Literature DB >> 19574903

The Boston Keratoprosthesis: comparing corneal epithelial cell compatibility with titanium and PMMA.

Jared D Ament1, Sandra J Spurr-Michaud, Claes H Dohlman, Ilene K Gipson.   

Abstract

PURPOSE: To determine in vitro whether titanium is superior in corneal cell compatibility to standard polymethyl-methacrylate (PMMA) for the Boston Keratoprosthesis (KPro).
METHODS: Human corneal-limbal epithelial (HCLE) cells were cultured 24, 48, 72, 96, 120, 144, or 168 hours in culture plates alone (controls) or with PMMA or titanium discs. Experiments were performed in triplicate and repeated (final n = 6). To determine if a soluble, toxic factor is emitted from materials, concurrent experiments at 48 and 144 hours were performed with discs placed in Transwell Supports, with HCLE cells plated beneath. As an additional test for soluble factors, cells were incubated 24 hours with disc-conditioned media, and number of viable cells per well was quantified at each timepoint by proliferation assay. To determine if delayed cell proliferation was attributable to cell death, HCLE cell death was measured under all conditions and quantified at each timepoint by cytotoxicity assay. The effects of material on HCLE cell proliferation over time was determined by repeated measures ANOVA. P < 0.05 was statistically significant.
RESULTS: HCLE cell proliferation was greater in wells with titanium discs compared to PMMA. Differences between the test discs and control non-disc cocultures were statistically significant over time for both cell proliferation (P = 0.001) and death (P = 0.0025). No significant difference was found using Transwells (P = 0.9836) or disc-conditioned media (P = 0.36).
CONCLUSION: This in vitro HCLE cell model demonstrates significantly increased cell proliferation and decreased cell death with cell/titanium contact compared to cell/PMMA contact. Moreover, differences are unlikely attributable to a soluble factor. Prospective in vivo analysis of the two KPro biomaterials is indicated.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19574903      PMCID: PMC4263318          DOI: 10.1097/ICO.0b013e31819670ac

Source DB:  PubMed          Journal:  Cornea        ISSN: 0277-3740            Impact factor:   2.651


  22 in total

Review 1.  Biomaterials surface characterization and modification.

Authors:  M Morra; C Cassinelli
Journal:  Int J Artif Organs       Date:  2006-09       Impact factor: 1.595

2.  A 1-year study on carbon, titanium surface-modified intraocular lens in rabbit eyes.

Authors:  Zhaoxu Yuan; Huimin Sun; Jiaqin Yuan
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2004-09-10       Impact factor: 3.117

3.  Using MTT viability assay to test the cytotoxicity of antibiotics and steroid to cultured porcine corneal endothelial cells.

Authors:  H Z Wang; C H Chang; C P Lin; M C Tsai
Journal:  J Ocul Pharmacol Ther       Date:  1996       Impact factor: 2.671

4.  Minimization of the inevitable residual monomer in denture base acrylic.

Authors:  C Y K Lung; B W Darvell
Journal:  Dent Mater       Date:  2005-07-22       Impact factor: 5.304

5.  PMMA versus titanium cage after anterior cervical discectomy - a prospective randomized trial.

Authors:  J Schröder; F Grosse-Dresselhaus; C Schul; H Wassmann
Journal:  Zentralbl Neurochir       Date:  2007-02

6.  Neutral red assay of the cytotoxicity of fluorocarbon-coated polymethylmethacrylate intraocular lenses in vitro.

Authors:  L Werner; J M Legeais; M D Nagel; G Renard
Journal:  J Biomed Mater Res       Date:  1999

7.  Cytotoxic effects of residual chemicals from polymeric biomaterials for artificial soft intraocular lenses.

Authors:  T V Chirila; L N Walker; I J Constable; D E Thompson; G D Barrett
Journal:  J Cataract Refract Surg       Date:  1991-03       Impact factor: 3.351

8.  Mucin gene expression in immortalized human corneal-limbal and conjunctival epithelial cell lines.

Authors:  Ilene K Gipson; Sandra Spurr-Michaud; Pablo Argüeso; Ann Tisdale; Tat Fong Ng; Cindy Leigh Russo
Journal:  Invest Ophthalmol Vis Sci       Date:  2003-06       Impact factor: 4.799

9.  PMMA particles and pressure--a study of the osteolytic properties of two agents proposed to cause prosthetic loosening.

Authors:  Björn Skoglund; Per Aspenberg
Journal:  J Orthop Res       Date:  2003-03       Impact factor: 3.494

10.  Importance of nutrition to corneal grafts when used as a carrier of the Boston Keratoprosthesis.

Authors:  Mona Harissi-Dagher; Bilal F Khan; Debra A Schaumberg; Claes H Dohlman
Journal:  Cornea       Date:  2007-06       Impact factor: 2.651

View more
  11 in total

1.  Titanium back plate for a PMMA keratoprosthesis: clinical outcomes.

Authors:  Amit Todani; Joseph B Ciolino; Jared D Ament; Kathryn A Colby; Roberto Pineda; Michael W Belin; James V Aquavella; James Chodosh; Claes H Dohlman
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2011-04-26       Impact factor: 3.117

2.  Boston type 1 keratoprosthesis: the New York Eye and Ear experience.

Authors:  A P Patel; E I Wu; D C Ritterband; J A Seedor
Journal:  Eye (Lond)       Date:  2011-12-16       Impact factor: 3.775

3.  Preparation and characterization of TiO2-coated polymerization of methyl methacrylate (PMMA) for biomedical applications: In vitro study.

Authors:  Leila Mohammadi Amirabad; Mohammadreza Tahriri; Payam Zarrintaj; Reza Ghaffari; Lobat Tayebi
Journal:  Asia Pac J Chem Eng       Date:  2022-03-07       Impact factor: 1.777

4.  In vivo performance of a drug-eluting contact lens to treat glaucoma for a month.

Authors:  Joseph B Ciolino; Cristina F Stefanescu; Amy E Ross; Borja Salvador-Culla; Priscila Cortez; Eden M Ford; Kate A Wymbs; Sarah L Sprague; Daniel R Mascoop; Shireen S Rudina; Sunia A Trauger; Fabiano Cade; Daniel S Kohane
Journal:  Biomaterials       Date:  2013-10-04       Impact factor: 12.479

5.  In vitro effect of a corrosive hostile ocular surface on candidate biomaterials for keratoprosthesis skirt.

Authors:  Xiao Wei Tan; Andri Riau; Zhi Long Shi; Anna C S Tan; Koon Gee Neoh; Khiam Aik Khor; Roger W Beuerman; Donald Tan; Jodhbir S Mehta
Journal:  Br J Ophthalmol       Date:  2012-07-16       Impact factor: 4.638

6.  Epithelial growth over the optic surface of the type 1 Boston Keratoprosthesis: histopathology and implications for biointegration.

Authors:  Yousuf M Khalifa; Don Davis; Nick Mamalis; Majid Moshirfar
Journal:  Clin Ophthalmol       Date:  2010-10-05

7.  Thin-film electroencephalographic electrodes using multi-walled carbon nanotubes are effective for neurosurgery.

Authors:  Kousuke Awara; Ryuhei Kitai; Makoto Isozaki; Hiroyuki Neishi; Kenichiro Kikuta; Naoki Fushisato; Akira Kawamoto
Journal:  Biomed Eng Online       Date:  2014-12-15       Impact factor: 2.819

8.  Titanium Coating of the Boston Keratoprosthesis.

Authors:  Borja Salvador-Culla; Kyung Jae Jeong; Paraskevi Evi Kolovou; Homer H Chiang; James Chodosh; Claes H Dohlman; Daniel S Kohane
Journal:  Transl Vis Sci Technol       Date:  2016-04-28       Impact factor: 3.283

Review 9.  Boston keratoprosthesis - Clinical outcomes with wider geographic use and expanding indications - A systematic review.

Authors:  Khalid Al Arfaj
Journal:  Saudi J Ophthalmol       Date:  2015-02-10

10.  Intrinsic Optical Properties of Boston Keratoprosthesis.

Authors:  Pui-Chuen Hui; Leonardo A Pereira; Renald Dore; Shengtong Chen; Elise Taniguchi; James Chodosh; Claes H Dohlman; Eleftherios I Paschalis
Journal:  Transl Vis Sci Technol       Date:  2020-11-03       Impact factor: 3.283

View more