Literature DB >> 24232315

Flap characteristics, predictability, and safety of the Ziemer FEMTO LDV femtosecond laser with the disposable suction ring for LASIK.

J Pietilä1, A Huhtala2, P Mäkinen1, H Uusitalo3.   

Abstract

AIMS: The outcomes of laser-assisted in situ keratomileusis (LASIK) operations performed with the Classic FEMTO LDV femtosecond laser using the plastic single-use suction ring (Ziemer Ophthalmic Systems) and the Allegretto Wave Concerto 500 Hz excimer laser (Wavelight AG) are presented in terms of accuracy, predictability, and safety of the operation.
METHODS: A FEMTO LDV plastic suction ring was used for flap creation in 342 eyes of 179 patients. The intended flap thickness was 90 μm. The size of the suction ring varied from 9.0 to 10.0 mm. Flap dimensions were measured and correlated to preoperative characteristics.
RESULTS: Mean flap thickness was very constant, 89.6 ± 2.0 μm (range 84-97). In 163 bilateral operations, the second flap was 1.1 μm thinner than the one cut first (P<0.0001). Mean flap diameter was 9.4 ± 0.2 mm (range 8.1-9.9). Mean hinge length was 3.9 ± 0.2 mm (range 3.0-4.2). In hyperopic eyes, flap thickness correlated negatively with keratometric power K1 and flap diameter. In hyperopic eyes, flap diameter correlated positively with spherical equivalent refraction and with keratometric power K1 as well as hinge length both in myopic and hyperopic eyes. Complications were reported in 12 (3.5%) eyes. Complications were very mild and none of them prevented further refractive laser treatment. Two Snellen lines of corrected distance visual acuity were lost in one (0.3%) eye.
CONCLUSION: The FEMTO LDV plastic single-use suction rings yielded accurate and reproducible flaps and were safe for the creation of thin corneal flaps.

Entities:  

Mesh:

Year:  2013        PMID: 24232315      PMCID: PMC3890769          DOI: 10.1038/eye.2013.244

Source DB:  PubMed          Journal:  Eye (Lond)        ISSN: 0950-222X            Impact factor:   3.775


  13 in total

1.  Comparison of the IntraLase femtosecond laser and mechanical keratomes for laser in situ keratomileusis.

Authors:  Guy M Kezirian; Karl G Stonecipher
Journal:  J Cataract Refract Surg       Date:  2004-04       Impact factor: 3.351

2.  Flap dimensions created with the IntraLase FS laser.

Authors:  Perry S Binder
Journal:  J Cataract Refract Surg       Date:  2004-01       Impact factor: 3.351

3.  Comparison of the Ziemer FEMTO LDV femtosecond laser and Moria M2 mechanical microkeratome.

Authors:  Yuehua Zhou; Jing Zhang; Lei Tian; Changbin Zhai
Journal:  J Refract Surg       Date:  2012-03       Impact factor: 3.573

4.  Femtosecond laser-assisted corneal flap cuts: morphology, accuracy, and histopathology.

Authors:  Mike P Holzer; Tanja M Rabsilber; Gerd U Auffarth
Journal:  Invest Ophthalmol Vis Sci       Date:  2006-07       Impact factor: 4.799

5.  Comparison of the IntraLase femtosecond laser and mechanical microkeratome for laser in situ keratomileusis.

Authors:  Taehyung Lim; Sungjae Yang; Myoungjoon Kim; Hungwon Tchah
Journal:  Am J Ophthalmol       Date:  2006-05       Impact factor: 5.258

6.  Accuracy and precision of LASIK flap thickness using the IntraLase femtosecond laser in 1000 consecutive cases.

Authors:  Gerard Sutton; Chris Hodge
Journal:  J Refract Surg       Date:  2008-10       Impact factor: 3.573

7.  LASIK for myopia using the Zeiss VisuMax femtosecond laser and MEL 80 excimer laser.

Authors:  Marcus Blum; Kathleen Kunert; Annika Gille; Walter Sekundo
Journal:  J Refract Surg       Date:  2009-04       Impact factor: 3.573

8.  Very high-frequency digital ultrasound measurement of the LASIK flap thickness profile using the IntraLase femtosecond laser and M2 and Carriazo-Pendular microkeratomes.

Authors:  Jorge L Alió; David P Piñero
Journal:  J Refract Surg       Date:  2008-01       Impact factor: 3.573

9.  Reproducibility of flap thickness with IntraLase FS and Moria LSK-1 and M2 microkeratomes.

Authors:  Jonathan H Talamo; Jeremy Meltzer; John Gardner
Journal:  J Refract Surg       Date:  2006-06       Impact factor: 3.573

10.  US trends in refractive surgery: 2004 ISRS/AAO Survey.

Authors:  Richard J Duffey; David Leaming
Journal:  J Refract Surg       Date:  2005 Nov-Dec       Impact factor: 3.573

View more
  4 in total

1.  Comparison of Flap Characteristics Created with Two Different Methods in Laser in Situ Keratomileusis (LASIK).

Authors:  Melisa Ahmedbegovic-Pjano; Alma Biscevic; Amila Alikadic-Husovic; Nita Bejdic; Maja Bohac
Journal:  Med Arch       Date:  2021-06

Review 2.  Advances in refractive corneal lenticule extraction.

Authors:  Matthias Fuest; Jodhbir S Mehta
Journal:  Taiwan J Ophthalmol       Date:  2021-04-24

3.  Femtosecond laser versus mechanical microkeratome-assisted flap creation for LASIK: a prospective, randomized, paired-eye study.

Authors:  Bojan Pajic; Iraklis Vastardis; Brigitte Pajic-Eggspuehler; Zisis Gatzioufas; Farhad Hafezi
Journal:  Clin Ophthalmol       Date:  2014-09-22

4.  The Impact of Learning Curve on Flap Thickness Outcome in Femtosecond Laser-assisted Laser In Situ Keratomileusis Performed with New LenSx Dual Platform.

Authors:  Bilgehan Sezgin Asena; Oya Donmez
Journal:  Korean J Ophthalmol       Date:  2022-01-24
  4 in total