognosis of stent fracture after sirolimus-eluting stent implantation.

Yasushi Ino1, Yasuaki Toyoda, Atsushi Tanaka, Shuuta Ishii, Yoshio Kusuyama, Takashi Kubo, Shigeho Takarada, Hironori Kitabata, Takashi Tanimoto, Masato Mizukoshi, Toshio Imanishi, Takashi Akasaka.   

Abstract

BACKGROUND: Stent fracture is a cause of in-stent restenosis (ISR) after sirolimus-eluting stent (SES) implantation, so this study investigated the incidence, predictors and prognosis of stent fracture. METHODS AND
RESULTS: The 273 consecutive patients (364 lesions) after SES implantation and who had 6-9 month' scheduled follow-up coronary angiography (CAG) were divided into groups with and without stent fracture. Deltaangle was defined as the difference in the angle in the target lesion between diastole and systole before the procedure. The incidence of stent fracture was 4.9% (18 of 364 lesions). Deltaangle in the target lesion was larger in the fracture group (28.3 +/-11.5 degrees vs 12.3 +/-9.0 degrees , P<0.0001). Independent predictors of stent fracture were Deltaangle and total stent length. The rates of binary restenosis and target lesion revascularization (TLR) were higher in the fracture group (33% vs 4.0%, P=0.0002 and 28% vs 3.5%, P=0.0007, respectively). There were no major adverse cardiac events (MACE) in the fracture group during a mean 24-month follow-up after follow-up CAG.
CONCLUSIONS: Predictors of stent fracture were Deltaangle and total stent length. Although stent fracture was associated with ISR and TLR, it was not associated with MACE during long-term follow-up.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19713648     DOI: 10.1253/circj.cj-09-0343

Source DB:  PubMed          Journal:  Circ J        ISSN: 1346-9843            Impact factor:   2.993


  10 in total

1.  Prevalence and prognosis of coronary stent gap detected by multi-detector CT: a follow-up study.

Authors:  Xinghua Zhang; Li Yang; Haiyue Ju; Fan Zhang; Jian Wu; Bin He; Yundai Chen
Journal:  Eur Radiol       Date:  2012-04-27       Impact factor: 5.315

2.  Application of a new three-dimensional method for the measurement of coronary stent angulation and comparison with conventional methods.

Authors:  Huaiyu Qiao; Luyue Gai; Minhan Wang; Bin He; Shuoyang Zhang; Xia Yang; Qinhua Jin; Zhiwei Guan
Journal:  Int J Cardiovasc Imaging       Date:  2012-11-21       Impact factor: 2.357

3.  Impact of the bifurcation angle on major cardiac events after cross-over single stent strategy in unprotected left main bifurcation lesions: 3-dimensional quantitative coronary angiographic analysis.

Authors:  Kisaki Amemiya; Takenori Domei; Masashi Iwabuchi; Shinichi Shirai; Kenji Ando; Masahiko Goya; Hiroyoshi Yokoi; Masakiyo Nobuyoshi
Journal:  Am J Cardiovasc Dis       Date:  2014-12-29

4.  Structural Mechanics Predictions Relating to Clinical Coronary Stent Fracture in a 5 Year Period in FDA MAUDE Database.

Authors:  Kay D Everett; Claire Conway; Gerard J Desany; Brian L Baker; Gilwoo Choi; Charles A Taylor; Elazer R Edelman
Journal:  Ann Biomed Eng       Date:  2015-10-14       Impact factor: 3.934

5.  Expansion of the Multi-Link Frontier™ coronary bifurcation stent: micro-computed tomographic assessment in human autopsy and porcine heart samples.

Authors:  Stefan Kralev; Benjamin Haag; Jens Spannenberger; Siegfried Lang; Marc A Brockmann; Soenke Bartling; Alexander Marx; Karl-Konstantin Haase; Martin Borggrefe; Tim Süselbeck
Journal:  PLoS One       Date:  2011-07-21       Impact factor: 3.240

6.  Coronary stent fracture in an octogenarian patient: from bad to worse.

Authors:  Andreas S Triantafyllis; Petros N Fountoulakis; Georgios Charalampidis; Konstantinos Kotinas; Pavlos Tsinivizov; Dimitrios Varvarousis; Andreas Giannakopoulos; Sofia Apollonatou; Areti Stefanidou; Stamatios Chatzopoulos; Dimitrios Tsiptsios; Konstantinos Tsamakis; Konstantinos Kyfnidis; Leonidas E Poulimenos
Journal:  J Geriatr Cardiol       Date:  2021-11-28       Impact factor: 3.327

7.  Incidence of coronary drug-eluting stent fracture: A systematic review and meta-analysis.

Authors:  Yang Chen; Dandan Li; Yanhui Liao; Xiongda Yao; Yuehua Ruan; Kai Zou; Hanhui Liao; Jingwen Ding; Hao Qin; Zuozhong Yu; Yuanbin Zhao; Longlong Hu; Renqiang Yang
Journal:  Front Cardiovasc Med       Date:  2022-08-23

8.  Inelastic Deformation of Coronary Stents: Two-Level Model.

Authors:  Pavel S Volegov; Nikita A Knyazev; Roman M Gerasimov; Vadim V Silberschmidt
Journal:  Materials (Basel)       Date:  2022-10-07       Impact factor: 3.748

9.  Stent fracture: how frequently is it recognized?

Authors:  Mohammed Khalil Mohsen; Awad Alqahtani; Jassim Al Suwaidi
Journal:  Heart Views       Date:  2013-04

Review 10.  Coronary stent fracture: a recently appreciated phenomenon with clinical relevance.

Authors:  Madjid Chinikar; Parham Sadeghipour
Journal:  Curr Cardiol Rev       Date:  2014-11
  10 in total