Reduction of dose to the female breast as a result of spectral optimisation for high-contrast thoracic CT imaging: a phantom study.

S V Vollmar1, W A Kalender.   

Abstract

Various approaches to reduce dose to the female breast in thoracic CT have been investigated. We evaluated the potential for reduction of dose to the breast by optimal choice of the X-ray spectra. The effect of X-ray energy variation on dose to the female breast in thoracic CT was examined by simulations and measurements of image contrast, image noise and radiation dose. A standard thorax phantom was used with various extension rings and breasts added and the following contrast inserts: iodine, calcium hydroxyapatite and a pure soft-tissue density difference. Three-dimensional dose distributions were determined by a validated Monte Carlo tool. The contrast-to-noise ratio per unit dose (CNRD) was determined for tube voltages of 40-200 kV by simulations and for 60-140 kV by measurements on a clinical CT scanner. CNRD curves did not show significant variations in soft-tissue density contrast, but considerable optimisation potential for iodine and skeletal imaging at reduced energies. Exact values depend on the patient's cross-section and X-ray spectrum. For example, reducing the tube voltage from 120 kV to 80 kV on the scanner reduced dose to the female breast typically by 50% without deterioration of the CNR. This method exceeds the dose reduction potential of other measures. We conclude that tube voltages in thoracic CT can be lowered for contrast medium and skeletal imaging without affecting the CNR but with a significant decrease in dose to the female breast.

Entities:  

Mesh:

Year:  2009        PMID: 19470573     DOI: 10.1259/bjr/28017710

Source DB:  PubMed          Journal:  Br J Radiol        ISSN: 0007-1285            Impact factor:   3.039


  10 in total

1.  The effects of bismuth breast shields in conjunction with automatic tube current modulation in CT imaging.

Authors:  Sabah Servaes; Xiaowei Zhu
Journal:  Pediatr Radiol       Date:  2013-05-23

2.  Cone beam breast CT with a high pitch (75 μm), thick (500 μm) scintillator CMOS flat panel detector: visibility of simulated microcalcifications.

Authors:  Youtao Shen; Yuncheng Zhong; Chao-Jen Lai; Tianpeng Wang; Chris C Shaw
Journal:  Med Phys       Date:  2013-10       Impact factor: 4.071

3.  Dose reduction in oncological staging multidetector CT: effect of iterative reconstruction.

Authors:  M Karpitschka; D Augart; H-C Becker; M Reiser; A Graser
Journal:  Br J Radiol       Date:  2013-01       Impact factor: 3.039

4.  Experimentally determined spectral optimization for dedicated breast computed tomography.

Authors:  Nicolas D Prionas; Shih-Ying Huang; John M Boone
Journal:  Med Phys       Date:  2011-02       Impact factor: 4.071

5.  Coronary CTA using scout-based automated tube potential and current selection algorithm, with breast displacement results in lower radiation exposure in females compared to males.

Authors:  Harshna Vadvala; Phillip Kim; Thomas Mayrhofer; Oleg Pianykh; Mannudeep Kalra; Udo Hoffmann; Brian Ghoshhajra
Journal:  Cardiovasc Diagn Ther       Date:  2014-12

6.  Chest circumference as a predictive parameter of computed tomography coronary angiography radiation doses from dual-source computed tomography.

Authors:  M Erdem Sagsoz; Ummugulsum Bayraktutan; Hayri Ogul; Mecit Kantarci
Journal:  Eurasian J Med       Date:  2013-02

7.  Predicting the image noise level of prospective ECG-triggered coronary computed tomography angiography: quantitative measurement of thoracic component versus body mass index.

Authors:  Hyeongmin Kim; Chul Hwan Park; Kyung Hwa Han; Tae Hoon Kim
Journal:  Int J Cardiovasc Imaging       Date:  2015-10-27       Impact factor: 2.357

8.  Contrast agent and radiation dose reduction in abdominal CT by a combination of low tube voltage and advanced image reconstruction algorithms.

Authors:  Nico Buls; Gert Van Gompel; Toon Van Cauteren; Koenraad Nieboer; Inneke Willekens; Guy Verfaillie; Paul Evans; Sven Macholl; Ben Newton; Johan de Mey
Journal:  Eur Radiol       Date:  2014-11-29       Impact factor: 5.315

9.  An Investigation of Calibration Phantoms for CT Scanners with Tube Voltage Modulation.

Authors:  Jing Zou; Xiaodong Hu; Hanyu Lv; Xiaotang Hu
Journal:  Int J Biomed Imaging       Date:  2013-12-25

10.  A dose-neutral image quality comparison of different CBCT and CT systems using paranasal sinus imaging protocols and phantoms.

Authors:  Ari-Petteri Ronkainen; Ali Al-Gburi; Timo Liimatainen; Hanna Matikka
Journal:  Eur Arch Otorhinolaryngol       Date:  2022-01-27       Impact factor: 3.236

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.