Glucose monitoring by microdialysis: performance in a multicentre study.

J K Nielsen1, G Freckmann, C Kapitza, G Ocvirk, K H Koelker, U Kamecke, R Gillen, I Amann-Zalan, N Jendrike, J S Christiansen, T Koschinsky, L Heinemann.   

Abstract

AIMS: The aim of this study was to assess the performance of the Continuous Research Tool (CRT) in a multicentre clinical-experimental study.
METHODS: Three patient groups totalling 28 subjects with diabetes [group A 10 Type 1 (Ulm), group B 10 Type 1 (Neuss), group C eight Type 2 (Aarhus)] participated in this trial. Two CRT microdialysis probes were inserted in parallel in the abdominal subcutaneous tissue for 120 h in each subject. In subjects in group A, glucose excursions were induced on one study day and those in group B underwent a glucose clamp (eu-, hypo- or hyperglycaemic) on one study day. CRT data were calibrated once with a retrospective calibration model based on a run-in time of 24 h and three blood glucose measurements per day.
RESULTS: All analysable experiments, covering a broad range of blood glucose values, yielded highly accurate data for the complete experimental time with a mean relative absolute difference of 12.8 +/- 6.0% and a predictive residual error sum of squares of 15.6 +/- 6.3 (mean +/- SD). Of all measurement results, 98.2% were in zones A and B of the error grid analysis. The average absolute differences were 1.14 mmol/l for Type 1 and 0.88 mmol/l for Type 2 diabetic patients. Relative absolute differences were 16.0% for Type 1 and 12.6% for Type 2 diabetic patients.
CONCLUSIONS: These results demonstrate that this microdialysis system allows reliable continuous glucose monitoring in patients with diabetes of either type.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19573121     DOI: 10.1111/j.1464-5491.2009.02750.x

Source DB:  PubMed          Journal:  Diabet Med        ISSN: 0742-3071            Impact factor:   4.359


  9 in total

Review 1.  Technologies for continuous glucose monitoring: current problems and future promises.

Authors:  Santhisagar Vaddiraju; Diane J Burgess; Ioannis Tomazos; Faquir C Jain; Fotios Papadimitrakopoulos
Journal:  J Diabetes Sci Technol       Date:  2010-11-01

2.  Performance evaluation of three continuous glucose monitoring systems: comparison of six sensors per subject in parallel.

Authors:  Guido Freckmann; Stefan Pleus; Manuela Link; Eva Zschornack; Hans-Martin Klötzer; Cornelia Haug
Journal:  J Diabetes Sci Technol       Date:  2013-07-01

3.  Benefits and Limitations of MARD as a Performance Parameter for Continuous Glucose Monitoring in the Interstitial Space.

Authors:  Lutz Heinemann; Michael Schoemaker; Günther Schmelzeisen-Redecker; Rolf Hinzmann; Adham Kassab; Guido Freckmann; Florian Reiterer; Luigi Del Re
Journal:  J Diabetes Sci Technol       Date:  2019-06-19

4.  Use of microdialysis-based continuous glucose monitoring to drive real-time semi-closed-loop insulin infusion.

Authors:  Guido Freckmann; Nina Jendrike; Stefan Pleus; Harvey Buck; Steven Bousamra; Paul Galley; Ajay Thukral; Robin Wagner; Stefan Weinert; Cornelia Haug
Journal:  J Diabetes Sci Technol       Date:  2014-09-09

Review 5.  Biocompatible materials for continuous glucose monitoring devices.

Authors:  Scott P Nichols; Ahyeon Koh; Wesley L Storm; Jae Ho Shin; Mark H Schoenfisch
Journal:  Chem Rev       Date:  2013-02-07       Impact factor: 60.622

6.  Glucose Measurement by Affinity Sensor and Pulsed Measurements of Fluidic Resistances: Proof of Principle.

Authors:  Uwe Beyer; Thomas Wyss; Franck Robin; Lutz Heinemann
Journal:  J Diabetes Sci Technol       Date:  2014-01-01

7.  Performance evaluations of continuous glucose monitoring systems: precision absolute relative deviation is part of the assessment.

Authors:  Karin Obermaier; Günther Schmelzeisen-Redeker; Michael Schoemaker; Hans-Martin Klötzer; Harald Kirchsteiger; Heino Eikmeier; Luigi del Re
Journal:  J Diabetes Sci Technol       Date:  2013-07-01

8.  Performance evaluation of a continuous glucose monitoring system under conditions similar to daily life.

Authors:  Stefan Pleus; Christina Schmid; Manuela Link; Eva Zschornack; Hans-Martin Klötzer; Cornelia Haug; Guido Freckmann
Journal:  J Diabetes Sci Technol       Date:  2013-07-01

Review 9.  Glucose biosensors: an overview of use in clinical practice.

Authors:  Eun-Hyung Yoo; Soo-Youn Lee
Journal:  Sensors (Basel)       Date:  2010-05-04       Impact factor: 3.576

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.