Literature DB >> 1961153

Physical performance characteristics of spiral CT scanning.

W A Kalender1, A Polacin.   

Abstract

CT scanning in spiral geometry is achieved by continuously transporting the patient through the gantry in synchrony with continuous data acquisition over a multitude of 360-deg scans. Data for reconstruction of images in planar geometry are estimated from the spiral data by interpolation. The influence of spiral scanning on image quality is investigated. Most of the standard physical performance parameters, e.g., spatial resolution, image uniformity, and contrast, are not affected; results differ for pixel noise and slice sensitivity profiles. For linear interpolation, pixel noise is expected to be reduced by a factor of 0.82; reduction factors of 0.81 to 0.83 were measured. Slice sensitivity profiles are changed as a function of table feed d, measured in millimeters per 360-deg scan; they are smoothed as the original profile is convolved with the object motion function. The motion function is derived for linear interpolation that constitutes a triangle with a base line width of 2d and a maximal height equal to 1/d. Calculations of both the full width at half-maximum and the shape of the profiles were in good agreement with experimental results. The effect of the widened profiles, in particular of their extended tail ends, on image quality is demonstrated in phantom measurements.

Entities:  

Mesh:

Year:  1991        PMID: 1961153     DOI: 10.1118/1.596607

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  18 in total

Review 1.  Cross-sectional vascular imaging with CT and MR angiography.

Authors:  Hasan K Kabul; Klaus D Hagspiel
Journal:  J Nucl Cardiol       Date:  2006 May-Jun       Impact factor: 5.952

Review 2.  Anniversary paper. Development of x-ray computed tomography: the role of medical physics and AAPM from the 1970s to present.

Authors:  Xiaochuan Pan; Jeffrey Siewerdsen; Patrick J La Riviere; Willi A Kalender
Journal:  Med Phys       Date:  2008-08       Impact factor: 4.071

3.  Image quality and dose in spiral computed tomography.

Authors:  F R Verdun; R A Meuli; F O Bochud; C Imsand; S Raimondi; P Schnyder; J F Valley
Journal:  Eur Radiol       Date:  1996       Impact factor: 5.315

4.  Spiral CT of the temporal bone: value of image reconstruction at submillimetric table increments.

Authors:  R Hermans; G Marchal; L Feenstra; A L Baert
Journal:  Neuroradiology       Date:  1995-02       Impact factor: 2.804

5.  CT angiography of the common carotid artery bifurcation: comparison between two techniques and conventional angiography.

Authors:  M Castillo; J D Wilson
Journal:  Neuroradiology       Date:  1994-11       Impact factor: 2.804

6.  Spiral CT of musculoskeletal pathology: preliminary observations.

Authors:  E K Fishman; S H Wyatt; D A Bluemke; B A Urban
Journal:  Skeletal Radiol       Date:  1993       Impact factor: 2.199

7.  Use of three-dimensional spiral computed tomography imaging for staging and surgical planning of head and neck cancer.

Authors:  C Franca; D Levin-Plotnik; V Sehgal; G T Chen; R G Ramsey
Journal:  J Digit Imaging       Date:  2000-05       Impact factor: 4.056

8.  Multidetector row computed tomography evaluation of the micropig kidney as a potential renal donor.

Authors:  Woong Yoon; Min Young Lee; Jung Min Ryu; Yong Ju Moon; Sang Hun Lee; Jae Hong Park; Seung Pil Yun; Min Woo Jang; Sung Su Park; Ho Jae Han
Journal:  J Vet Sci       Date:  2010-03       Impact factor: 1.672

9.  The quality of reconstructed 3D images in multidetector-row helical CT: experimental study involving scan parameters.

Authors:  Ji Hoon Shin; Ho Kyu Lee; Choong Gon Choi; Dae Chul Suh; Tae Hwan Lim; Weechang Kang
Journal:  Korean J Radiol       Date:  2002 Jan-Mar       Impact factor: 3.500

10.  64-channel multi-detector row CT angiographic evaluation of the micropigs for potential living donor lung transplantation.

Authors:  Woong Yoon; Jung Min Ryu; Min Young Lee; Yong Ju Moon; Sang Hun Lee; Jae Hong Park; Seung Pil Yun; Min Woo Jang; Sung Su Park; Ho Jae Han
Journal:  J Vet Sci       Date:  2010-09       Impact factor: 1.672

View more