Literature DB >> 19431925

LASIK for myopia using the Zeiss VisuMax femtosecond laser and MEL 80 excimer laser.

Marcus Blum1, Kathleen Kunert, Annika Gille, Walter Sekundo.   

Abstract

PURPOSE: To assess the efficacy, safety, and predictability of the VisuMax femtosecond laser (Carl Zeiss Meditec).
METHODS: In a prospective, multicenter clinical trial, 32 eyes of 17 patients underwent LASIK and flap creation with the VisuMax femtosecond laser and ablation with the MEL 80 excimer laser (Carl Zeiss Meditec). Mean patient age was 35 years (range: 23 to 52 years). Mean preoperative spherical equivalent refraction was -4.04 +/- 1.39 diopters (D) (range: -1.50 to -7.25 D), and the planned flap thickness was 110, 120, 140, and 150 microm.
RESULTS: At the 3-month postoperative examination, all eyes had best spectacle-corrected visual acuity of 20/25 and uncorrected visual acuity (UCVA) of 20/40; 91% of eyes reached UCVA of 20/20 or better. Ninety-four percent of eyes were within +/- 0.50 D of the planned correction. Complications were limited to one case of suction loss, and no other adverse events were noted at 3 months. The flaps were planar and the mean achieved central flap thickness was 118 +/-9.2 microm for an attempted thickness of 120 microm.
CONCLUSIONS: The investigators achieved LASIK outcomes exceeding the US Food and Drug Administration guidelines for refractive procedures in this series of myopic eyes using the VisuMax femtosecond laser and MEL 80 excimer laser. The curved contact lens of the applanation glass combined with low suction permitted continuous fixation during treatment.

Entities:  

Mesh:

Year:  2009        PMID: 19431925     DOI: 10.3928/1081597X-20090401-01

Source DB:  PubMed          Journal:  J Refract Surg        ISSN: 1081-597X            Impact factor:   3.573


  14 in total

1.  Femtosecond lenticule extraction (ReLEx) for correction of hyperopia - first results.

Authors:  Marcus Blum; Kathleen S Kunert; Urs Voßmerbäumer; Walter Sekundo
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2012-06-14       Impact factor: 3.117

2.  [Femtosecond lenticule extraction (FLEx)].

Authors:  M Blum; W Sekundo
Journal:  Ophthalmologe       Date:  2010-10       Impact factor: 1.059

Review 3.  [Refractive lenticule extraction - The ReLEx/SMILE technique : Video article].

Authors:  M Blum; W Sekundo
Journal:  Ophthalmologe       Date:  2017-09       Impact factor: 1.059

4.  Laser in situ keratomileusis enhancements with the Ziemer FEMTO LDV femtosecond laser following previous LASIK treatments.

Authors:  Juhani Pietilä; Anne Huhtala; Petri Mäkinen; Hannu Uusitalo
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2012-07-20       Impact factor: 3.117

5.  Femtosecond lenticule extraction for the correction of myopia: preliminary 6-month results.

Authors:  Marcus Blum; Kathleen Kunert; Michael Schröder; Walter Sekundo
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-02-04       Impact factor: 3.117

6.  Flap characteristics, predictability, and safety of the Ziemer FEMTO LDV femtosecond laser with the disposable suction ring for LASIK.

Authors:  J Pietilä; A Huhtala; P Mäkinen; H Uusitalo
Journal:  Eye (Lond)       Date:  2013-11-15       Impact factor: 3.775

7.  Comparison of clinical outcomes in PRK with a standard and aspherical optimized profile: a full case analysis of 100 eyes with 1-year follow-up.

Authors:  Dieter Dausch; Burglinde Dausch; Matthias Wottke; Georg Sluyterman van Langeweyde
Journal:  Clin Ophthalmol       Date:  2014-11-24

8.  Visumax femtolasik versus Moria M2 microkeratome in mild to moderate myopia: efficacy, safety, predictability, aberrometric changes and flap thickness predictability.

Authors:  Magda A Torky; Yousif A Al Zafiri; Abeer M Khattab; Rania K Farag; Eman A Awad
Journal:  BMC Ophthalmol       Date:  2017-07-17       Impact factor: 2.209

9.  Control-matched comparison of refractive and visual outcomes between small incision lenticule extraction and femtosecond laser-assisted LASIK.

Authors:  Takahiro Kataoka; Tomoya Nishida; Azusa Murata; Mayuka Ito; Naoki Isogai; Rie Horai; Takashi Kojima; Yoko Yoshida; Tomoaki Nakamura
Journal:  Clin Ophthalmol       Date:  2018-05-10

10.  Femtosecond lenticule extraction for correction of myopia: a 6 month follow-up study.

Authors:  Ahmet Demirok; Alper Agca; Engin Bilge Ozgurhan; Ercument Bozkurt; Ugur Celik; Ali Demircan; Nimet Burcu Guleryuz; Kadir İlker Cankaya; Omer Faruk Yilmaz
Journal:  Clin Ophthalmol       Date:  2013-06-07