| Literature DB >> 19636390 |
Brendon O Watson1, Volodymyr Nikolenko, Rafael Yuste.
Abstract
Two-photon imaging has become a useful tool for optical monitoring of neural circuits, but it requires high laser power and serial scanning of each pixel in a sample. This results in slow imaging rates, limiting the measurements of fast signals such as neuronal activity. To improve the speed and signal-to-noise ratio of two-photon imaging, we introduce a simple modification of a two-photon microscope, using a diffractive optical element (DOE) which splits the laser beam into several beamlets that can simultaneously scan the sample. We demonstrate the advantages of DOE scanning by enhancing the speed and sensitivity of two-photon calcium imaging of action potentials in neurons from neocortical brain slices. DOE scanning can easily improve the detection of time-varying signals in two-photon and other non-linear microscopic techniques.Entities:
Keywords: beam-splitter; calcium; photostimulation; scanning; uncaging; video-rate
Year: 2009 PMID: 19636390 PMCID: PMC2715267 DOI: 10.3389/neuro.04.006.2009
Source DB: PubMed Journal: Front Neural Circuits ISSN: 1662-5110 Impact factor: 3.492
Figure 1Optical path and microscope design. (A) Conceptual schematic demonstrating beam splitting using a diffractive optical element (DOE). These beams are brought together onto the galvanometer mirrors which may scan and which transmit beams for exciting the sample underneath the objective. While the telescope does not change the collimation of the individual beamlets, it does change the size of the outgoing beamlets and so a pre-sizing telescope is used before the DOE-telescope complex (at left). Image collection can be performed using either a photomultiplier tube (PMT) or a CCD camera. (B) Detailed diagram of all elements of our optical and image acquisition setup
Titanium sapphire femtosecond source laser (Chameleon Ultra, Coherent Inc., Santa Clara, CA, USA)
Pockels cell is controlled by voltage input and regulates excitation laser intensity – in our case works essentially as a fast “shutter”. (Pockels cell: Model 350-160, Conoptics Inc., Danbury, CT, USA)
Beam-sizing telescope (in combination with 8 it provides the beam of convenient size at the input port of the microscope – to properly fill the back aperture of the microscope objective). This is composed of lenses (model BK7 with anti-reflection coating permitting near IR), holders and cage frames from Thorlabs (Newton, NJ, USA)
An optional half-wave plate (depending on type of DOE used, it could be employed for changing diffraction efficiency and making intensity of zero diffraction order beamlet equal to other beamlets). Thorlabs HWP05M-950 achromatic λ/2 plate, 690–1200 nm.
Periscope – mirrors to deliver laser beam to the input port of an upright microscope
Slow mechanical safety shutter (part of original Fluoview 200 system Olympus America Inc, Center Valley, PA, USA)
DOE [11 beamlets SLH-511D-1.5-(785) or 5 beamlets SLH-505X-(0.23)-(780) from StockerYale, Dollard-Des-Ormeaux, QC, Canada]
DOE imaging telescope relays the image of the DOE surface to approximately the plane of scanning mirrors (Thorlabs including 1 inch diameter initial lens and 2 inch diameter second lens).
For 11-beamlet DOE we used first lens with focal distance F1 = 75 mm, and the second lens has focal distance F2 = 125 mm; distance between the DOE and the first lens is 70 mm to image the surface of DOE to the plane between galvanometer mirrors (and subsequently onto the plane of the back aperture of the microscope objective).
For 5-beamlet DOE we used first F1 = 100 mm, and F2 = 500 mm; distance between the DOE and the first lens is 60 mm. We were modifying the magnification of this relay telescope in order to achieve desired physical beamlet separation on the sample plane according to used magnification of the microscope objective. For purposes of convenience five-beamlet DOE and the relay telescope was actually placed on the optical table before the periscope mirrors.
An optional zero-order beam block (we used a thin metal rod for this purpose). This is often necessary in the case of DOEs that are designed to produce a pattern without a beamlet in the center location but which produce an unintended zero-order beamlet.
Scanning mirrors – original galvanometer scanners from the Fluoview 200 system controlled by native Olympus confocal software
Scan (or “pupil-transfer”) lens – an original part of Fluoview laser scanning system (Olympus part number FVX-PL-IBX50/T).
Upright microscope (Olympus BX50WI)
a. Standard dichroic for two-photon fluorescence detection
b. Tube lens – an essential part of the upright microscope
c. Microscope objective
d. Microscope bright field illumination condenser
Detection system (detects shorter-wavelength light signified by green arrows)
a. Band-pass filter to include only emission-wavelength light (Chroma, Rockingham, VT, USA)
b. A standard microscope trinocular tube to switch between imaging ports: one of them is used for traditional whole-filed PMT detection, another – for fast imaging using the camera
c. Cooled CCD camera (C9100-12, Hamamatsu, Bridgewater, NJ, USA)
d. Optional PMT for traditional slow scanning imaging (Hamamatsu H7422-40P cooled GaAs)
Current to voltage converters/signal amplifiers for PMT detection. (PE 5113 preamplifier Signal Recovery AMETEK Advanced Measurement Technology, Wokingham, UK.)
Data acquisition unit – the part of original Fluoview 200 system
Optional 2nd detector (Camera or PMT) for gathering optical signals in forward direction [two-photon excited fluorescence or second harmonics generation (SHG) signal)
PC for data acquisition and equipment control.
Figure 2Two forms of DOE scanning. Schematic representations of strategies for utilizing a diffractive element to enhance two-photon imaging. Upper left: traditional raster scan imaging requires time for a mirror system to scan an excitation beam across a field of view and has only a certain ceiling quantity of excitation possible before photodamage occurs. Upper right: using a diffractive optical element (DOE) one can split our beam into an arbitrary number of beamlets. Using beamlets spread horizontally it is possible to image in a way that allows for summation of the signals excited by the sample at each pixel to yield greater signal and increased signal to noise ratio. Horizontal DOE mode or “excitation boost” may be used for either line scans or full-field frame scans (at the expense of some spatial resolution). Bottom: Vertical DOE mode or “speed boost” achieves greater speed of scanning than traditional raster scanning. By spacing DOE-created beamlets widely over the vertical aspect of the field of view and scanning each beamlet simultaneously horizontally across a narrow strip of the field of view it is possible to excite the full field of view in 1/(number of beamlets) the amount of time required for a single beam. As indicated by arrows at left of each representation of a field of view, in the standard raster scan mode, a single beam must make horizontal line scans along the entire height of the field (long arrow), whereas in speed boost, each beamlet simultaneously scans a fraction of the vertical aspect of the field (short arrows). This method requires a camera or any other similar wide-field light collection device (such as a photodiode array or multianode photomultiplier tube) with a resolution equal to at least the number of beamlets used, since more than one area is excited simultaneously.
Figure 3Line scanning with a horizontal DOE. Line scan imaging using horizontal DOE scanning. (A) Full-frame raster scan images of a neuron filled with Fura 2 pentapotassium salt acquired using traditional single beam excitation (left) and five beamlet excitation (right). Note that with multiple beamlets there is slight blurring of the image of the neuron, however the narrow spacing minimizes this blurring to the extent that the cell is clearly distinguishable. These images were used to select a level for line scanning (white horizontal line in each image). (B) Line scans acquired at 80 scans per second using single beam (left) and multibeam (right) excitation, time is represented in the horizontal axis, each vertical column represents one scan. A limited spatial extent of the line scan which includes the cell soma is shown for display purposes. The neuron was patch clamped and driven to fire action potentials during the optical recording. (C) Intensity versus time profiles of linescans shown in (B). Raw brightness intensity scales identical in left and right graphs in arbitrary brightness units. Calcium transients are visible corresponding to times of action potential firing indicated in the whole cell patch clamp current clamp trace shown in (E). The cell was induced to fire increasing numbers of action potentials starting with 1 and ending with 5. Calcium transients grew monotonically in correspondence to the number of action potentials in both imaging regimes but can be seen more clearly in the five beamlet case. (D) Same data as in (C) but re-plotted such that percent change from baseline (DF/F) is equivalent for both traces. Note that calcium transients induced by a given number of spikes are the same amplitude in both conditions, however noise is decreased in the five beamlet case, rendering a greater signal to noise ratio with DOE imaging. Signal to noise ratio was improved by 1.91 ± 0.24 fold (n = 5 signals) in the excitation boost DOE imaging results shown versus the single beam imaging.
Figure 4Enhanced frame scan with a DOE. Full-frame calcium imaging using horizontal DOE mode. (A) Full-frame raster scan images of a population of neurons bulk loaded with Fura 2-AM calcium indicator dye acquired using traditional single beam excitation (left) and five beamlet excitation (right). Horizontal DOE mode slightly decreases spatial resolution but not to the extent that resolution of single cells is problematic. Images in the top row are the product of average pixel-wise projection of a movie. (B) Intensity versus time profiles of time-lapse movies of shown field of view for the cell indicated by arrow in (A). Scales identical in left and right graphs, vertical axis in arbitrary brightness units, horizontal axis in seconds. The neuron of interest was patch clamped was driven to fire sets of three action potentials during the optical recording and calcium transients are visible corresponding to times of action potential firing (indicated by arrows and vertical lines). Signal to noise ratio was improved by an average of 1.95 ± 0.72 fold in the excitation boost multibeam excitation boost frame scan mode relative to single beam imaging.
Figure 5Scanning with vertical DOE. Full-frame calcium imaging using vertical (speed-boost) DOE scanning. (A) Progressive full-field scanning of a paper sample using many beams spread vertically over the field of view and captured with a CCD camera. At left is a “line scan” which creates a number of lines of excitation equal to the number of beamlets. Moving right, a demonstration of the scanning performed by this system: the beamlets are allowed to scan for a longer time with each image, eventually covering the entire field of view with each frame capture by the camera. (B) Full-frame scanning of a patch clamped neuron using a traditional single beam with PMT detection. Movie acquired at one frame per second. Below is intensity versus time profile and patch clamp recording for cell indicated by arrow which was driven to bursts of increasing numbers of action potentials (number of action potentials per burst indicated below trace and timing indicated in imaging trace with dotted vertical line). Vertical axis of brightness trace in arbitrary brightness units, horizontal axis in seconds. (C) Vertical DOE scanning of a population of neurons using 11 beamlet excitation and imaging with a CCD camera. Movie collected at 10 frames per second. Note that calcium transients corresponding to times of bursts of two or three action potentials are easily visible in the calcium indicator tracing produced by DOE-based imaging and are less clear with single beam raster scan. Furthermore, while calcium transients produced by 9–11 action potentials per burst were easily distinguishable in both cases, they were found o have a 2.15-fold greater signal to noise ratio in the case of speed boost imaging.