Wear of contemporary total knee replacements--a knee simulator study of six current designs.

Sandra Utzschneider1, Norbert Harrasser, Christian Schroeder, Farhad Mazoochian, Volkmar Jansson.   

Abstract

BACKGROUND: Compared to conventional ultra-high molecular weight polyethylene (UHMWPE), crosslinked polyethylene showed significantly reduced wear rate in hip simulation and early prospective-randomized clinical studies. The crosslinking process can reduce the mechanical properties of UHMWPE, particularly the fatigue strength. UHMWPE fatigue occurs more frequently in the knee joint than in the hip joint due to its higher contact stresses and there is therefore an increased concern of mechanical failure. Therefore, the purpose of this study was to determine the wear behaviour and the wear rates of different current knee designs and bearing materials.
METHODS: In a knee-joint-simulator four kinds of crosslinked polyethylene (one produced by sequential irradiation and annealing process, three by different remelting processes, including fixed- and mobile-bearing-types) and two UHMWPE- (fixed- and mobile-bearing) inserts were tested with the appropriate femoral and tibial component recommended from the manufacturer.
FINDINGS: All types of crosslinked polyethylene produced statistically significant (P<0.05) lower wear rates than the conventional UHMWPEs without any traces of fatigue. There were no differences in the wear rates between fixed-and mobile-bearings (crosslinked polyethylenes and UHMWPEs; P>0.05). The crosslinked polyethylene tibial inserts manufactured by sequential irradiation and annealing (X3) combined with the Scorpio-knee-design had the lowest wear rates (P<0.05) overall.
INTERPRETATION: Fixed- as well as mobile-bearing crosslinked polyethylene tibial inserts are suitable for total knee arthroplasty and showed reduced wear rates compared with conventional UHMWPE. The combination of the fixed-bearing Scorpio-knee-design with a sequential irradiated and annealed crosslinked polyethylene tibial insert (X3) seems to have an advantage in wear generation compared with other fixed- and mobile-bearing knee designs.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19450910     DOI: 10.1016/j.clinbiomech.2009.04.007

Source DB:  PubMed          Journal:  Clin Biomech (Bristol, Avon)        ISSN: 0268-0033            Impact factor:   2.063


  12 in total

1.  Mobility of the rotating platform in low contact stress knee arthroplasty is durable.

Authors:  Arthur Zürcher; Kim van Hutten; Jaap Harlaar; Ruud Pöll
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2015-10-16       Impact factor: 4.342

Review 2.  Polyethylene and metal wear particles: characteristics and biological effects.

Authors:  Isabelle Catelas; Markus A Wimmer; Sandra Utzschneider
Journal:  Semin Immunopathol       Date:  2011-01-26       Impact factor: 9.623

3.  Biological activity and migration of wear particles in the knee joint: an in vivo comparison of six different polyethylene materials.

Authors:  S Utzschneider; V Lorber; M Dedic; A C Paulus; C Schröder; O Gottschalk; M Schmitt-Sody; V Jansson
Journal:  J Mater Sci Mater Med       Date:  2014-02-22       Impact factor: 3.896

4.  CORR Insights(®): Surface Damage Is Not Reduced With Highly Crosslinked Polyethylene Tibial Inserts at Short-term.

Authors:  Harry A McKellop
Journal:  Clin Orthop Relat Res       Date:  2015-07-08       Impact factor: 4.176

5.  [Principles of tribological analysis of endoprostheses].

Authors:  J P Kretzer; C Zietz; C Schröder; J Reinders; L Middelborg; A Paulus; R Sonntag; R Bader; S Utzschneider
Journal:  Orthopade       Date:  2012-10       Impact factor: 1.087

Review 6.  Polyethylene in knee arthroplasty: A review.

Authors:  Gautam Chakrabarty; Mayank Vashishtha; Daniel Leeder
Journal:  J Clin Orthop Trauma       Date:  2015-02-19

7.  Surface Damage Is Not Reduced With Highly Crosslinked Polyethylene Tibial Inserts at Short-term.

Authors:  Tong Liu; Christina Esposito; Marcella Elpers; Timothy Wright
Journal:  Clin Orthop Relat Res       Date:  2016-01       Impact factor: 4.176

8.  Comparison of in vivo polyethylene wear particles between mobile- and fixed-bearing TKA in the same patients.

Authors:  Yukihide Minoda; Kanako Hata; Mitsuhiko Ikebuchi; Shigekazu Mizokawa; Yoichi Ohta; Hiroaki Nakamura
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2016-02-04       Impact factor: 4.342

Review 9.  [Possibilities and limits of modern polyethylenes. With respect to the application profile].

Authors:  S Utzschneider; A C Paulus; C Schröder; V Jansson
Journal:  Orthopade       Date:  2014-06       Impact factor: 1.087

10.  The influence of simulator input conditions on the wear of total knee replacements: An experimental and computational study.

Authors:  Claire L Brockett; Abdellatif Abdelgaied; Tony Haythornthwaite; Catherine Hardaker; John Fisher; Louise M Jennings
Journal:  Proc Inst Mech Eng H       Date:  2016-05       Impact factor: 1.617

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.