Literature DB >> 19455702

Unusual presentation of nickel allergy requiring explantation of an Amplatzer atrial septal occluder device.

David G Rabkin1, Kevin J Whitehead, Andrew D Michaels, Douglas L Powell, S V Karwande.   

Abstract

Systemic allergic reaction to a percutaneous patent foramen ovale (PFO) occlusion device is a rare event with only scattered reports in the literature. Serious allergic reactions to these devices have a poorly defined incidence, presentation, and natural history. We present a woman with a previously unknown nickel allergy who developed severe chest pain beginning the morning after percutaneous device closure of the PFO. Despite multiple visits to her cardiologists and primary care physicians, the cause of her chest pain remained unclear. After seeking a second opinion at our medical center, skin testing showed a severe reaction to nickel. These symptoms were refractory to treatment until device explantation 18 mo later. This case highlights the importance of recognizing nickel allergy as a cause of chest pain following implantation of certain types of devices used for closure of PFOs and other heart defects.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19455702      PMCID: PMC6653388          DOI: 10.1002/clc.20427

Source DB:  PubMed          Journal:  Clin Cardiol        ISSN: 0160-9289            Impact factor:   2.882


  10 in total

Review 1.  Atrial septum defect closure device in a beating heart, from the perspective of a researcher in artificial organs.

Authors:  Yasuko Tomizawa
Journal:  J Artif Organs       Date:  2012-06-23       Impact factor: 1.731

Review 2.  Surgical solution to an intracorporeal nickel allergy.

Authors:  Andrew Stamm; Paul Kozlowski; Jared Brandenberger
Journal:  Rev Urol       Date:  2017

3.  Endocarditis and Incomplete Endothelialization 12 Years after Amplatzer Septal Occluder Deployment.

Authors:  Allan K Nguyen; Brian A Palafox; Joanne P Starr; Richard N Gates; Farbouch Berdjis
Journal:  Tex Heart Inst J       Date:  2016-06-01

4.  Current practices in corrosion, surface characterization, and nickel leach testing of cardiovascular metallic implants.

Authors:  Srinidhi Nagaraja; Matthew Di Prima; David Saylor; Erica Takai
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2016-02-16       Impact factor: 3.368

5.  Stent thrombosis caused by metal allergy complicated by protein S deficiency and heparin-induced thrombocytopenia: a case report and review of the literature.

Authors:  Takao Konishi; Tadashi Yamamoto; Naohiro Funayama; Beni Yamaguchi; Seiichiro Sakurai; Hiroshi Nishihara; Koko Yamazaki; Yusuke Kashiwagi; Yasuki Sasa; Mitsuru Gima; Hideichi Tanaka; Daisuke Hotta; Kenjiro Kikuchi
Journal:  Thromb J       Date:  2015-07-23

6.  Can Computer Tomography Predict Compromise of Cardiac Structures After Percutaneous Closure of Interatrial Septal Defects?

Authors:  Philipp Wagdi
Journal:  Cardiol Res       Date:  2011-11-20

Review 7.  Patent Foramen Ovale Closure for Stroke Prevention and Other Disorders.

Authors:  Fareed Moses S Collado; Marie-France Poulin; Joshua J Murphy; Hani Jneid; Clifford J Kavinsky
Journal:  J Am Heart Assoc       Date:  2018-06-17       Impact factor: 5.501

8.  Cocoon devices for transcatheter closure of atrial septal defect and patent ductus arteriosus in children: Single center experience.

Authors:  Hyam Mahmoud; Alin Nicolescu; Cristina Filip; Gabriela Duica; Georgiana Nicolae; Eliza Elena Cinteza
Journal:  Medicine (Baltimore)       Date:  2019-03       Impact factor: 1.817

Review 9.  Nickel Hypersensitivity to Atrial Septal Occluders: Smoke Without Fire?

Authors:  Anastasios Apostolos; Maria Drakopoulou; Stamatios Gregoriou; Andreas Synetos; George Trantalis; Georgios Tsivgoulis; Spyridon Deftereos; Konstantinos Tsioufis; Konstantinos Toutouzas
Journal:  Clin Rev Allergy Immunol       Date:  2021-06-15       Impact factor: 8.667

10.  Severe generalized dermatitis in a nickel-allergic patient with a popliteal artery nitinol stent.

Authors:  Andres Guerra; Melissa Kirkwood
Journal:  J Vasc Surg Cases Innov Tech       Date:  2017-03-06
  10 in total