Effect of steel and teflon infusion catheters on subcutaneous adipose tissue blood flow and infusion counter pressure in humans.

Lise Højbjerre1, Camilla Skov-Jensen, Peter Kaastrup, Per Elgård Pedersen, Bente Stallknecht.   

Abstract

BACKGROUND: Subcutaneous tissue is an important target for drug deposition or infusion. A local trauma may induce alterations in local microcirculation and diffusion barriers with consequences for drug bioavailability. We examined the influence of infusion catheters' wear time on local microcirculation and infusion counter pressure.
METHODS: One steel catheter and one Teflon (Dupont, Wilmington, DE) catheter were inserted in subcutaneous, abdominal adipose tissue (SCAAT) in 10 healthy, lean men. The catheters were infused with isotonic saline at a rate of 10 microL/h for 48 h. Another steel catheter and a Teflon catheter were inserted contralateral to the previous catheters after 48 h. The infusion counter pressure was measured during a basal infusion rate followed by a bolus infusion. The measurements during a basal rate infusion were repeated after the bolus infusion. Adipose tissue blood flow (ATBF) was measured in SCAAT continuously.
RESULTS: A significant increase in ATBF was observed with wear time for Teflon but not for steel catheters. Mean infusion pressure during the bolus phase increased significantly from 0 to 48 h for Teflon but not for steel catheters. ATBF and infusion counter pressure was similar between Teflon and steel catheters after acute catheter implantation and after wear time of 48 h. The maximum value of pressure during the bolus phase increased with wear time of a catheter.
CONCLUSIONS: ATBF and bolus mean infusion pressure increased significantly with a wear time of 48 h in Teflon but not in steel catheters. The maximal pressure required to deliver a bolus infusion increased with wear time of a catheter. A higher maximal pressure was required to deliver a bolus infusion through a Teflon than through a steel catheter. We propose that the difference in infusion counter pressure and ATBF between Teflon and steel catheters with wear time may be explained by better biocompatibility of steel than Teflon.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19425878     DOI: 10.1089/dia.2008.0061

Source DB:  PubMed          Journal:  Diabetes Technol Ther        ISSN: 1520-9156            Impact factor:   6.118


  8 in total

Review 1.  Moving Toward a Unified Platform for Insulin Delivery and Sensing of Inputs Relevant to an Artificial Pancreas.

Authors:  Anneke Graf; Sybil A McAuley; Catriona Sims; Johanna Ulloa; Alicia J Jenkins; Gayane Voskanyan; David N O'Neal
Journal:  J Diabetes Sci Technol       Date:  2016-12-13

2.  Randomized trial of infusion set function: steel versus teflon.

Authors:  Parul J Patel; Kari Benasi; Gina Ferrari; Mark G Evans; Satya Shanmugham; Darrell M Wilson; Bruce A Buckingham
Journal:  Diabetes Technol Ther       Date:  2013-10-03       Impact factor: 6.118

3.  Insulin depot formation in subcutaneoue tissue.

Authors:  James P Leuenberger Jockel; Philipp Roebrock; Oliver A Shergold
Journal:  J Diabetes Sci Technol       Date:  2013-01-01

4.  Tissue Response to Subcutaneous Infusion Catheter.

Authors:  Ershuai Zhang; Zhiqiang Cao
Journal:  J Diabetes Sci Technol       Date:  2019-03-31

5.  Pilot study for assessment of optimal frequency for changing catheters in insulin pump therapy-trouble starts on day 3.

Authors:  Volkmar Schmid; Cloth Hohberg; Marcus Borchert; Thomas Forst; Andreas Pfützner
Journal:  J Diabetes Sci Technol       Date:  2010-07-01

Review 6.  Insulin infusion set: the Achilles heel of continuous subcutaneous insulin infusion.

Authors:  Lutz Heinemann; Lars Krinelke
Journal:  J Diabetes Sci Technol       Date:  2012-07-01

7.  Using Insulin Infusion Sets in CSII for Longer Than the Recommended Usage Time Leads to a High Risk for Adverse Events: Results From a Prospective Randomized Crossover Study.

Authors:  Andreas Pfützner; Daniela Sachsenheimer; Marco Grenningloh; Matthias Heschel; Lene Walther-Johannesen; Rabi Gharabli; David Klonoff
Journal:  J Diabetes Sci Technol       Date:  2015-09-03

8.  Insulin induces a progressive increase in the resistance of subcutaneous tissue to fluid flow: Implications for insulin pump therapy.

Authors:  Werner Regittnig; Mathias Tschaikner; Alexandru-Cristian Tuca; Amra Simic; Jürgen Feiel; Roland Schaller-Ammann; Alexander H Licht; Miró Jungklaus; Thomas R Pieber
Journal:  Diabetes Obes Metab       Date:  2021-11-17       Impact factor: 6.408

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.