Literature DB >> 26880035

Current practices in corrosion, surface characterization, and nickel leach testing of cardiovascular metallic implants.

Srinidhi Nagaraja1, Matthew Di Prima1, David Saylor1, Erica Takai1.   

Abstract

In an effort to better understand current test practices and improve nonclinical testing of cardiovascular metallic implants, the Food and Drug Administration (FDA) held a public workshop on Cardiovascular Metallic Implants: corrosion, surface characterization, and nickel leaching. The following topics were discussed: (1) methods used for corrosion assessments, surface characterization techniques, and nickel leach testing of metallic cardiovascular implant devices, (2) the limitations of each of these in vitro tests in predicting in vivo performance, (3) the need, utility, and circumstances when each test should be considered, and (4) the potential testing paradigms, including acceptance criteria for each test. In addition to the above topics, best practices for these various tests were discussed, and knowledge gaps were identified. Prior to the workshop, discussants had the option to provide feedback and information on issues relating to each of the topics via a voluntary preworkshop assignment. During the workshop, the pooled responses were presented and a panel of experts discussed the results. This article summarizes the proceedings of this workshop and background information provided by workshop participants. Published 2016. This article is a U.S. Government work and is in the public domain in the USA. J Biomed Mater Res Part B: Appl Biomater, 105B: 1330-1341, 2017. Published 2016. This article is a U.S. Government work and is in the public domain in the USA.

Entities:  

Keywords:  cardiovascular; corrosion; metal (alloys); metal ions

Mesh:

Substances:

Year:  2016        PMID: 26880035      PMCID: PMC6026850          DOI: 10.1002/jbm.b.33630

Source DB:  PubMed          Journal:  J Biomed Mater Res B Appl Biomater        ISSN: 1552-4973            Impact factor:   3.368


  38 in total

1.  In vitro host response assessment of biomaterials for cardiovascular stent manufacture.

Authors:  Matteo Santin; Lyuba Mikhalovska; Andrew W Lloyd; Sergey Mikhalovsky; Louise Sigfrid; Stephen P Denyer; Susanna Field; Dennis Teer
Journal:  J Mater Sci Mater Med       Date:  2004-04       Impact factor: 3.896

2.  Susceptibility of nitinol to localized corrosion.

Authors:  Bruce G Pound
Journal:  J Biomed Mater Res A       Date:  2006-04       Impact factor: 4.396

3.  A case of allergic contact dermatitis following transcatheter closure of patent ductus arteriosus using Amplatzer ductal occluder.

Authors:  Kye Hun Kim; Jong Chun Park; Nam Sik Yoon; Jae Youn Moon; Young Joon Hong; Hyung Wook Park; Ju Han Kim; Youngkeun Ahn; Myung Ho Jeong; Jeong Gwan Cho; Jung Chaee Kang
Journal:  Int J Cardiol       Date:  2007-10-22       Impact factor: 4.164

4.  Serum levels of nickel and chromium after instrumented posterior spinal arthrodesis.

Authors:  Young-Jo Kim; Farid Kassab; Sigurd H Berven; David Zurakowski; M Timothy Hresko; John B Emans; James R Kasser
Journal:  Spine (Phila Pa 1976)       Date:  2005-04-15       Impact factor: 3.468

Review 5.  Nickel.

Authors:  D G Barceloux
Journal:  J Toxicol Clin Toxicol       Date:  1999

6.  Effect of surface treatment and sterilization processes on the corrosion behavior of NiTi shape memory alloy.

Authors:  B Thierry; M Tabrizian; C Trepanier; O Savadogo; L Yahia
Journal:  J Biomed Mater Res       Date:  2000-09-15

7.  Evaluation of metal allergies in patients with coronary stents.

Authors:  U Hillen; M Haude; R Erbel; M Goos
Journal:  Contact Dermatitis       Date:  2002-12       Impact factor: 6.600

8.  Metal allergy to Amplatzer occluder device presented as severe bronchospasm.

Authors:  Reza A Khodaverdian; Kent W Jones
Journal:  Ann Thorac Surg       Date:  2009-12       Impact factor: 4.330

9.  In vivo human electrochemical properties of a NiTi-based alloy (Nitinol) used for minimally invasive implants.

Authors:  Luana B Pértile; Patrícia M S Silva; Vinícius B Peccin; Ricardo Peres; Pierre G Silveira; Cristiano Giacomelli; Fernando C Giacomelli; Márcio C Fredel; Almir Spinelli
Journal:  J Biomed Mater Res A       Date:  2009-06-15       Impact factor: 4.396

10.  Nickel release, a possible indicator for the duration of antiplatelet treatment, from a nickel cardiac device in vivo: a study in patients with atrial septal defects implanted with an Amplatzer occluder.

Authors:  M Burian; T Neumann; M Weber; R Brandt; G Geisslinger; V Mitrovic; C Hamm
Journal:  Int J Clin Pharmacol Ther       Date:  2006-03       Impact factor: 1.366

View more
  3 in total

Review 1.  Extracellular Matrix Bioscaffolds as Immunomodulatory Biomaterials<sup/>.

Authors:  Jenna L Dziki; Luai Huleihel; Michelle E Scarritt; Stephen F Badylak
Journal:  Tissue Eng Part A       Date:  2017-05-19       Impact factor: 3.845

2.  The effect of fatigue on the corrosion resistance of common medical alloys.

Authors:  Matthew Di Prima; Erick Gutierrez; Jason D Weaver
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2016-07-04       Impact factor: 3.368

3.  On the High Sensitivity of Corrosion Resistance of NiTi Stents with Respect to Inclusions: An Experimental Evidence.

Authors:  Fan Sun; Laurence Jordan; Valérie Albin; Virginie Lair; Armelle Ringuedé; Frédéric Prima
Journal:  ACS Omega       Date:  2020-02-10
  3 in total