Abstract
In 8 patients with implanted pacemaker (3 unipolar, 5 bipolar systems) the risk of interference of a regular pacemaker function and an energy source with a capacity of 110 Watt and a frequency of 434 MHz was investigated. The results show that thermotherapy in patients with prostatic hypertrophy can be used without the risk of interference with regular pacemaker function.Entities:
Year: 1997 PMID: 19495680 DOI: 10.1007/BF03042480
Source DB: PubMed Journal: Herzschrittmacherther Elektrophysiol ISSN: 0938-7412