Stochastic noise characteristics in matrix inversion tomosynthesis (MITS).

Devon J Godfrey1, H P McAdams, James T Third Dobbins.   

Abstract

Matrix inversion tomosynthesis (MITS) uses known imaging geometry and linear systems theory to deterministically separate in-plane detail from residual tomographic blur in a set of conventional tomosynthesis ("shift-and-add") planes. A previous investigation explored the effect of scan angle (ANG), number of projections (N), and number of reconstructed planes (NP) on the MITS impulse response and modulation transfer function characteristics, and concluded that ANG = 20 degrees, N = 71, and NP = 69 is the optimal MITS imaging technique for chest imaging on our prototype tomosynthesis system. This article examines the effect of ANG, N, and NP on the MITS exposure-normalized noise power spectra (ENNPS) and seeks to confirm that the imaging parameters selected previously by an analysis of the MITS impulse response also yield reasonable stochastic properties in MITS reconstructed planes. ENNPS curves were generated for experimentally acquired mean-subtracted projection images, conventional tomosynthesis planes, and MITS planes with varying combinations of the parameters ANG, N, and NP. Image data were collected using a prototype tomosynthesis system, with 11.4 cm acrylic placed near the image receptor to produce lung-equivalent beam hardening and scattered radiation. Ten identically acquired tomosynthesis data sets (realizations) were collected for each selected technique and used to generate ensemble mean images that were subtracted from individual image realizations prior to noise power spectra (NPS) estimation. NPS curves were normalized to account for differences in entrance exposure (as measured with an ion chamber), yielding estimates of the ENNPS for each technique. Results suggest that mid- and high-frequency noise in MITS planes is fairly equivalent in magnitude to noise in conventional tomosynthesis planes, but low-frequency noise is amplified in the most anterior and posterior reconstruction planes. Selecting the largest available number of projections (N = 71) does not incur any appreciable additive electronic noise penalty compared to using fewer projections for roughly equivalent cumulative exposure. Stochastic noise is minimized by maximizing N and NP but increases with increasing ANG. The noise trend results for NP and ANG are contrary to what would be predicted by simply considering the MITS matrix conditioning and likely result from the interplay between noise correlation and the polarity of the MITS filters. From this study, the authors conclude that the previously determined optimal MITS imaging strategy based on impulse response considerations produces somewhat suboptimal stochastic noise characteristics, but is probably still the best technique for MITS imaging of the chest.

Entities:  

Mesh:

Year:  2009        PMID: 19544768      PMCID: PMC2736704          DOI: 10.1118/1.3103399

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  8 in total

1.  Detection of subtle lung nodules: relative influence of quantum and anatomic noise on chest radiographs.

Authors:  E Samei; M J Flynn; W R Eyler
Journal:  Radiology       Date:  1999-12       Impact factor: 11.105

2.  A framework for noise-power spectrum analysis of multidimensional images.

Authors:  J H Siewerdsen; I A Cunningham; D A Jaffray
Journal:  Med Phys       Date:  2002-11       Impact factor: 4.071

Review 3.  Digital x-ray tomosynthesis: current state of the art and clinical potential.

Authors:  James T Dobbins; Devon J Godfrey
Journal:  Phys Med Biol       Date:  2003-10-07       Impact factor: 3.609

4.  Optimization of the matrix inversion tomosynthesis (MITS) impulse response and modulation transfer function characteristics for chest imaging.

Authors:  Devon J Godfrey; H P McAdams; James T Dobbins
Journal:  Med Phys       Date:  2006-03       Impact factor: 4.071

5.  Digital tomosynthesis of the chest for lung nodule detection: interim sensitivity results from an ongoing NIH-sponsored trial.

Authors:  T Dobbins James; H Page McAdams; Jae-Woo Song; Christina M Li; Devon J Godfrey; David M DeLong; Sang-Hyun Paik; Santiago Martinez-Jimenez
Journal:  Med Phys       Date:  2008-06       Impact factor: 4.071

6.  Digital tomosynthesis of hand joints for arthritis assessment.

Authors:  J Duryea; J T Dobbins; J A Lynch
Journal:  Med Phys       Date:  2003-03       Impact factor: 4.071

7.  DQE(f) of four generations of computed radiography acquisition devices.

Authors:  J T Dobbins; D L Ergun; L Rutz; D A Hinshaw; H Blume; D C Clark
Journal:  Med Phys       Date:  1995-10       Impact factor: 4.071

8.  Tomosynthesis: a three-dimensional radiographic imaging technique.

Authors:  D G Grant
Journal:  IEEE Trans Biomed Eng       Date:  1972-01       Impact factor: 4.538

  8 in total
  3 in total

1.  Task-based assessment of breast tomosynthesis: effect of acquisition parameters and quantum noise.

Authors:  I Reiser; R M Nishikawa
Journal:  Med Phys       Date:  2010-04       Impact factor: 4.071

2.  A dual-view digital tomosynthesis imaging technique for improved chest imaging.

Authors:  Yuncheng Zhong; Chao-Jen Lai; Tianpeng Wang; Chris C Shaw
Journal:  Med Phys       Date:  2015-09       Impact factor: 4.071

3.  The effect of averaging adjacent planes for artifact reduction in matrix inversion tomosynthesis.

Authors:  Devon J Godfrey; H Page McAdams; James T Dobbins
Journal:  Med Phys       Date:  2013-02       Impact factor: 4.071

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.