Literature DB >> 19420426

Experimental validation of a Monte Carlo proton therapy nozzle model incorporating magnetically steered protons.

S W Peterson1, J Polf, M Bues, G Ciangaru, L Archambault, S Beddar, A Smith.   

Abstract

The purpose of this study is to validate the accuracy of a Monte Carlo calculation model of a proton magnetic beam scanning delivery nozzle developed using the Geant4 toolkit. The Monte Carlo model was used to produce depth dose and lateral profiles, which were compared to data measured in the clinical scanning treatment nozzle at several energies. Comparisons were also made between measured and simulated off-axis profiles to test the accuracy of the model's magnetic steering. Comparison of the 80% distal dose fall-off values for the measured and simulated depth dose profiles agreed to within 1 mm for the beam energies evaluated. Agreement of the full width at half maximum values for the measured and simulated lateral fluence profiles was within 1.3 mm for all energies. The position of measured and simulated spot positions for the magnetically steered beams agreed to within 0.7 mm of each other. Based on these results, we found that the Geant4 Monte Carlo model of the beam scanning nozzle has the ability to accurately predict depth dose profiles, lateral profiles perpendicular to the beam axis and magnetic steering of a proton beam during beam scanning proton therapy.

Mesh:

Substances:

Year:  2009        PMID: 19420426     DOI: 10.1088/0031-9155/54/10/017

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  12 in total

1.  Verification of proton range, position, and intensity in IMPT with a 3D liquid scintillator detector system.

Authors:  L Archambault; F Poenisch; N Sahoo; D Robertson; A Lee; M T Gillin; R Mohan; S Beddar
Journal:  Med Phys       Date:  2012-03       Impact factor: 4.071

2.  An MCNPX Monte Carlo model of a discrete spot scanning proton beam therapy nozzle.

Authors:  Gabriel O Sawakuchi; Dragan Mirkovic; Luis A Perles; Narayan Sahoo; X Ron Zhu; George Ciangaru; Kazumichi Suzuki; Michael T Gillin; Radhe Mohan; Uwe Titt
Journal:  Med Phys       Date:  2010-09       Impact factor: 4.071

3.  Optimizing a three-stage Compton camera for measuring prompt gamma rays emitted during proton radiotherapy.

Authors:  S W Peterson; D Robertson; J Polf
Journal:  Phys Med Biol       Date:  2010-11-03       Impact factor: 3.609

4.  TOPAS: an innovative proton Monte Carlo platform for research and clinical applications.

Authors:  J Perl; J Shin; J Schumann; B Faddegon; H Paganetti
Journal:  Med Phys       Date:  2012-11       Impact factor: 4.071

Review 5.  The physics of proton therapy.

Authors:  Wayne D Newhauser; Rui Zhang
Journal:  Phys Med Biol       Date:  2015-03-24       Impact factor: 3.609

6.  Monte Carlo study of the potential reduction in out-of-field dose using a patient-specific aperture in pencil beam scanning proton therapy.

Authors:  Stephen J Dowdell; Benjamin Clasie; Nicolas Depauw; Peter Metcalfe; Anatoly B Rosenfeld; Hanne M Kooy; Jacob B Flanz; Harald Paganetti
Journal:  Phys Med Biol       Date:  2012-04-19       Impact factor: 3.609

7.  Experimental validation of proton physics models of Geant4 for calculating stopping power ratio.

Authors:  Ruirui Liu; Xiandong Zhao; Maria Medrano
Journal:  J Radiol Prot       Date:  2022-06-28       Impact factor: 1.559

8.  Detecting prompt gamma emission during proton therapy: the effects of detector size and distance from the patient.

Authors:  Jerimy C Polf; Dennis Mackin; Eunsin Lee; Stephen Avery; Sam Beddar
Journal:  Phys Med Biol       Date:  2014-04-15       Impact factor: 3.609

Review 9.  Monte Carlo methods for device simulations in radiation therapy.

Authors:  Hyojun Park; Harald Paganetti; Jan Schuemann; Xun Jia; Chul Hee Min
Journal:  Phys Med Biol       Date:  2021-09-14       Impact factor: 4.174

Review 10.  Range Verification Methods in Particle Therapy: Underlying Physics and Monte Carlo Modeling.

Authors:  Aafke Christine Kraan
Journal:  Front Oncol       Date:  2015-07-07       Impact factor: 6.244

View more